Determination of mycophenolic acid in human plasma by ultra performance liquid chromatography tandem mass spectrometry

被引:1
作者
Vivek Upadhyay [1 ]
Vikas Trivedi [1 ]
Gaurang Shah [1 ]
Manish Yadav [1 ]
Pranav SShrivastav [1 ,2 ]
机构
[1] Chemistry Department,Kadi Sarva Vishwavidyalaya,Gandhinagar ,India
[2] Department of Chemistry,School of Sciences,Gujarat University
关键词
Mycophenolic acid; Mycophenolic acid-d3; UPLC–MS/MS; In-source conversion; Incurred sample reanalysis;
D O I
暂无
中图分类号
R927.2 [药物含量测定];
学科分类号
1007 ;
摘要
A simple, sensitive and high throughput ultra performance liquid chromatography tandem mass spectrometry method has been developed for the determination of mycophenolic acid in human plasma. The method involved simple protein precipitation of MPA along with its deuterated analog as an internal standard(IS) from 50 mL of human plasma. The chromatographic analysis was done on Acquity UPLC C18(100 mm 2.1 mm, 1.7 mm) column under isocratic conditions using acetonitrile and 10 mM ammonium formate, pH 3.00(75:25, v/v) as the mobile phase. A triple quadrupole mass spectrometer operating in the positive ionization mode was used for quantitation. In-source conversion of mycophenolic glucuronide metabolite to the parent drug was selectively controlled by suitable optimization of cone voltage, cone gas flow and desolvation temperature. The method was validated over a wide concentration range of 15–15000 ng/mL. The mean extraction recovery for the analyte and IS was 495%. Matrix effect expressed as matrix factors ranged from 0.97 to 1.02. The method was successfully applied to support a bioequivalence study of 500 mg mycophenolate mofetil tablet in 72 healthy subjects.
引用
收藏
页码:205 / 216
页数:12
相关论文
共 26 条
[1]  
Simultaneous quantification of prodrug oseltamivir and its metabolite oseltamivir carboxylate in human plasma by LC-MS/MS to support a bioequivalence study[J]. Ajay Gupta,Swati Guttikar,Pranav S. Shrivastav,Mallika Sanyal.Journal of Pharmaceutical Analysis. 2013(03)
[2]  
Liquid chromatography–tandem mass spectrometry method for the quantification of mycophenolic acid and its phenolic glucuronide in saliva and plasma using a standardized saliva collection device[J] . Journal of Chromatography A . 2012
[3]  
Single-dose, Two-way Crossover, Bioequivalence Study of Mycophenolate Mofetil 500 mg Tablet Under Fasting Conditions in Healthy Male Subjects[J] . Shashank Patel,Vijay Chauhan,Jayanta Mandal,Samir Shah,Keyur Patel,Dipak Saptarshi,Kirti Maheshwari,Pankaj Kumar Jha,Prashant Kale,Kamlesh Patel,Philip Mathew.Clinical Therapeutics . 2011 (3)
[4]  
Incurred sample reanalysis (ISR): a decisive tool in bioanalytical research[J] . Manish Yadav,Pranav S Shrivastav.Bioanalysis . 2011 (9)
[5]  
Determination of mycophenolic acid and its phenyl glucuronide in human plasma, ultrafiltrate, blood, DBS and dried plasma spots[J] . Katja Heinig,Franz Bucheli,Roswitha Hartenbach,Almudena Gajate-Perez.Bioanalysis . 2010 (8)
[6]  
Mycophenolate mofetil 500-mg tablet under fasting conditions: Single-dose, randomized-sequence, open-label, four-way replicate crossover, bioequivalence study in healthy subjects[J] . Clinical Therapeutics . 2010 (3)
[7]  
Bioequivalence and pharmacokinetic comparison of two mycophenolate mofetil formulations in healthy Chinese male volunteers: An open-label, randomized-sequence, single-dose, two-way crossover study[J] . Qian Zhang,Yifu Tao,Yubing Zhu,Dingchun Zhu.Clinical Therapeutics . 2010 (1)
[8]  
Modification of the EMIT immunoassay for the measurement of unbound mycophenolic acid in plasma[J] . Noemí Rebollo,M. Victoria Calvo,Ana Martín-Suárez,Alfonso Domínguez-Gil.Clinical Biochemistry . 2010 (2)
[9]  
UPLC MS/MS method for quantification of mycophenolic acid and metabolites in human plasma: Application to pharmacokinetic study[J] . Xavier Delavenne,Ludivine Juthier,Bertrand Pons,Christophe Mariat,Thierry Basset.Clinica Chimica Acta . 2010 (1)
[10]  
Sample cleanup-free determination of mycophenolic acid and its glucuronide in serum and plasma using the novel technology of ultra-performance liquid chromatography–electrospray ionization tandem mass spectrometry[J] . Talanta . 2009 (5)