Monte Carlo simulation for performance evaluation of detector model with a monolithic LaBr3(Ce) crystal and SiPM array for γ radiation imaging

被引:1
作者
Wei Lu [1 ]
Lei Wang [1 ]
Yong Yuan [2 ]
Ting Zhang [1 ]
Ge Tang [1 ]
Ming-Zhe Liu [1 ]
Xian-Guo Tuo [3 ]
Jing Ning [2 ]
机构
[1] Chengdu University of Technology
[2] Beijing Institute of Radiation Medicine
[3] Sichuan University of Science and Engineering
基金
中国国家自然科学基金;
关键词
D O I
暂无
中图分类号
TL812 [闪烁探测技术和仪器];
学科分类号
082704 ;
摘要
In recent years, LaBr3(Ce) crystals and silicon photomultipliers(SiPMs) have been increasingly used in radiation imaging. This study involved the establishment of a detector model with a monolithic LaBr3(Ce) crystal and SiPM array for γ-radiation imaging on the GEANT4 platform. The optical process included in the detector model was defined by key parameters, such as the emission spectrum, scintillation yield, and intrinsic resolution of the LaBr3:5% Ce crystal, as well as the detection efficiency of the SiPM array. The response of the detector model to57Co flooded field irradiation was simulated and evaluated. The radiation images generated by the detector model exhibited a compression effect that was very close to that on images acquired by the physical detector. The spatial resolution of the simulated detector closely approximates that of the physical experiment. A detector model without the optical process was also established for comparison with a detector using the optical process. Both were used in a near-field modified uniform redundant array(MURA) imaging system to acquire images of a point source and a ring source of57Co at the center of the field-of-view of the imaging system. The spatial resolution and signal-to-noise ratio of the images that were reconstructed using the two detector models were determined and compared. Compared with the detector model without optical processes, although the images from the proposed detector model have slightly inferior signal-to-noise ratios and more artifacts, they are more consistent with the reconstructed versions of images acquired in real physical experiments. The results confirm that the detector model can be used to design a γ-radiation imaging detector and to develop an imaging algorithm that can significantly shorten the development time and reduce the cost.
引用
收藏
页码:165 / 176
页数:12
相关论文
共 12 条
[1]   基于MURA编码孔径准直器核辐射成像系统设计 [J].
陈立宏 ;
李勇平 ;
赵翠兰 ;
袁超 .
核技术, 2013, 36 (08) :64-67
[2]   双模测量法消除编码孔成像的近场伪影 [J].
肖洒 ;
兰明聪 ;
党晓军 ;
张连平 ;
韦孟伏 .
核技术, 2013, 36 (08) :45-49
[3]   Investigation of high resolution compact gamma camera module based on a continuous scintillation crystal using a novel charge division readout method [J].
代秋声 ;
赵翠兰 ;
张华林 ;
漆玉金 .
中国物理C, 2010, (08) :1148-1152
[4]  
Near-field high-resolution coded aperture gamma-ray imaging with separable masks[J] . Shifeng Sun,Yang Liu,Xiaoping Ouyang.Nuclear Inst. and Methods in Physics Research, A . 2020 (C)
[5]  
A portable internal contamination monitor with dual detectors for screening in a large-scale radiological incident[J] . Lei Wang,Yi-ni Wu,Jie Pan,Jing Ning,Xin Liu,Xue-lan Yuan,Ting Zhang,Wei Lu,Xian-guo Tuo.Applied Radiation and Isotopes . 2019 (C)
[6]   Adapting clinical gamma cameras for body monitoring in the event of a large-scale radiological incident [J].
Scuffham, J. W. ;
Yip-Braidley, M. ;
Shutt, A. L. ;
Hinton, P. J. ;
Nisbet, A. ;
Bradley, D. A. .
JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (02) :363-381
[7]  
The Experimental Study of Lanthanum Bromide Detector in the Gamma Ray Imaging[J] . Cui Lan Zhao,Yong Ping Li,Li Hong Chen,Chao Yuan,Yu Jin Qi.Applied Mechanics and Materials . 2013 (336-)
[8]  
MgO reflectance data for Monte Carlo simulation of LaBr 3 :Ce scintillation crystals[J] . Raffaele Scafè,Roberto Pani,Rosanna Pellegrini,Maria Nerina Cinti,Paolo Bennati,Sergio Lo Meo.Nuclear Inst. and Methods in Physics Research, A . 2013
[9]  
Scintillation properties of LaBr 3 :Ce 3+ crystals: fast, efficient and high-energy-resolution scintillators[J] . E.V.D van Loef,P Dorenbos,C.W.E van Eijk,K.W Kr?mer,H.U Güdel.Nuclear Inst. and Methods in Physics Research, A . 2002 (1)
[10]  
A coded aperture for high-resolution nuclear medicine planar imaging with a conventional Anger camera: experimental results[J] . R. Accorsi,F. Gasparini,R.C. Lanza.IEEE Transactions on Nuclear Science . 2001 (6P3)