Respiratory signal estimation for cardiac perfusion SPECT using deep learning

被引:4
作者
Chen, Yuan [1 ,3 ]
Pretorius, P. Hendrik [1 ]
Lindsay, Clifford [1 ]
Yang, Yongyi [2 ]
King, Michael A. [1 ]
机构
[1] Univ Massachusetts, Dept Radiol, Med Sch, Worcester, MA USA
[2] IIT, Dept Elect & Comp Engn, Chicago, IL USA
[3] UMass Chan Med Sch, Dept Radiol, 55 Lake Ave North, Worcester, MA 01545 USA
基金
美国国家卫生研究院;
关键词
cardiac SPECT; deep learning; respiratory signal; RIGID-BODY MOTION; CENTER-OF-MASS; COMPENSATION; HEART; RECONSTRUCTION; TRACKING; PET; DEFORMATION; ATTENUATION;
D O I
10.1002/mp.16653
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundRespiratory motion induces artifacts in reconstructed cardiac perfusion SPECT images. Correction for respiratory motion often relies on a respiratory signal describing the heart displacements during breathing. However, using external tracking devices to estimate respiratory signals can add cost and operational complications in a clinical setting. PurposeWe aim to develop a deep learning (DL) approach that uses only SPECT projection data for respiratory signal estimation. MethodsA modified U-Net was implemented that takes temporally finely sampled SPECT sub-projection data (100 ms) as input. These sub-projections are obtained by reframing the 20-s list-mode data, resulting in 200 sub-projections, at each projection angle for each SPECT camera head. The network outputs a 200-time-point motion signal for each projection angle, which was later aggregated over all angles to give a full respiratory signal. The target signal for DL model training was from an external stereo-camera visual tracking system (VTS). In addition to comparing DL and VTS, we also included a data-driven approach based on the center-of-mass (CoM) strategy. This CoM method estimates respiratory signals by monitoring the axial changes of CoM for counts in the heart region of the sub-projections. We utilized 900 subjects with stress cardiac perfusion SPECT studies, with 302 subjects for testing and the remaining 598 subjects for training and validation. ResultsThe Pearson's correlation coefficient between the DL respiratory signal and the reference VTS signal was 0.90, compared to 0.70 between the CoM signal and the reference. For respiratory motion correction on SPECT images, all VTS, DL, and CoM approaches partially de-blured the heart wall, resulting in a thinner wall thickness and increased recovered maximal image intensity within the wall, with VTS reducing blurring the most followed by the DL approach. Uptake quantification for the combined anterior and inferior segments of polar maps showed a mean absolute difference from the reference VTS of 1.7% for the DL method for patients with motion >12 mm, compared to 2.6% for the CoM method and 8.5% for no correction. ConclusionWe demonstrate the capability of a DL approach to estimate respiratory signal from SPECT projection data for cardiac perfusion imaging. Our results show that the DL based respiratory motion correction reduces artefacts and achieves similar regional quantification to that obtained using the stereo-camera VTS signals. This may enable fully automatic data-driven respiratory motion correction without relying on external motion tracking devices.
引用
收藏
页码:1217 / 1231
页数:15
相关论文
共 53 条
  • [1] Noncontact Respiratory Monitoring Using Depth Sensing Cameras: A Review of Current Literature
    Addison, Anthony P.
    Addison, Paul S.
    Smit, Philip
    Jacquel, Dominique
    Borg, Ulf R.
    [J]. SENSORS, 2021, 21 (04) : 1 - 16
  • [2] Agarap AF, 2018, ARXIV, V1803
  • [3] [Anonymous], KERN DENS EST US GAU
  • [4] Regularized B-spline deformable registration for respiratory motion correction in PET images
    Bai, Wenjia
    Brady, Michael
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) : 2719 - 2736
  • [5] Respiratory Motion Detection and Correction in ECG-Gated SPECT: a New Approach
    Bitarafan, Ahmad
    Rajabi, Hossein
    Gruy, Bornhard
    Rustgou, Feridoon
    Sharafi, Ali Akbar
    Firoozabady, Hasan
    Yaghoobi, Nahid
    Malek, Hadi
    Pirich, Christian
    Langesteger, Werner
    Beheshti, Mohsen
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2008, 9 (06) : 490 - 497
  • [6] Correction of the respiratory motion of the heart by tracking of the center of mass of thresholded projections: A simulation study using the dynamic MCAT phantom
    Bruyant, PP
    King, MA
    Pretorius, PH
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) : 2159 - 2166
  • [7] Cross-vender, cross-tracer, and cross-protocol deep transfer learning for attenuation map generation of cardiac SPECT
    Chen, Xiongchao
    Hendrik Pretorius, P.
    Zhou, Bo
    Liu, Hui
    Johnson, Karen
    Liu, Yi-Hwa
    King, Michael A.
    Liu, Chi
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2022, 29 (06) : 3379 - 3391
  • [8] Automatic attenuation map estimation from SPECT data only for brain perfusion scans using convolutional neural networks
    Chen, Yuan
    Goorden, Marlies C.
    Beekman, Freek J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (06)
  • [9] Lung motion correction on respiratory gated 3-D PET/CT images
    Dawood, M
    Lang, N
    Jiang, XY
    Schäfers, KP
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (04) : 476 - 485
  • [10] Impact of variability in respiration on estimation and correction of respiratory motion of the heart in SPECT
    Dey, Joyoni
    Johnson, Karen J.
    Pretorius, P. Hendrik
    Mitra, Joyeeta
    King, Michael A.
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4503 - 4509