Development and characterization of fission chamber neutron detectors in Isfahan miniature neutron source reactor

被引:12
作者
Teimoory, E. [1 ]
Allaf, M. Athari [1 ]
Mokhtari, J. [2 ]
Dastjerdi, M. H. Choopan [2 ]
机构
[1] Islamic Azad Univ, Sci & Res Branch, POB 14515-775, Tehran, Iran
[2] Nucl Sci & Technol Res Inst, Reactor & Nucl Safety Res Sch, AEOI, Tehran, Iran
关键词
Neutron detection; Fission chamber detector; MNSR; Detection sensitivity; Neutron beamline; CURRENT-MODE; OPTIMIZATION; DESIGN;
D O I
10.1016/j.radphyschem.2023.111360
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
In this research, two fission chamber neutron detectors have been designed and constructed, and their performance has been evaluated in the radiation field of the Isfahan miniature reactor neutron beam line. Each detector consists of two coaxial cylinders called anode and cathode made of stainless steel. The outer layer of the anode is covered with a very thin layer of uranium-235 oxide as fissile material, and a mixture of argon and nitrogen gas is used as the filling gas. The thickness of the deposited layer on the anode is different, so that on one of them, named FC400A, the thickness of the uranium oxide layer was more than 5 mu m, and on the other one, named FC400B, it was less than 1.5 mu m. A spectrum collection software has also been designed and developed in LabVIEW. The developed detectors are irradiated in the neutron field resulting from the neutron radiography beamline of the miniature neutron source reactor in the range of neutron flux of about 2E+02 to 2E+05 n.cm(-2).s(-1). In order to determine the sensitivity of the detection response, the results of these two detectors are also compared with a commercial fission chamber neutron detector. The results showed that the working voltage range of these detectors is in the range of 400-600 V and both of these detectors have a suitable linear response in the mentioned neutron flux range. Also, the sensitivity of FC400A and FC400B detectors is 0.1 and 1 counts/ 1E+04 neutrons, respectively.
引用
收藏
页数:5
相关论文
共 30 条
[21]  
Miller J.F., 1999, Wiley Encyclopedia of Electrical and Electronics Engineering, P1
[22]   Improvement of the quality of neutron radiography beamlines in Isfahan MNSR [J].
Mokhtari, J. ;
Dastjerdi, M. H. Choopan ;
Soleimani, B. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2023, 1056
[23]   Development and characterization of a large thermal neutron beam for neutron radiography at Isfahan MNSR [J].
Mokhtari, J. ;
Dastjerdi, M. H. Choopan .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2023, 1051
[24]   Design and optimization of the new LEU MNSR for neutron radiography using thermal column to upgrade thermal flux [J].
Mokhtari, J. ;
Faghihi, F. ;
Khorsandi, J. .
PROGRESS IN NUCLEAR ENERGY, 2017, 100 :221-232
[25]   Feasibility study of Isfahan MNSR as a calibration thermal neutron source [J].
Moslehi, A. ;
Dastjerdi, M. Choopan ;
Torkzadeh, F. ;
Mokhtari, J. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1024
[26]   Recent developments in neutron detection [J].
Peurrung, AJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 443 (2-3) :400-415
[27]   Modeling of the saturation current of a fission chamber taking into account the distorsion of electric field due to space charge effects [J].
Poujade, O ;
Lebrun, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 433 (03) :673-682
[28]  
Rahmati E., 2023, Journal of Instrumentation, V18
[29]   A new fission chamber dedicated to Prompt Fission Neutron Spectra measurements [J].
Taieb, J. ;
Laurent, B. ;
Belier, G. ;
Sardet, A. ;
Varignon, C. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 833 :1-7
[30]   Design of an in-tank thermal neutron beam for PGNAA application at Isfahan MNSR [J].
Vatani, M. ;
Hassanvand, M. ;
Mokhtari, J. ;
Dastjerdi, M. H. Choopan .
NUCLEAR ENGINEERING AND DESIGN, 2023, 412