A Headrest Made of Extruded Polystyrene Reduces the Influence of Attenuation Correction on Human Brain SPECT Images

被引:0
作者
Ohba, Makoto [1 ]
Kokubo, Yasuaki [2 ]
Suzuki, Koji [1 ]
Kanoto, Masafumi [3 ]
Sonoda, Yukihiko [2 ]
机构
[1] Yamagata Univ Hosp, Dept Radiol, Yamagata, Japan
[2] Yamagata Univ, Fac Med, Dept Neurosurg, Yamagata, Japan
[3] Yamagata Univ, Dept Radiol, Div Diagnost Radiol, Grad Sch Med Sci, Yamagata, Japan
关键词
SPECT; headrest; image reconstruction; attenuation correction; XPS; extruded polystyrene; SCATTER CORRECTION;
D O I
10.2967/jnmt.122.264729
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Previous reports suggest that a headrest made of carbon signifi-cantly influences cerebral blood flow in the anterior and posterior regions by image reconstruction and attenuation correction (AC). The present study aimed to develop a headrest that reduces the influence of the AC process on human brain SPECT. Methods: To validate the performance of a headrest made of extruded polysty-rene (XPS), 10 healthy controls and 43 patients with cerebrovascu-lar disease underwent 99mTc-ethyl cysteinate dimer SPECT using a carbon headrest and an XPS headrest. We evaluated the anterior -to-posterior and middle-to-posterior ratio of the brain regions in filtered backprojection (FBP) Chang AC, ordered-subset expecta-tion maximization (OSEM) Chang AC, and OSEM CT-based AC. Results: The anterior-to-posterior ratio was significantly higher with the carbon headrest than with the XPS headrest in FBP Chang AC and OSEM Chang AC (P , 0.001). There was no significant differ-ence between the materials in OSEM CT-based AC. The middle -to-posterior ratio did not differ to a statistically significant extent in any correction process. Conclusion: Acquisition of brain SPECT images with an XPS headrest and processing by the FBP or OSEM Chang AC method enables the influence of the headrest to be reduced, especially in anterior and posterior brain regions.
引用
收藏
页码:44 / 48
页数:5
相关论文
共 7 条
[1]   CT Nonuniform Attenuation and TEW Scatter Corrections in Brain Tc-99m ECD SPECT [J].
Farid, Karim ;
Habert, Marie-Odile ;
Martineau, Antoine ;
Caillat-Vigneron, Nadine ;
Sibon, Igor .
CLINICAL NUCLEAR MEDICINE, 2011, 36 (08) :665-668
[2]   CT-based attenuation and scatter correction compared with uniform attenuation correction in brain perfusion SPECT imaging for dementia [J].
Gillen, Rebecca ;
Firbank, Michael J. ;
Lloyd, Jim ;
O'Brien, John T. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (17) :6775-6787
[3]  
Hayashi M, 2005, J NUCL MED TECHNOL, V33, P224
[4]   Impact of CT attenuation correction by SPECT/CT in brain perfusion images [J].
Ishii, Kazunari ;
Hanaoka, Kohei ;
Okada, Masahiro ;
Kumano, Seishi ;
Komeya, Yoshihiro ;
Tsuchiya, Norio ;
Hosono, Makoto ;
Murakami, Takamichi .
ANNALS OF NUCLEAR MEDICINE, 2012, 26 (03) :241-247
[5]   Influence of a Headrest on Reconstruction and Attenuation Correction of Human Brain SPECT Images [J].
Ohba, Makoto ;
Kokubo, Yasuaki ;
Suzuki, Koji ;
Kanoto, Masafumi ;
Sonoda, Yukihiko .
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2021, 49 (01) :54-57
[6]   SPECT/CT Physical Principles and Attenuation Correction [J].
Patton, James A. ;
Turkington, Timothy G. .
JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2008, 36 (01) :1-10
[7]   Usefulness of a three-dimensional stereotaxic ROI template on anatomically standardised 99mTc-ECD SPET [J].
Takeuchi, R ;
Yonekura, Y ;
Matsuda, H ;
Konishi, J .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (03) :331-341