Evaluation of Safety and Efficacy of Qinming8631 DR Implantable Cardiac Pacemaker in Chinese Patients: A Prospective, Multicenter, Randomized Controlled Trial of the First Domestically Developed Pacemaker of China

被引:0
作者
Xiang MeiXiang
Wang DongQi
Xu Jing
Zhang Zheng
Hu JianXin
Wang DongMei
Gu Xiang
Liu HePing
Guo Tao
Yang XiangJun
Ling Feng
Lin JiaFeng
Cai ShangLang
Zhu GuoBin
Wang JianAn
Department of Cardiology Cardiovascular Key Laboratory of Zhejiang Province The Second Affiliated Hospital of Zhejiang University School of Medicine Hangzhou Zhejiang China [310009 ]
Department of Cardiology The First Affiliated Hospital of Xian Jiaotong University Xian Shaanxi China [710061 ]
Department of Cardiology Tianjin Chest Hospital Tianjin China [300051 ]
Department of Cardiology The First Hospital of Lanzhou University Lanzhou Gansu China [730000 ]
Department of Cardiology The Second Affiliated Hospital of Nanchang University Nanchang Jiangxi China [330008 ]
Department of Cardiology Bethune International Peace Hospital of the Chinese PLA Shijiazhuang Hebei China [50082 ]
Department of Cardiology Northern Jiangsu Peoples Hospital Yangzhou Jiangsu China [225001 ]
Department of Cardiology Jilin Province Peoples Hospital Changchun Jilin China [130021 ]
Department of Cardiology The First Affiliated Hospital of Kunming Medical University Kunming Yunnan China [650032 ]
Department of Cardiology The First Affiliated Hospital of Suzhou University Suzhou Jiangsu China [215006 ]
Department of Cardiology The First Peoples Hospital of Hangzhou Hangzhou Zhejiang China [310006 ]
Department of Cardiology The Second Affiliated Hospital of Wenzhou Medical University Wenzhou Zhejiang China [325027 ]
Department of Cardiology The Affiliated Hospital of Qingdao University Medical College Qingdao Shandong China [266003 ]
Department of Cardiology The Second Hospital of Shanxi Medical University Taiyuan Shanxi China [30001 ]
机构
[1] Department of Cardiology, Bethune International Peace Hospital of the Chinese PLA, Shijiazhuang, Hebei 050082, China
[2] Department of Cardiology, Cardiovascular Key Laboratory of Zhejiang Province, The Second Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, Zhejiang 310009, China
[3] Department of Cardiology, Jilin Province People’s Hospital, Changchun, Jilin 130021, China
[4] Department of Cardiology, Northern Jiangsu People’s Hospital, Yangzhou, Jiangsu 225001, China
[5] Department of Cardiology, The Affiliated Hospital of Qingdao University Medical College, Qingdao, Shandong 266003, China
[6] Department of Cardiology, The First Affiliated Hospital of Kunming Medical University, Kunming, Yunnan 650032, China
[7] Department of Cardiology, The First Affiliated Hospital of Suzhou University, Suzhou, Jiangsu 215006, China
[8] Department of Cardiology, The First Affiliated Hospital of Xi’an Jiaotong University, Xi’an, Shaanxi 710061, China
[9] Department of Cardiology, The First Hospital of Lanzhou University, Lanzhou, Gansu 730000, China
[10] Department of Cardiology, The First People’s Hospital of Hangzhou, Hangzhou, Zhejiang 310006, China
[11] Department of Cardiology, The Second Affiliated Hospital of Nanchang University, Nanchang, Jiangxi 330008, China
[12] Department of Cardiology, The Second Affiliated Hospital of Wenzhou Medical University, Wenzhou, Zhejiang 325027, China
[13] Department of Cardiology, The Second Hospital of Shanxi Medical University, Taiyuan, Shanxi 030001, China
[14] Department of Cardiology, Tianjin Chest Hospital, Tianjin 300051, China
关键词
Bradyarrhythmia; Cardiac Pacemaker; Qinming8631; Safety and Efficacy;
D O I
暂无
中图分类号
R541.7 [心律失常];
