An inversion decomposition method for better energy resolution of NaI(Tl) scintillation detectors based on a Gaussian response matrix

被引:1
作者
Jian-Feng He [1 ,2 ]
Yao-Zong Yang [3 ]
Jin-Hui Qu [1 ]
Qi-Fan Wu [2 ]
Hai-Ling Xiao [1 ]
Cong-Cong Yu [4 ]
机构
[1] Fundamental Science on Radioactive Geology and Exploration Technology Laboratory,East China Institute of Technology
[2] Department of Engineering Physics,Tsinghua University
[3] Jiujiang Power Supply Subsidiary Company of State Grid Jianxi Electronic Power Company
[4] Sichuan Integrative Medicine Hospital Sub-health Testing and Evaluation Laboratory
基金
中国国家自然科学基金;
关键词
Na I(Tl) scintillation detector; Gamma-ray energy spectrum; Gaussian response matrix; Inversion decomposition;
D O I
暂无
中图分类号
TL812 [闪烁探测技术和仪器];
学科分类号
082704 ;
摘要
NaI(T1) scintillation detectors have been widely applied for gamma-ray spectrum measurements owing to advantages such as high detection efficiency and low price.However,the mitigation of the limited energy resolution of these detectors,which detracts from an accurate analysis of the instrument spectra obtained,remains a crucial need.Based on the physical properties and spectrum formation processes of NaI(T1) scintillation detectors,the detector response to gamma photons with different energies is represented by photopeaks that are approximately Gaussian in shape with unique full-width-at-half-maximum(FWHM) values.The FWHM is established as a detector parameter based on resolution calibrations and is used in the construction of a general Gaussian response matrix,which is employed for the inverse decomposition of gamma spectra obtained from the detector.The Gold and Boosted Gold iterative algorithms are employed to accelerate the decomposition of the measured spectrum.Tests of the inverse decomposition method on multiple simulated overlapping peaks and on experimentally obtained U and Th radionuclide series spectra verify the practicability of the method,particularly in the low-energy region of the spectrum,providing for the accurate qualitative and quantitative analysis of radionuclides.
引用
收藏
页码:111 / 120
页数:10
相关论文
共 11 条
  • [1] Implicit FWHM calibration for gamma-ray spectra[J]. WANG Yiming,WEI Yixiang.Nuclear Science and Techniques. 2013(02)
  • [2] Least square fitting of low resolution gamma ray spectra with cubic B-spline basis functions
    祝梦华
    刘良钢
    齐东旭
    尤众
    许敖敖
    [J]. 中国物理C, 2009, (01) : 24 - 30
  • [3] 碘化钠探测器和高纯锗探测器γ能谱仪性能比较
    方晓明
    李欣年
    [J]. 上海大学学报(自然科学版), 2004, (04) : 389 - 392+399
  • [4] Use of an iterative convolution approach for qualitative and quantitative peak analysis in low resolution gamma-ray spectra
    Gardner, Robin P.
    Ai, Xianyun
    Peeples, Cody R.
    Wang, Jiaxin
    Lee, Kyoung
    Peeples, Johanna L.
    Calderon, Adan
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01) : 544 - 549
  • [5] Implementation of Gold deconvolution for enhanced energy resolution in EEL spectra[J] . Staffan Wachtmeister,Stefan Csillag.Ultramicroscopy . 2010 (2)
  • [6] Complete positive deconvolution of spectrometric data[J] . Digital Signal Processing . 2008 (3)
  • [7] An algorithm for determination of peak regions and baseline elimination in spectroscopic data[J] . Nuclear Inst. and Methods in Physics Research, A . 2008 (2)
  • [8] Comparison of LaBr 3 :Ce and NAI(Tl) scintillators for radio-isotope identification devices[J] . B.D. Milbrath,B.J. Choate,J.E. Fast,W.K. Hensley,R.T. Kouzes,J.E. Schweppe.Nuclear Inst. and Methods in Physics Research, A . 2007 (2)
  • [9] Background elimination methods for multidimensional coincidence γ-ray spectra[J] . Nuclear Inst. and Methods in Physics Research, A . 1997 (1)
  • [10] Study of the Van Cittert and Gold iterative methods of deconvolution and their application in the deconvolution of experimental spectra of positron annihilation[J] . Nuclear Inst. and Methods in Physics Research, A . 1997 (2)