Passive shimming based on mixed integer programming for MRI magnet

被引:0
作者
YOU XiaoFei [1 ,2 ]
WANG Zheng [1 ]
ZHANG XiaoBing [1 ]
YANG WenHui [1 ]
SONG Tao [1 ]
机构
[1] Institute of Electrical Engineering,Chinese Academy of Sciences
[2] University of Chinese Academy of Sciences
关键词
mixed integer programming; passive shimming; magnetic resonance imaging(MRI);
D O I
暂无
中图分类号
O482.531 []; O221.4 [整数规划];
学科分类号
070205 ; 0805 ; 080502 ; 0809 ; 070105 ; 1201 ;
摘要
A mathematical model based on mixed integer programming is presented in this paper for the passive shimming of magnet in magnetic resonance imaging(MRI) scanner.In this model,the magnetic field inhomogeneity tolerance and the central value of the magnetic field after shimming are programmed together with the volume of each shim piece as the variables,which increases the degree of freedom and guarantees a better solution.The magnetic field inhomogeneity tolerance after shimming with a weighting coefficient and the total volume of shim pieces are both contained in the objective function of the model.By assigning different values to the weighting coefficient in the objective function,different shimming plans with different emphases can be obtained.A simulation test has been carried out on a small permanent magnet with frame structure.Several solutions are given and compared in this paper,which indicates that a practical shimming plan can be obtained quickly by solving this model.
引用
收藏
页码:1208 / 1212
页数:5
相关论文
共 3 条
[1]  
Active ferromagnetic shimming of the permanent magnet for magnetic resonance imaging scanner[J]. 唐昕,洪礼明,俎栋林.Chinese Physics B. 2010(07)
[2]   基于动态规划的永磁MRI磁体的无源匀场 [J].
谢德馨 ;
王恩龙 ;
张艳丽 .
沈阳工业大学学报, 2007, (04) :396-399
[3]   开放式永磁机构无源匀场方法 [J].
张艳丽 ;
谢德馨 ;
王恩龙 ;
曾林锁 ;
白保东 .
电工技术学报, 2007, (06) :1-5