摘要
Background Since October 1997, an international standard for hepatitis C virus (HCV) nucleic acid amplification technology assay, 96/790, has been available. We compared a series of lyophilized standards with known HCV RNA concentrations against the international standard in fluorescence quantitative PCR detection. Methods A series of lyophilized sera were calibrated by ROCHE COBAS AMPLICOR HCV Monitor test against the international standard and sent to various manufacturers to analyse the samples using their own kits. Then calibration curves from the series were compared with that obtained from the external standard calibration curve with the manufacture’s series. Results The standard calibration curve with the series of lyophilized serum showed an excellent correlation (R2>0.98), slope and intercept that were similar to those from the manufacture’s series. When the standard calibration curve from the series of lyophilized standards were used to define the values of the given sample, lower coefficients of variation between kits from different manufactures were obtained. Conclusion The results showed that the lyophilized standards could be used to setup the standard calibration curve for clinical HCV RNA quantitative PCR detection.
引用
收藏
页码:1910 / 1914
页数:5
相关论文
共 1 条
[1]   Comparison of HCV RNA levels by branched DNA probe assay and by competitive polymerase chain reaction to predict effectiveness of interferon treatment for patients with chronic hepatitis C virus [J].
Hayashi, J ;
Kawakami, Y ;
Nabeshima, A ;
Kishihara, Y ;
Furusyo, N ;
Sawayama, Y ;
Kinukawa, N ;
Kashiwagi, S .
DIGESTIVE DISEASES AND SCIENCES, 1998, 43 (02) :384-391