Evaluation of fetal dose during pelvimetry CT scan procedure by Monte Carlo using GATE

被引:9
作者
Aabid, M. [1 ]
Semghouli, S. [2 ]
Amaoui, B. [3 ]
Choukri, A.
机构
[1] Ibn Tofail Univ, Fac Sci, Dept Phys, NPTT, POB 133, Kenitra, Morocco
[2] Higher Inst Nursing Profess & Hlth Tech ISPITS, Res Lab Hlth & Environm Sci LabReSSE, Team Hlth Tech ETechS, Agadir, Morocco
[3] Univ Ibn Zohr, Fac Med & Pharm, Agadir, Morocco
关键词
COMPUTED-TOMOGRAPHY; PREGNANT-WOMEN; INDEX CTDI; DOSIMETRY; MOROCCO; RISKS;
D O I
10.1016/j.radphyschem.2023.111042
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Pelvimetry is performed at the late pregnancy to measure the size of the pelvis through its different diameters in order to determine the possible modalities of delivery: by natural way or by programmed cesarean section. When a pregnant woman undergoes a radiological examination, the risk associated with radiation exposure is of concern. In general, the fetal dose cannot be measured directly, so Monte Carlo (MC) methods are used for this purpose. Therefore the objective of this study is to evaluate the fetal dose during a Computed Tomography (CT) pelvimetry examination by means of a MC method using Geant4 Application for Tomographic Emission code (GATE). A simulation of a CT pelvimetry examination was performed by modeling a Hitachi Supria 16-slice scanner that follows a CT pelvimetry protocol. A 16-diameter phantom is used to model the fetus inside a 32diameter phantom that represents the patient's body. The fetal dose in this simulation was 0.95 mGy. Other studies on fetal dose from pelvic CT scans found that it does not exceed a few mGys, which is not a real concern. On another hand, the dose showed a dependence on the variation of the tube current and also the voltage. Therefore, these parameters can be used as a basis for action to reduce exposure and limit risk during such examinations.
引用
收藏
页数:5
相关论文
共 24 条
[1]   Assessment of Computed Tomography Dose Index (CTDI) During CT Pelvimetry Using Monte Carlo Simulation [J].
Aabid, M. ;
Semghouli, S. ;
Choukri, A. .
ATOM INDONESIA, 2023, 49 (01) :21-25
[2]  
Aabid M., 2020, LECT NOTES INTELLIGE
[3]  
Amaoui Bouchra, 2023, [Radiation Medicine and Protection, 放射医学与防护], V4, P54
[4]   Radiation exposures in pregnancy, health effects and risks to the embryo/foetus-information to inform the medical management of the pregnant patient [J].
Applegate, Kimberly E. ;
Findlay, Una ;
Fraser, Louise ;
Kinsella, Yvonne ;
Ainsbury, Liz ;
Bouffler, Simon .
JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) :S522-S539
[5]   Can three-dimensional pelvimetry using low-dose stereoradiography replace low-dose CT pelvimetry? [J].
Aubry, S. ;
Padoin, P. ;
Petegnief, Y. ;
Vidal, C. ;
Riethmuller, D. ;
Delabrousse, E. .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2018, 99 (09) :569-576
[6]   Ionizing Radiation-Induced Epigenetic Modifications and Their Relevance to Radiation Protection [J].
Belli, Mauro ;
Tabocchini, Maria Antonella .
INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2020, 21 (17) :1-34
[7]   Comparative dose levels between CT-scanner and slot-scanning device (EOS system) in pregnant women pelvimetry [J].
Ben Abdennebi, A. ;
Aubry, S. ;
Ounalli, L. ;
Fayache, M. S. ;
Delabrousse, E. ;
Petegnief, Y. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 33 :77-86
[8]   Retrospective analysis of dose delivered to the uterus during CT examination in pregnant women [J].
Dabli, Djamel ;
Linard, Melinee ;
Durand, Quentin ;
Frandon, Julien ;
de Oliveira, Fabien ;
Beregi, Jean Paul ;
Greffier, Joel .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2022, 103 (7-8) :331-337
[9]   Advances in Geant4 applications in medicine [J].
Incerti, Sebastien ;
Brown, Jeremy M. C. ;
Guatelli, Susanna .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 :224-227
[10]  
K C., 1997, PHYS ENG MED REP, V78