Four-Dimensional Computerized Tomography (4D-CT) Reconstruction Based on the Similarity Measure of Spatial Adjacent Images

被引:6
|
作者
ZHANG Shu-xu1
机构
关键词
phantom; four-dimensional computed tomography (4D-CT) reconstruction; similarity; motion artifacts; mutual information;
D O I
暂无
中图分类号
TP391.41 [];
学科分类号
080203 ;
摘要
Objective:To investigate the feasibility of a 4D-CT reconstruction method based on the similarity principle of spatial adjacent images and mutual information measure. Methods:A motor driven sinusoidal motion platform made in house was used to create one-dimensional periodical motion that was along the longitudinal axis of the CT couch. The amplitude of sinusoidal motion was set to an amplitude of ±1 cm. The period of the motion was adjustable and set to 3.5 s. Phantom objects of two eggs were placed in a Styrofoam block, which in turn were placed on the motion platform. These objects were used to simulate volumes of interest undergoing ideal periodic motion. CT data of static phantom were acquired using a multi-slice general electric (GE) LightSpeed 16-slice CT scanner in an axial mode. And the CT data of periodical motion phantom were acquired in an axial and cine-mode scan. A software program was developed by using VC++ and VTK software tools to resort the CT data and reconstruct the 4D-CT. Then all of the CT data with same phase were sorted by the program into the same series based on the similarity principle of spatial adjacent images and mutual information measure among them, and 3D reconstruction of different phase CT data were completed by using the software. Results:All of the CT data were sorted accurately into different series based on the similarity principle of spatial adjacent images and mutual information measures among them. Compared with the unsorted CT data, the motion artifacts in the 3D reconstruction of sorted CT data were reduced significantly, and all of the sorted CT series result in a 4D-CT that reflected the characteristic of the periodical motion phantom. Conclusion:Time-resolved 4D-CT reconstruction can be implemented with any general multi-slice CT scanners based on the similarity principle of spatial adjacent images and mutual information measure.The process of the 4D-CT data acquisition and reconstruction were not restricted to the hardware or software of the CT scanner and has the feasibility ,which extensive applicability.
引用
收藏
页码:106 / 113
页数:8
相关论文
共 50 条
  • [1] Image reconstruction algorithm for four-dimensional CT (4D-CT)
    Taguchi, K
    RADIOLOGY, 2001, 221 : 543 - 544
  • [2] Evaluation of four-dimensional computed tomography (4D-CT)-based pulmonary ventilation: The high correlation between 4D-CT ventilation and 81mKr-planar images was found
    Kanai, Takayuki
    Kadoya, Noriyuki
    Ito, Kengo
    Kishi, Kazuma
    Dobashi, Suguru
    Yamamoto, Takaya
    Umezawa, Rei
    Matsushita, Haruo
    Takeda, Ken
    Jingu, Keiichi
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 (03) : 444 - 448
  • [3] Four-dimensional computed tomography (4D-CT)-based analysis of intra-fractionall esophageal motion
    Michalski, D.
    de Andrade, R. S.
    Heron, D. E.
    Huq, M. S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S518 - S519
  • [4] Evaluation of weighted FDK algorithms applied to four-dimensional CT (4D-CT)
    Mori, S
    Endo, M
    Tsunoo, T
    Murase, K
    Fujiwara, H
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3243 - 3245
  • [5] Dose profile measurement of a four-dimensional CT (4D-CT) including scattered radiation
    Endo, M
    Mori, S
    Tsunoo, T
    Nishizawa, K
    Aoyama, T
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 596 - 609
  • [6] Original Four-dimensional computed tomography (4D-CT) for preoperative parathyroid localization: A good study but are we using it?
    Kuo, Lindsay E.
    Bird, Sarah H.
    Lubitz, Carrie C.
    Pandian, T. K.
    Parangi, Sareh
    Stephen, Antonia E.
    AMERICAN JOURNAL OF SURGERY, 2022, 223 (04): : 694 - 698
  • [7] Four-dimensional CT (4D-CT) image guidance for stereotactic body radiation therapy (SBRT)
    Fuss, M.
    Srisuthep, A.
    Kalpathy-Cramer, J.
    D'Souza, W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S440 - S440
  • [8] Time-Adjusted Internal Target Volume (TTV) based on Four-Dimensional Computed Tomography (4D-CT) for Radiotherapy Planning of Lung Cancer
    Nishibuchi, I.
    Kimura, T.
    Nakashima, T.
    Ochi, Y.
    Murakami, Y.
    Kenjo, M.
    Kaneyasu, Y.
    Nagata, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S535 - S535
  • [9] Introduction of a laser-based optical surface scanning system for retrospective four-dimensional CT (4D-CT): feasibility study
    Maffei, M.
    Haller, M.
    Ferrari, P.
    Stefan, H.
    Lukas, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 : 62 - 62
  • [10] Role of four-dimensional computer tomography (4D-CT) in non-localising and discordant first-line imaging in primary hyperparathyroidism
    Sayed, S.
    Das, A.
    Turner, B.
    Wadhwa, V. S.
    Pathak, K. A.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2023, 105 (08) : 739 - 746