Using GPU to calculate electron dose for hybrid pencil beam model

被引:0
|
作者
GOU Chengjun LI Xia HOU Qing WU Zhangwen~* Key Laboratory for Radiation Physics and Technology of Ministry of Education
机构
关键词
GPU; HPBM; Dose calculation;
D O I
10.13538/j.1001-8042/nst.22.349-352
中图分类号
R815 [放射疗法];
学科分类号
1001 ; 100105 ; 100207 ; 100602 ;
摘要
Hybrid pencil beam model(HPBM) offers an efficient approach to calculate the three-dimension dose distribution from a clinical electron beam.Still,clinical radiation treatment activity desires faster treatment plan process.Our work presented the fast implementation of HPBM-based electron dose calculation using graphics processing unit(GPU).The HPBM algorithm was implemented in compute unified device architecture running on the GPU,and C running on the CPU,respectively.Several tests with various sizes of the field,beamlet and voxel were used to evaluate our implementation.On an NVIDIA GeForce GTX470 GPU card,we achieved speedup factors of 2.18- 98.23 with acceptable accuracy,compared with the results from a Pentium E5500 2.80 GHz Dual-core CPU.
引用
收藏
页码:349 / 352
页数:4
相关论文
共 50 条
  • [1] Using GPU to calculate electron dose for hybrid pencil beam model
    Gou Chengjun
    Li Xia
    Hou Qing
    Wu Zhangwen
    NUCLEAR SCIENCE AND TECHNIQUES, 2011, 22 (06) : 349 - 352
  • [2] Electron beam dose calculation: Hybrid pencil beam model extended for small fields
    Gou, CJ
    Zeng, G
    Luo, ZM
    ACTA PHYSICA SLOVACA, 2003, 53 (03) : 259 - 265
  • [3] Three-dimensional electron dose calculation using an improved hybrid pencil beam model
    Gou, CJ
    Wu, ZW
    Luo, ZM
    Jette, D
    MEDICAL PHYSICS, 2003, 30 (03) : 415 - 423
  • [4] Electron dose calculation using multiple-scattering theory: A hybrid electron pencil-beam model
    Luo, ZM
    Jette, D
    Walker, S
    MEDICAL PHYSICS, 1998, 25 (10) : 1954 - 1963
  • [5] The application of hybrid-pencil beam in electron dose calculation
    Gou, CJ
    Yang, DL
    Zeng, G
    Luo, ZM
    ACTA PHYSICA SINICA, 2002, 51 (11) : 2649 - 2655
  • [6] The application of hybrid-pencil beam in electron dose calculation
    Gou, Cheng-Jun
    Yang, Dai-Lun
    Zeng, Ge
    Lou, Zheng-Ming
    Wuli Xuebao/Acta Physica Sinica, 2002, 51 (11): : 2654 - 2655
  • [7] GPU-based ultra-fast dose calculation using a finite size pencil beam model
    Gu, Xuejun
    Choi, Dongju
    Men, Chunhua
    Pan, Hubert
    Majumdar, Amitava
    Jiang, Steve B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (20): : 6287 - 6297
  • [8] EVALUATION OF ELECTRON PENCIL BEAM DOSE CALCULATIONS
    HOGSTROM, KR
    MEDICAL PHYSICS, 1985, 12 (04) : 554 - 554
  • [9] USING BACKGROUND SUBTRACTION FOR MEASURING THE DOSE DISTRIBUTION OF AN ELECTRON PENCIL BEAM
    MCPARLAND, BJ
    MEDICAL PHYSICS, 1987, 14 (03) : 406 - 409