Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements

被引:0
作者
Stella Briggs [1 ]
Marwa Bin Moammar [2 ]
机构
[1] College of Applied Medical Sciences, King Saud University
[2] Ophthalmology Clinic, Security Forces Hospital
关键词
central corneal thickness; fluorescein; Scheimpflug device;
D O I
暂无
中图分类号
R770.4 [眼科诊断学];
学科分类号
100212 ;
摘要
AIM: To assess central corneal thickness(CCT)changes measured with Scheimpflug device following instillation of 2% fluorescein in normal subjects.·METHODS: This was a prospective randomized study of 60 hospital volunteers. After baseline CCT measurements of both eyes of 40 subjects were obtained using Scheimpflug system, a drop of preservative-free2% fluorescein, was instilled in one eye and in other eye,one drop of normal saline(control). Measurements were repeated after 1, 2, 5, 10, 20, 30, 40, 50 and 60min(continuous assessment group). Twenty subjects had baseline CCT taken, then fluorescein was instilled in one eye and measurements were taken at 1min. Ten eyes had saline rinse after 1min and 10 other eyes did not,measurements were repeated at 2min(eye rinse group).· RESULTS: The mean baseline CCT for continuous assessment group was 546.2±32.1 μm(range, 489.0-606.0),control eyes was 546.6±30.7 μm(range, 489.0-602.0). At1 min after fluorescein instillation, CCT significantly increased by 37.0 ±34.0 μm(P <0.001), then decreased gradually, reaching baseline at 60 min. CCT variations were not significant in control group(P >0.05). For eye rinse group, CCT mean differences between baseline and 2min were 18.2 μm(95 % CI:-54.7 to 18.3) with rinse and 26.5 μm(95% CI:-62.9 to 9.9) without rinse; paired sample tests were not significant(P >0.05).·CONCLUSION: The presence of fluorescein increased CCT value to a clinically relevant level of 6.8%. Eye rinse did not significantly reduce the effect at 2min post fluorescein timepoint.
引用
收藏
页码:239 / 242
页数:4
相关论文
共 11 条
  • [1] Effect of Fluorescein Dye Staining of the Tear Film on Scheimpflug Measurements of Central Corneal Thickness
    Hirnschall, Nino
    Crnej, Alja
    Gangwani, Vinod
    Findl, Oliver
    [J]. CORNEA, 2012, 31 (01) : 18 - 20
  • [2] Measurement of central corneal thickness by high‐resolution Scheimpflug imaging, Fourier‐domain optical coherence tomography and ultrasound pachymetry[J] . Shihao Chen,Jinhai Huang,Daizong Wen,Wuhe Chen,David Huang,Qinmei Wang.Acta Ophthalmologica . 2010 (5)
  • [3] Risk assessment for ectasia after corneal refractive surgery
    Randleman, J. Bradley
    Woodward, Maria
    Lynn, Michael J.
    Stulting, R. Doyle
    [J]. OPHTHALMOLOGY, 2008, 115 (01) : 37 - 50
  • [4] Comparison of Pentacam Scheimpflug Camera with Ultrasound Pachymetry and Noncontact Specular Microscopy in Measuring Central Corneal Thickness[J] . Miyuki Fujioka,Makoto Nakamura,Yasuko Tatsumi,Azusa Kusuhara,Hidetaka Maeda,Akira Negi.Current Eye Research . 2007 (2)
  • [5] Comparison of Central Corneal Thickness Measurements by Rotating Scheimpflug Camera, Ultrasonic Pachymetry, and Scanning-Slit Corneal Topography[J] . Shiro Amano,Norihiko Honda,Yuki Amano,Satoru Yamagami,Takashi Miyai,Tomokazu Samejima,Miyuki Ogata,Kazunori Miyata.Ophthalmology . 2006 (6)
  • [6] Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry[J] . Yaniv Barkana,Yariv Gerber,Uri Elbaz,Shulamit Schwartz,Gie Ken-Dror,Isaac Avni,David Zadok.Journal of Cataract & Refractive Surgery . 2005 (9)
  • [7] Relation between optical coherence tomography and optical pachymetry measurements of corneal swelling induced by hypoxia
    Wang, JH
    Fonn, D
    Simpson, TL
    Jones, L
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2002, 134 (01) : 93 - 98
  • [8] Comparison of corneal pachymetry using ultrasound and Orbscan II
    Fakhry, MA
    Artola, A
    Belda, JL
    Ayala, MJ
    Alió, JL
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) : 248 - 252
  • [9] Accuracy of Orbscan optical pachymetry in corneas with haze
    Boscia, F
    La Tegola, MG
    Alessio, G
    Sborgia, C
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) : 253 - 258
  • [10] The repeatability of corneal thickness measures
    Marsich, MM
    Bullimore, MA
    [J]. CORNEA, 2000, 19 (06) : 792 - 795