Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines

被引:0
作者
Mustafa Atas [1 ]
Süleyman Demircan [1 ]
Arzu Seyhan Karatepe Hashas [1 ]
Ahmet Gülhan [1 ]
Gkmen Zararsιz [2 ]
机构
[1] Department of Ophthalmology and Vision Sciences,Kayseri Traning and Research Hospital  2. Department of Biostatistics,Faculty of Medicine,Erciyes University
关键词
endothelial cell changes; transversal; torsional; phacoemulsification;
D O I
暂无
中图分类号
R779.6 [眼外科手术学];
学科分类号
100212 ;
摘要
AIM:To compare and evaluate the phacoemulsification parameters and postoperative endothelial cell changes of two different phacoemulsification machines, each with different modes, but also to assess the relationship between postoperative endothelial cell loss and the phacoemulsification parameters, as well as the other factors in both groups.METHODS:This prospective observational study was comprised of consecutive eligible cataract patients operated with phacoemulsification technique performed by the same surgeon using either a WHITESTAR Signature Ellips FX(transversal, group 1) or Infiniti OZil IP(torsional, group 2) machine.RESULTS:The study included 86 patients. Baseline characteristics in the groups were similar. The median nuclear sclerosis grade was 3(2-4) in the first group and2(2-4) in the second group(P =0.265). Both groups had similar phacoemulsification needle times(group 1: 60.63±36 s; group 2: 55.98±30 s; P =0.789). The percentage of endothelial cell loss 30 d after surgery ranged from 3% to15% with a median of 7% in group 1, and from 2% to13% with a median of 6% in group 2; however, there was no statistically significant difference between the groups(P =0.407). Hexagonality(P =0.794) and the coefficient of variation(CV; P =0.142) did not differ significantly between the groups before and 30 d after surgery. Asignificant positive correlation was found between the endothelial cell loss and nuclear sclerosis grade(group1: P <0.001; group 2: P <0.001) and between the endothelial cell loss and average phacoemulsification power(group 1: P =0.007; group 2: P =0.008).CONCLUSION:Both of these machines were efficient,with similar endothelial cell loss. This endothelial cell loss was related to the increased nuclear sclerosis grade and increased phacoemulsification power.
引用
收藏
页码:822 / 827
页数:6
相关论文
共 16 条
  • [1] The phaco machine: analysing new technology[J] . William J. Fishkind.Current Opinion in Ophthalmology . 2013 (1)
  • [2] Comparative analysis of corneal morphological changes after transversal and torsional phacoemulsification through 2.2 mm corneal incision[J] . Maged Maher Salib Roshdy,Assaf.Clinical Ophthalmology . 2013 (defa)
  • [3] Stroke dynamics and frequency of 3 phacoemulsification machines
    Tognetto, Daniele
    Cecchini, Paolo
    Leon, Pia
    Di Nicola, Marta
    Ravalico, Giuseppe
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (02) : 333 - 342
  • [4] Relationship between endothelial cell loss and microcoaxial phacoemulsification parameters in noncomplicated cataract surgery[J] . Mohamed Mahdy,Hafez,Mohammed,Bhatia,Eid.Clinical Ophthalmology . 2012 (defa)
  • [5] Intraoperative performance and postoperative outcome comparison of longitudinal, torsional, and transversal phacoemulsification machines[J] . Panos G. Christakis,Rosa M. Braga-Mele.Journal of Cataract & Refractive Surgery . 2011 (2)
  • [6] Cataract surgery in eyes with low corneal endothelial cell density
    Hayashi, Ken
    Yoshida, Motoaki
    Manabe, Shin-ichi
    Hirata, Akira
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (08) : 1419 - 1425
  • [7] Thermal Comparison of Infiniti OZil and Signature Ellips Phacoemulsification Systems
    Schmutz, Joseph S.
    Olson, Randall J.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 149 (05) : 762 - 767
  • [8] Comparison of endothelial changes and power settings between torsional and longitudinal phacoemulsification
    Reuschel, Anna
    Bogatsch, Holger
    Barth, Thomas
    Wiedemann, Renate
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (11) : 1855 - 1861
  • [9] Thermal study of longitudinal and torsional ultrasound phacoemulsification[J] . Bokkwan Jun,John P. Berdahl,Terry Kim.Journal of Cataract & Refractive Surgery . 2010 (5)
  • [10] Comparison of torsional and longitudinal modes using phacoemulsification parameters
    Rekas, Marek
    Montes-Mico, Robert
    Krix-Jachym, Karolina
    Klus, Adam
    Stankiewicz, Andrzej
    Ferrer-Blasco, Teresa