Heterogeneous breast phantom for computed tomography and magnetic resonance imaging

被引:5
作者
Saleh, Gameel [1 ]
Abuelhaija, Ashraf [2 ]
Alfaris, Budour [1 ]
Aljabr, Aljohara [1 ]
Zainalabedin, Maryam [1 ]
Mhareb, M. H. A. [3 ]
Alhashim, Maryam [4 ]
Alenezi, Salma [4 ]
机构
[1] Imam Abdulrahman Bin Faisal Univ, Coll Engn, Dept Biomed Engn, Dammam, Saudi Arabia
[2] Appl Sci Private Univ, Fac Engn & Technol, Dept Elect Engn, Amman, Jordan
[3] Imam Abdulrahman Bin Faisal Univ, Coll Sci, Dept Phys, Dammam, Saudi Arabia
[4] King Fahad Specialist Hosp KFSH, Dammam, Saudi Arabia
关键词
TISSUE; MULTIPURPOSE;
D O I
10.1371/journal.pone.0284531
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
In this article, a heterogeneous multimodal anthropomorphic breast phantom with carcinoma is introduced to meet the response of the natural breast tissue when imaged using ionizing and non-ionizing machines. The skin, adipose, fibroglandular, pectoral muscle, and carcinoma tissue were mimicked. A T1-weighted breast magnetic resonance image with BI-RADS I tissue segmentation was used for molds creation. The tissue-mimicking materials (TMMs) were tailored in terms of their elemental composition weight fractions and their response to ionization radiation parameters. These are the mass attenuation coefficient (MAC), electron density (n(e)) and effective atomic number (Z(eff)). The behaviour of the TMMs, when exposed to a wide range of ionization radiation energy, was investigated analytically and numerically using X-COM. The achieved results showed an excellent agreement with the corresponding properties of the natural breast elemental compositions as reported by the International Commission on Radiation Units and Measurements (ICRU). The MAC of the TMMs and the ICRU-based breast tissue were found to be consistent. The maximum percentage of error in n(e) and Z(eff) amounts to only 2.93% and 5.76%, respectively. For non-ionizing imaging, the TMMs were characterized in term of T1 and T2 relaxation times. Using our preclinical MRI unit, the TMMs relaxation times were measured and compared to the natural tissue. The fabricated phantom was validated experimentally using CT, MRI, and Mammographic machines. The achieved images of the TMMs were in alignment with the real tissue in terms of CT HU values and grayscale colors. T1W and T2W images on MRI revealed the expected contrast between TMMs as in natural tissue.
引用
收藏
页数:23
相关论文
共 27 条
[1]  
AlFares B., 2021 IEEE INT IOT EL, V2021, P1
[2]  
Browne J, 2020, U.S. Patent, Patent No. [20200170612A1, 20200170612]
[3]   MRI-Derived 3-D-Printed Breast Phantom for Microwave Breast Imaging Validation [J].
Burfeindt, Matthew J. ;
Colgan, Timothy J. ;
Mays, R. Owen ;
Shea, Jacob D. ;
Behdad, Nader ;
Van Veen, Barry D. ;
Hagness, Susan C. .
IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2012, 11 :1610-1613
[4]   Development of a physical 3D anthropomorphic breast phantom [J].
Carton, Ann-Katherine ;
Bakic, Predrag ;
Ullberg, Christer ;
Derand, Helen ;
Maidment, Andrew D. A. .
MEDICAL PHYSICS, 2011, 38 (02) :891-896
[5]   Computed tomography on lung cancer screening is useful for adjuvant comorbidity diagnosis in developing countries [J].
de Mattos, Juliane Nascimento ;
Santiago Escovar, Carlos Eugenio ;
Zereu, Manuela ;
Rubin, Adalberto Sperb ;
Camargo, Spencer Marcantonio ;
Mohammed, Tan-Lucien ;
dos Santos, Ricardo Sales ;
Verma, Nupur ;
Pereira, Diana Penha ;
Pinto, Erique Guedes ;
Machuca, Tiago ;
Medeiros, Tassia Machado ;
Hochhegger, Bruno .
ERJ OPEN RESEARCH, 2022, 8 (02)
[6]   Characterization of a dielectric phantom for high-field magnetic resonance imaging applications [J].
Duan, Qi ;
Duyn, Jeff H. ;
Gudino, Natalia ;
de Zwart, Jacco A. ;
van Gelderen, Peter ;
Sodickson, Daniel K. ;
Brown, Ryan .
MEDICAL PHYSICS, 2014, 41 (10)
[7]   Recent advances on the development of phantoms using 3D printing for imaging with CT, MRI, PET, SPECT, and ultrasound [J].
Filippou, Valeria ;
Tsoumpas, Charalampos .
MEDICAL PHYSICS, 2018, 45 (09) :E740-E760
[8]   Advances in digital and physical anthropomorphic breast phantoms for x-ray imaging [J].
Glick, Stephen J. ;
Ikejimba, Lynda C. .
MEDICAL PHYSICS, 2018, 45 (10) :E870-E885
[9]   3D-printed breast phantom for multi-purpose and multi-modality imaging [J].
He, Yaoyao ;
Liu, Yulin ;
Dyer, Brandon A. ;
Boone, John M. ;
Liu, Shanshan ;
Chen, Tiao ;
Zheng, Fenglian ;
Zhu, Ye ;
Sun, Yong ;
Rong, Yi ;
Qiu, Jianfeng .
QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (01) :63-74
[10]   Design of a breast phantom for quantitative MRI [J].
Keenan, Kathryn E. ;
Wilmes, Lisa J. ;
Aliu, Sheye O. ;
Newitt, David C. ;
Jones, Ella F. ;
Boss, Michael A. ;
Stupic, Karl F. ;
Russek, Stephen E. ;
Hylton, Nola M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 44 (03) :610-619