Error induced by the estimation of the corneal power and the effective lens position with a rotationally asymmetric refractive multifocal intraocular lens

被引:0
作者
David P.Piero [1 ,2 ,3 ]
Vicente J.Camps [1 ]
María L.Ramón [2 ]
Verónica Mateo [1 ]
Rafael J.Pérez-Cambrodí [2 ,3 ]
机构
[1] Grupo de óptica y Percepción Visual (GOPV).Department of Optics, Pharmacology and Anatomy, University of Alicante, San Vicente del Raspeig, Alicante 03690, Spain
[2] Department of Ophthalmology (Oftalmar), Vithas Medimar International Hospital
[3] Foundation for the Visual Quality (FUNCAVIS: Fundación para la Calidad Visual), Alicante 03016, Spain
关键词
Mplus; multifocalintraocularlens; keratometry; effective lens position; intraocular lens power;
D O I
暂无
中图分类号
R772.2 [角膜疾病];
学科分类号
100212 ;
摘要
AIM : To evaluate the prediction error in intraocular lens(IOL) power calculation for a rotationally asymmetric refractive multifocal IOL and the impact on this error of the optimization of the keratometric estimation of the corneal power and the prediction of the effective lens position(ELP).METHODS: Retrospective study including a total of 25 eyes of 13 patients(age, 50 to 83y) with previous cataract surgery with implantation of the Lentis Mplus LS-312 IOL(Oculentis Gmb H, Germany). In all cases, an adjusted IOL power(P IOLadj) was calculated based on Gaussian optics using a variable keratometric index value(n kadj) for the estimation of the corneal power(P kadj) and on a new value for ELP(ELP adj) obtained by multiple regression analysis.This P IOLadj was compared with the IOL power implanted(P IOLReal) and the value proposed by three conventional formulas(Haigis, Hoffer Q and Holladay Ⅰ).RESULTS: P IOLReal was not significantly different than P IOLadj and Holladay IOL power(P >0.05). In the Bland and Altman analysis, P IOLadj showed lower mean difference(-0.07 D) and limits of agreement(of 1.47 and-1.61 D)when compared to P IOLReal than the IOL power value obtained with the Holladay formula. Furthermore, ELP adj was significantly lower than ELP calculated with other conventional formulas(P <0.01) and was found to be dependent on axial length, anterior chamber depth and P kadj. CONCLUSION: Refractive outcomes after cataract surgery with implantation of the multifocal IOL Lentis Mplus LS-312 can be optimized by minimizing thekeratometric error and by estimating ELP using a mathematical expression dependent on anatomical factors.
引用
收藏
页码:501 / 507
页数:7
相关论文
共 23 条
  • [1] Autorefraction versus subjective refraction in a radially asymmetric multifocal intraocular lens
    van der Linden, Jan Willem
    Vrijman, Violette
    El-Saady, Rana
    van der Meulen, Ivanka J.
    Mourits, Maarten P.
    Lapid-Gortzak, Ruth
    [J]. ACTA OPHTHALMOLOGICA, 2014, 92 (08) : 764 - 768
  • [2] Miguel Faria-Ribeiro,Daniela Lopes-Ferreira,Norberto López-Gil,Jorge Jorge,José Manuel González-Méijome.Errors Associated with IOLMaster Biometry as a Function of Internal Ocular Dimensions[J].Journal of Optometry,2013
  • [3] Minimizing the IOL Power Error Induced by Keratometric Power
    Camps, Vicente J.
    Pinero, David P.
    de Fez, Dolores
    Mateo, Veronica
    [J]. OPTOMETRY AND VISION SCIENCE, 2013, 90 (07) : 639 - 649
  • [4] Jan A. Venter,Martina Pelouskova,Barrie M. Collins,Steven C. Schallhorn,Stephen J. Hannan.Visual outcomes and patient satisfaction in 9366 eyes using a refractive segmented multifocal intraocular lens[J].Journal of Cataract & Refractive Surgery,2013
  • [5] Visual function comparison of 2 aspheric multifocal intraocular lenses
    Alfonso, Jose F.
    Fernandez-Vega, Luis
    Blazquez, Jose I.
    Montes-Mico, Robert
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (02) : 242 - 248
  • [6] Anna-Lotta Engren,Anders Behndig.Anterior Chamber Depth, Intraocular Lens Position and Refractive Outcome after Cataract Surgery[J].Journal of Cataract & Refractive Surgery,2012
  • [7] Comparison of a New Refractive Multifocal Intraocular Lens with an Inferior Segmental Near Add and a Diffractive Multifocal Intraocular Lens
    Alio, Jorge L.
    Plaza-Puche, Ana B.
    Javaloy, Jaime
    Jose Ayala, Maria
    Moreno, Luis J.
    Pinero, David P.
    [J]. OPHTHALMOLOGY, 2012, 119 (03) : 555 - 563
  • [8] Visual function after bilateral implantation of a new zonal refractive aspheric multifocal intraocular lens
    Munoz, Gonzalo
    Albarran-Diego, Cesar
    Ferrer-Blasco, Teresa
    Sakla, Hani E.
    Garcia-Lazaro, Santiago
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (11) : 2043 - 2052
  • [9] Jan Willem van der Linden,Mirjam van Velthoven,Ivanka van der Meulen,Carla Nieuwendaal,Maarten Mourits,Ruth Lapid-Gortzak.Comparison of a new-generation sectorial addition multifocal intraocular lens and a diffractive apodized multifocal intraocular lens[J].Journal of Cataract & Refractive Surgery,2011(1)
  • [10] Comparative analysis of the clinical outcomes with 2 multifocal intraocular lens models with rotational asymmetry
    Alio, Jorge L.
    Plaza-Puche, Ana B.
    Pinero, David P.
    Javaloy, Jaime
    Jose Ayala, Maria
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (09) : 1605 - 1614