Measurements of Air-kerma in Mammography X-Ray Standard Using Free-air Chamber

被引:0
作者
GUO Siming
WU Jinjie
LIU Ying
机构
[1] NationalInstituteofMetrology
关键词
Metrology; Free-air Chamber; Air-kerma; Correction Factor; Degree Of Uncertainty;
D O I
10.15878/j.cnki.instrumentation.2017.02.002
中图分类号
R814.2 [X线剂量学];
学科分类号
1001 ; 100105 ; 100207 ; 100602 ;
摘要
Mammography x-rays standard is required to improve the quality of mammary machine and reduce the dose for w omen and staff. The molybdenum target x-rays machine is used as the radiation source and free-air cylindrical chamber is designed for detecting photon. Four standard radiation qualities are established w ith international standard documents as references. The Correction factors including air attenuation,scattered radiation,fluorescence,w all transmission,electron loss are studied in the four radiation quality. The uncertainty in absolute measurement of air-kerma includes air volume measurement of ionization chambers,current measurement,measurement of temperature and pressure,measurement of correction factors,physical constants and orientation of the reference distance. The relative combined standard uncertainty is 0.24%.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 5 条
[1]   乳腺癌影像诊断的现状与进展 [J].
王娇 ;
杜丽洁 ;
李健丁 .
中华临床医师杂志(电子版), 2016, 10 (04) :579-583
[2]   美国国家标准局自由空气电离室空气吸收改正项的测量 [J].
张友忠 ;
Paul J.Lamperti .
辐射防护, 1991, (03) :174-178
[3]  
Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the NIST, USA and the BIPM in mammography x-rays [J] . C Kessler,D T Burns,M O’Brien.&nbsp&nbspMetrologia . 2011 (1A)
[4]  
Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the NMIJ, Japan and the BIPM in mammography x-rays [J] . C Kessler,D T Burns,T Tanaka,T Kurosawa,N Saito.&nbsp&nbspMetrologia . 2010 (1A)
[5]   Free-air ionization chambers [J].
Burns, D. T. ;
Bueermann, L. B. .
METROLOGIA, 2009, 46 (02) :S9-S23