Comparison of retinal nerve fiber layer and macular thickness measurements with Stratus OCT and OPKO/OTI OCT devices in healthy subjects

被引:0
作者
Ahmet Ozkok [1 ]
Julide Canan Umurhan Akkan [2 ]
Nevbahar Tamcelik [1 ]
Mehmet Erdogan [1 ]
Didar Ucar Comlekoglu [1 ]
Rengin Yildirim [1 ]
机构
[1] Department of Ophthalmology,Cerrahpasa School of Medicine,Istanbul University
[2] Department of Ophthalmology,School of Medicine,Bezmialem Vakif University
关键词
optical coherence tomography; Stratus optical coherence tomography; OPKO/OTI OCT; retinal nerve fiber layer; macular thickness;
D O I
暂无
中图分类号
R770.4 [眼科诊断学];
学科分类号
100212 ;
摘要
AIM: To compare retinal nerve fiber layer(RNFL) and macular thickness measurements obtained with the Stratus optical coherence tomography(OCT) and OPKO/OTI OCT devices.METHODS: Included in the study were 59 eyes of 30 participants. All measurements for each eye were done on the same day with both devices. Student’s paired t-tests were used to compare the central macular thickness and RNFL measurements of the Stratus OCT and OPKO/OTI OCT. Pearson correlation was used to assess the relationship between the devices. Coefficient of variation(COV) was calculated to assess intersession repeatability.RESULTS: Using both the Stratus OCT and OPKO/OTI OCT, respectively, the measured mean average RNFL thicknesses were 98.9±11.1 μm and 115.1±9.6 μm(P =0.001),and the measured mean central retinal thicknesses(CRT)were 196.2 ±18.8 μm and 204.5 ±21.1 μm(P <0.001).Measured by the two devices, the RNFL thickness values were correlated in all quadrants, as were the retinal thickness values except the inferior outer sector. COV for average RNFL and CRT thickness were 2.9% and 4.6%for Stratus OCT, and 2.1% and 4.2% for OPKO/OTI OCT,respectively.CONCLUSION: We found good reproducibility of RNFL and retina thickness measurements for both Stratus OCT and OPKO/OTI OCT devices. However, even though the two OCT systems provided statistically correlated results,the values for both RNFL and macular thickness were statistically different. RNFL and macular thickness measurements with the OPKO/OTI OCT were higher than that of the Stratus OCT; therefore, the two OCT systems cannot be used interchangeably for the measurements of RNFL and macular thickness.
引用
收藏
页码:98 / 103
页数:6
相关论文
共 11 条
[1]  
Comparison of central macular thickness between two spectral-domain optical coherence tomography in elderly non-mydriatic eyes[J]. Xiao-Gang Wang 1,Qing Peng 2,Qiang Wu 1 1Department of Ophthalmology,Affiliated Sixth People’s Hospital of Shanghai Jiaotong University,Shanghai 200233,China 2Department of Ophthalmology,Xin hua Hospital Affiliated to Shanghai Jiao Tong University School of Medicine,Shanghai 200092,China.International Journal of Ophthalmology(English Edition). 2012(03)
[2]  
Comparison of Retinal Nerve Fiber Layer Thickness Measurements in Healthy Subjects Using Fourier and Time Domain Optical Coherence Tomography[J] . Isabel Pinilla,Elena Garcia-Martin,Miriam Idoipe,Eva Sancho,Isabel Fuertes,Robert J. Zawadzki.Journal of Ophthalmology . 2012
[3]  
Macular and Retinal Nerve Fiber Layer Thickness Measurements in Normal Eyes With the Stratus OCT, the Cirrus HD-OCT, and the Topcon 3D OCT-1000[J] . Jingjing Huang,Xing Liu,Ziqiang Wu,Xinxing Guo,Hongzhi Xu,Laurie Dustin,Srinivas Sadda.Journal of Glaucoma . 2011 (2)
[4]   Reproducibility of Retinal Nerve Fiber Thickness Measurements Using the Test-retest Function of Spectral OCT/SLO in Normal and Glaucomatous Eyes [J].
Lee, Sun Ho ;
Kim, Seok Hwan ;
Kim, Tae-Woo ;
Park, Ki Ho ;
Kim, Dong Myung .
JOURNAL OF GLAUCOMA, 2010, 19 (09) :637-642
[5]   Comparison of Retinal Nerve Fiber Layer Thickness in Normal Eyes Using Time-Domain and Spectral-Domain Optical Coherence Tomography [J].
Seibold, Leonard K. ;
Mandava, Naresh ;
Kahook, Malik Y. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 150 (06) :807-814
[6]   Comparison of Retinal Nerve Fiber Layer Measurements Using Time Domain and Spectral Domain Optical Coherent Tomography [J].
Knight, O'Rese J. ;
Chang, Robert T. ;
Feuer, William J. ;
Budenz, Donald L. .
OPHTHALMOLOGY, 2009, 116 (07) :1271-1277
[7]   Comparison of Retinal Nerve Fiber Layer Thickness Measured by Cirrus HD and Stratus Optical Coherence Tomography [J].
Sung, Kyung Rim ;
Kim, Dong Yoon ;
Park, Sung Bae ;
Kook, Michael S. .
OPHTHALMOLOGY, 2009, 116 (07) :1264-1270
[8]   Comparison of Macular Thickness Between Cirrus HD-OCT and Stratus OCT [J].
Kakinoki, Masashi ;
Sawada, Osamu ;
Sawada, Tomoko ;
Kawamura, Hajime ;
Ohji, Masahito .
OPHTHALMIC SURGERY LASERS & IMAGING, 2009, 40 (02) :135-140
[9]   Optical coherence tomography to detect and manage retinal disease and glaucoma [J].
Jaffe, GJ ;
Caprioli, J .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 137 (01) :156-169
[10]   Macular and retinal nerve fiber layer thickness measurement reproducibility using optical coherence tomography (OCT-3) [J].
Gürses-Özden, R ;
Teng, C ;
Vessani, R ;
Zafar, S ;
Liebmann, JM ;
Ritch, R .
JOURNAL OF GLAUCOMA, 2004, 13 (03) :238-244