Validation of Peek Acuity application in pediatric screening programs in Paraguay

被引:1
作者
Bryce de Venecia [1 ]
Yasmin Bradfield [1 ]
Ralph M?ller Trane
Alicia Bareiro [2 ]
Miguel Scalamogna [2 ]
机构
[1] Department of Ophthalmology and Visual Sciences,University of Wisconsin Madison
[2] Fundación Visión,Calle Ingavi y Coronel Cazal Departamento
关键词
pediatric ophthalmology; validation study; screening; mobile phone application;
D O I
暂无
中图分类号
R779.7 [小儿眼科学];
学科分类号
100212 ;
摘要
AIM: To validate the Peek Acuity mobile phone application in pediatric populations and compare its utility, both economic and diagnostic, against conventional screening methods using a pediatric ophthalmologist examination as the gold standard.METHODS: A cohort of 393 subjects from Fernando de la Mora, Paraguay(ages 6-16 y) were enrolled in the study. Subjects were randomly assigned a starting screening modality among: Peek Acuity, a single line of tumbling E optotypes set at 20/40, and Spot Vision Screener. Once completing the first screening modality, the subjects completed the two remaining techniques. Referral criteria were established based on the most current American Association of Pediatric Ophthalmology and Strabismus(AAPOS) recommendations: 20/40 for Peek Acuity and the tumbling E, and refractive error detection for the Spot Vision Screener. Subjects that failed to achieve the cutoff for any of the three screening techniques or subjects that passed the screening but were randomly selected to perform a comprehensive eye exam to determine the false negative rate, were evaluated by a pediatric ophthalmologist. This evaluation was considered the gold standard, and included vision assessment by a Snellen chart, strabismus evaluation, and cycloplegic refraction with dilated fundoscopy.RESULTS: We obtained 48% sensitivity, 83% specificity, 43% positive predictive value, and 86% negative predictive value for Peek Acuity’s ability to refer compared to evaluation by a pediatric ophthalmologist, failing to achieve a desired sensitivity for implementation. Peek Acuity trended to overestimate the subject’s visual acuity, providing a higher visual acuity that would not indicate referral for a comprehensive eye examination. However, its high specificity accurately predicted a significant number of children who did not need further evaluation. When comparing the three screening methods, no single screening modality outperformed the others. Peek Acuity represented a technology that was economically feasible compared to other screening modalities in low income settings, due to the prevalence of cell phone use.CONCLUSION: Peek Acuity represents an efficient tool that has potential for implementation in school screenings with different strategies aimed at pediatric populations due to its low cost and high specificity. An increase in sensitivity would improve detection of children with refractive errors.
引用
收藏
页码:1384 / 1389
页数:6
相关论文
共 18 条
[1]   Visual acuity measured with a smartphone app is more accurate than Snellen testing by emergency department providers [J].
Pathipati, Akhilesh S. ;
Wood, Edward H. ;
Lam, Carson K. ;
Sales, Christopher S. ;
Moshfeghi, Darius M. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2016, 254 (06) :1175-1180
[2]   Evaluation of the Spot Vision Screener in young children in Costa Rica [J].
Arana Mendez, Maricela ;
Arguello, Leslie ;
Martinez, Joaquin ;
Salas Vargas, Marisela ;
Alvarado Rodriguez, Ana Maria ;
Papa, Carrie E. ;
Peterseim, Carolyn M. ;
Otarola Viquez, Mauricio ;
Trivedi, Rupal H. ;
Peterseim, Mae Millicent W. .
JOURNAL OF AAPOS, 2015, 19 (05) :441-444
[3]  
Results of a pediatric vision screening program in western South Dakota[J] . Terveen Daniel C,Moser Jess M,Spencer Terrence S.South Dakota medicine : the journal of the South Dakota State Medical Association . 2015 (3)
[4]  
Empirical Studies on Usability of mHealth Apps: A Systematic Literature Review[J] . Belén Cruz Zapata,José Luis Fernández-Alemán,Ali Idri,Ambrosio Toval.Journal of Medical Systems . 2015 (2)
[5]   The effectiveness of the Spot Vision Screener in detecting amblyopia risk factors [J].
Peterseim, Mae Millicent W. ;
Papa, Carrie E. ;
Wilson, M. Edward ;
Davidson, Jennifer D. ;
Shtessel, Maria ;
Husain, Mavesh ;
Cheeseman, Edward W. ;
Wolf, Bethany J. ;
Trivedi, Rupal .
JOURNAL OF AAPOS, 2014, 18 (06) :539-542
[6]  
Validation of Spot screening device for amblyopia risk factors[J] . Glynnis A. Garry,Sean P. Donahue.Journal of AAPOS . 2014
[7]   Influence of uncorrected refractive error and unmet refractive error on visual impairment in a Brazilian population [J].
Ferraz, Fabio H. ;
Corrente, Jose E. ;
Opromolla, Paula ;
Schellini, Silvana A. .
BMC OPHTHALMOLOGY, 2014, 14
[8]  
Vision Screening by Teachers in Southern Indian Schools: Testing a New “All Class Teacher”Model[J] . Adhiseshan Priya,Kannusamy Veena,Ravilla Thulasiraj,Mouttapa Fredrick,Rengaraj Venkatesh,Sabyasachi Sengupta,Ken Bassett.Ophthalmic Epidemiology . 2014 (1)
[9]   Epidemiology, aetiology and management of visual impairment in children [J].
Solebo, Ameenat Lola ;
Rahi, Jugnoo .
ARCHIVES OF DISEASE IN CHILDHOOD, 2014, 99 (04) :375-379
[10]  
A new method to address verification bias in studies of clinical screening tests: cervical cancer screening assays as an example[J] . Xiaonan Xue,Mimi Y. Kim,Philip E. Castle,Howard D. Strickler.Journal of Clinical Epidemiology . 2013