A New Method for Blood NT-proBNP Determination Based on a Near-infrared Point of Care Testing Device with High Sensitivity and Wide Scope

被引:0
|
作者
ZHANG Xiao Guang [1 ]
SHU Yao Gen [2 ]
GAO Ju [3 ]
WANG Xuan [2 ]
LIU Li Peng [2 ]
WANG Meng [4 ]
CAO Yu Xi [1 ]
ZENG Yi [1 ]
机构
[1] National Institute for Viral Disease Control and Prevention,Chinese Center for Disease Control and Prevention
[2] Beijing Runbio Biotechnology Development Co.,Ltd  3. No.2 People's Hospital of Dali City  4. Department of Laboratory Medicine,Beijing Hospital,
关键词
NT-proBNP; Near-infrared POCT device; Lateral flow assay;
D O I
暂无
中图分类号
R446.1 [生物化学检验、临床检验];
学科分类号
100208 ;
摘要
Objective To develop a rapid, highly sensitive, and quantitative method for the detection of NT-proBNP levels based on a near-infrared point-of-care diagnostic(POCT) device with wide scope. Methods The lateral flow assay(LFA) strip of NT-proBNP was first prepared to achieve rapid detection. Then, the antibody pairs for NT-proBNP were screened and labeled with the near-infrared fluorescent dye Dylight-800. The capture antibody was fixed on a nitrocellulose membrane by a scribing device. Serial dilutions of serum samples were prepared using NT-proBNP-free serum series. The prepared test strips, combined with a near-infrared POCT device, were validated by known concentrations of clinical samples. The POCT device gave the output of the ratio of the intensity of the fluorescence signal of the detection line to that of the quality control line. The relationship between the ratio value and the concentration of the specimen was plotted as a work curve. The results of 62 clinical specimens obtained from our method were compared in parallel with those obtained from the Roche E411 kit. Results Based on the log-log plot, the new method demonstrated that there was a good linear relationship between the ratio value and NT-proBNP concentrations ranging from 20 pg/mL to 10 ng/mL. The results of the 62 clinical specimens measured by our method showed a good linear correlation with those measured by the Roche E411 kit. Conclusion The new LFA detection method of NT-proBNP levels based on the near-infrared POCT device was rapid and highly sensitive with wide scope and was thus suitable for rapid and early clinical diagnosis of cardiac impairment.
引用
收藏
页码:426 / 431
页数:6
相关论文
共 12 条
  • [1] A New Method for Blood NT-proBNP Determination Based on a Near-infrared Point of Care Testing Device with High Sensitivity and Wide Scope
    Zhang Xiao Guang
    Shu Yao Gen
    Gao Ju
    Wang Xuan
    Liu Li Peng
    Wang Meng
    Cao Yu Xi
    Zeng Yi
    BIOMEDICAL AND ENVIRONMENTAL SCIENCES, 2017, 30 (06) : 426 - 431
  • [2] Development and calibration of a new point-of-care test for the determination of NT-proBNP in whole blood
    Bertsch, Thomas
    Dikkeschei, Bert
    Gurr, Eberhard
    Hayen, Wiebke
    Jorgensen, Bo
    Lotz, Johannes
    Mueller-Bardorff, Margit
    Ordonez-Llanos, Jordi
    Schulz, Ilse
    Spinke, Juergen
    Thiele, Marcel
    Zerback, Rainer
    CLINICAL LABORATORY, 2007, 53 (7-8) : 423 - 431
  • [3] Multicentre evaluation of a new point-of-care test for the determination of NT-proBNP in whole blood
    Zugck, Christian
    Nelles, Manfred
    Katus, Hugo A.
    Collinson, Paul O.
    Gaze, David C.
    Dikkeschei, Bert
    Gurr, Eberhard
    Hayen, Wiebke
    Haass, Markus
    Hechler, Christoph
    van Hoof, Viviane
    Guerti, Khadija
    van Waes, Carl
    Printzen, Gert
    Klopprogge, Kai
    Schulz, Ilse
    Zerback, Rainer
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (10) : 1269 - 1277
  • [4] MEASUREMENT OF NT-PROBNP IN EMERGENCY DEPARTMENT: EVALUATION OF A NEW POINT-OF-CARE DEVICE
    Lepoutre, T.
    Henrard, C.
    Gruson, D.
    Thys, F.
    Verschuren, F.
    ACTA CLINICA BELGICA, 2012, 67 (02): : 146 - 146
  • [6] Performance of the StatusFirst™ CHF NT-proBNP point-of-care device for patients suspected of heart failure using plasma and whole blood
    Shea, E. N.
    Wu, A. H.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A72 - A73
  • [7] A near-infrared spectroscopy-based end-point determination method for the blending process of Dahuang soda tablets
    Wu, Si-jun
    Qiu, Ping
    Li, Pian
    Li, Zheng
    Li, Wen-long
    JOURNAL OF ZHEJIANG UNIVERSITY-SCIENCE B, 2020, 21 (11): : 897 - 910
  • [8] Image Monitoring of Pharmaceutical Blending Processes and the Determination of an End Point by Using a Portable Near-Infrared Imaging Device Based on a Polychromator-Type Near-Infrared Spectrometer with a High-speed and High-Resolution Photo Diode Array Detector
    Murayama, Kodai
    Ishikawa, Daitaro
    Genkawa, Takuma
    Sugino, Hiroyuki
    Komiyama, Makoto
    Ozaki, Yukihiro
    MOLECULES, 2015, 20 (03): : 4007 - 4019
  • [9] Prospective study of point of care device use in NT-proBNP therapeutic monitoring: substudy of GUIDing Evidence Based Therapy Using Biomarker Intensified Treatment in Heart Failure (GUIDE-IT) Trial
    Adams, K. F., Jr.
    Lee, A. K.
    Schwartz, T. A.
    Felker, G. M.
    Januzzi, J. L.
    Pina, I. L.
    Whellan, D. J.
    Masson, S.
    Fiuzat, M.
    Ezekowitz, J. A.
    Leifer, E. S.
    Cooper, L. S.
    Desvigne-Nickens, P.
    O'connor, C. M.
    EUROPEAN JOURNAL OF HEART FAILURE, 2021, 23 : 101 - 101
  • [10] Near-infrared spectroscopic determination of antioxidants and organic acids in rubbers assisted by a new multivariate calibration method based on direct orthogonalization
    Rodríguez, D
    Boschetti, CE
    Olivieri, AC
    ANALYST, 2002, 127 (02) : 304 - 309