Performance study of aluminum shielded room for ultra-low-field magnetic resonance imaging based on SQUID:Simulations and experiments

被引:0
|
作者
李波 [1 ,2 ,3 ]
董慧 [2 ,3 ]
黄小磊 [2 ,3 ,4 ]
邱阳 [1 ,2 ,3 ]
陶泉 [2 ,3 ]
朱建明 [1 ]
机构
[1] China Jiliang University
[2] State Key Laboratory of Functional Materials for Informatics, Shanghai Institute of Microsystem and Information Technology (SIMIT),Chinese Academy of Sciences (CAS)
[3] CAS Center for Excelle Nce in Superconducting Electronics (CENSE)
[4] University of Chinese Academy of Sciences
基金
中国国家自然科学基金;
关键词
shielding effectiveness; aluminum shielded room; eddy current cancellation technique; superconducting quantum interference device; ultra-low-field magnetic resonance imaging;
D O I
暂无
中图分类号
O482.531 [];
学科分类号
070205 ; 0805 ; 080502 ; 0809 ;
摘要
The aluminum shielded room has been an important part of ultra-low-field magnetic resonance imaging(ULF MRI)based on the superconducting quantum interference device(SQUID). The shielded room is effective to attenuate the external radio-frequency field and keep the extremely sensitive detector, SQUID, working properly. A high-performance shielded room can increase the signal-to-noise ratio(SNR) and improve image quality. In this study, a circular coil with a diameter of 50 cm and a square coil with a side length of 2.0 m was used to simulate the magnetic fields from the nearby electric apparatuses and the distant environmental noise sources. The shielding effectivenesses(SE) of the shielded room with different thicknesses of aluminum sheets were calculated and simulated. A room using 6-mm-thick aluminum plates with a dimension of 1.5 m×1.5 m×2.0 m was then constructed. The SE was experimentally measured by using three-axis SQUID magnetometers, with tranisent magnetic field induced in the aluminum plates by the strong pre-polarization pulses. The results of the measured SE agreed with that from the simulation. In addition, the introduction of a 0.5-mm gap caused the obvious reduction of SE indicating the importance of door design. The nuclear magnetic resonance(NMR) signals of water at 5.9 kHz were measured in free space and in a shielded room, and the SNR was improved from 3 to 15. The simulation and experimental results will help us design an aluminum shielded room which satisfies the requirements for future ULF human brain imaging. Finally, the cancellation technique of the transient eddy current was tried, the simulation of the cancellation technique will lead us to finding an appropriate way to suppress the eddy current fields.
引用
收藏
页码:277 / 283
页数:7
相关论文
共 11 条
  • [11] Adaptive suppression of power line interference in ultra-low field magnetic resonance imaging in an unshielded environment
    Huang, Xiaolei
    Dong, Hui
    Qiu, Yang
    Li, Bo
    Tao, Quan
    Zhang, Yi
    Krause, Hans-Joachim
    Offenhaeusser, Andreas
    Xie, Xiaoming
    JOURNAL OF MAGNETIC RESONANCE, 2018, 286 : 52 - 59