学科分类号
1002 ; 100201 ;
摘要
Background: High cost of imported pacemakers is a main obstacle for Chinese patients suffering from bradyarrhythmia, and a domestically developed pacemaker will help lower the burden. This study aimed to evaluate the safety and efficacy of Qinming8631 DR (Qinming Medical, Baoji, China), the first domestically developed dual-chamber pacemaker of China, compared with a commercially available pacemaker Talos DR (Biotronik, Berlin, Germany) in Chinese patients.Methods: A prospective randomized trial was conducted at 14 centers in China. Participants were randomized into trial (Qinming8631 DR) and control (Talos DR) groups. Parameters of the pacing systems were collected immediately after device implantation and during follow-ups. The effective pacing rate at 6-month follow-up was recorded as the primary end point. Electrical properties, magnet response, single- and double-pole polarity conversion, rate response function, and adverse events of the pacing system were analyzed. The Cochran-Mantel-Haenszel Chi-square test, pairedt-test, and Wilcoxon signed-rank test were used for measuring primary qualitative outcomes and comparing normally and abnormally distributed measurement data.Results: A total of 225 patients with a diagnosis of bradyarrhythmia and eligible for this study were randomly enrolled into the trial (n = 113) and control (n = 112) groups. They underwent successful pacemaker implantation with acceptable postoperative pacing threshold and sensitivity. Effective pacing rates of trial and control groups were comparable both in the full analysis set and the per protocol set (81.4% vs. 79.5%,P = 0.712 and 95.4% vs. 89.5%,P = 0.143, respectively). In both data sets, noninferiority of the trial group was above the predefined noninferiority limit (?9.5%).Conclusions: This study established the noninferiority of Qinming8631 DR to Talos DR. The safety and efficacy of Qinming8631 DR pacemaker were comparable to those of Talos DR in treating patients with cardiac bradyarrhythmia.
引用
收藏
页码:2659 / 2665
页数:7
相关论文
共 7 条
  • [1] 国产心脏起搏器Qingming 2312S/M与进口心脏起搏器临床效果研究
    孟天宇
    张勇
    袁记行
    [J]. 中国医药科学, 2014, (09) : 183 - 185
  • [2] 国产心脏起搏器Qinming 2312S/M临床应用的研究
    李晓燕
    谈红
    刘科卫
    李爱国
    高玉琪
    晋群
    韩淑芳
    [J]. 中国心脏起搏与心电生理杂志, 2013, 27 (01) : 11 - 12
  • [3] Initial Experience with MultiPoint Pacing Cardiac Resynchronization Therapy in China
    Hua, Wei
    Ding, Li-Gang
    Fan, Xiao-Han
    Liu, Zhi-Min
    Jiang, Chun-Lan
    Qu, Fu-Jian
    Zhang, Shu
    [J]. CHINESE MEDICAL JOURNAL, 2016, 129 (10) : 1241 - 1243
  • [4] Gender differences in dual-chamber pacemaker implantation indications and long-term outcomes
    Dbski, Maciej
    Ulman, Mateusz
    Zabek, Andrzej
    Haberka, Kazimierz
    Lelakowski, Jacek
    Malecka, Barbara
    [J]. ACTA CARDIOLOGICA, 2016, 71 (01) : 41 - 45
  • [5] Assessment of Adaptive Rate Response Provided by Accelerometer, Minute Ventilation and Dual Sensor Compared with Normal Sinus Rhythm During Exercise: A Self-controlled Study in Chronotropically Competent Subjects[J] . Yuanyuan Cao,Yiqun Zhang,Yangang Su,Jin Bai,Wei Wang,Junbo Ge. Chinese Medical Journal . 2015 (1)
  • [6] New Directions in Cardiac Arrhythmia Management: Present Challenges and Future Solutions[J] . Stanley Nattel,Jason Andrade,Laurent Macle,Lena Rivard,Katia Dyrda,Blandine Mondesert,Paul Khairy. Canadian Journal of Cardiology . 2014 (12)
  • [7] The Australian and New Zealand Cardiac Pacemaker and Implantable Cardioverter-Defibrillator Survey: Calendar Year 2013[J] . Harry G. Mond,Ian Crozier. Heart, Lung and Circulation . 2014