Swab culture monitoring of automated endoscope reprocessors after high-level disinfection

被引:0
|
作者
Lung-Sheng Lu [1 ]
Yi-Chun Chiu [1 ]
Ming-Tzung Lin [1 ]
Tsung-Hui Hu [1 ]
King-Wah Chiu [1 ]
机构
[1] Division of HepatoGastroenterology,Department of Internal Medicine,Chang Gung Memorial Hospital-Kaohsiung Medical Center,Chang Gung University College of Medicine
关键词
Automated endoscope reprocessor; Gastrointestinal scope; High-level disinfection; Swab culture; Monitoring; Decontamination;
D O I
暂无
中图分类号
R187 [消毒];
学科分类号
100401 ;
摘要
AIM:To conduct a bacterial culture study for monitoring decontamination of automated endoscope reprocessors(AERs) after high-level disinfection(HLD).METHODS:From February 2006 to January 2011,authors conducted randomized consecutive sampling each month for 7 AERs.Authors collected a total of 420 swab cultures,including 300 cultures from 5 gastroscope AERs,and 120 cultures from 2 colonoscope AERs.Swab cultures were obtained from the residual water from the AERs after a full reprocessing cycle.Samples were cultured to test for aerobic bacteria,anaerobic bacteria,and mycobacterium tuberculosis.RESULTS:The positive culture rate of the AERs was 2.0%(6/300) for gastroscope AERs and 0.8%(1/120) for colonoscope AERs.All the positive cultures,including 6 from gastroscope and 1 from colonoscope AERs,showed monofloral colonization.Of the gastroscopeAER samples,50%(3/6) were colonized by aerobic bacterial and 50%(3/6) by fungal contaminations.CONCLUSION:A full reprocessing cycle of an AER with HLD is adequate for disinfection of the machine.Swab culture is a useful method for monitoring AER decontamination after each reprocessing cycle.Fungal contamination of AERs after reprocessing should also be kept in mind.
引用
收藏
页码:1660 / 1663
页数:4
相关论文
共 28 条
  • [1] Swab culture monitoring of automated endoscope reprocessors after high-level disinfection
    Lu, Lung-Sheng
    Wu, Keng-Liang
    Chiu, Yi-Chun
    Lin, Ming-Tzung
    Hu, Tsung-Hui
    Chiu, King-Wah
    WORLD JOURNAL OF GASTROENTEROLOGY, 2012, 18 (14) : 1660 - 1663
  • [2] Surveillance cultures of samples obtained from biopsy channels and automated endoscope reprocessors after high-level disinfection of gastrointestinal endoscopes
    King-Wah Chiu
    Ming-Chao Tsai
    Keng-Liang Wu
    Yi-Chun Chiu
    Ming-Tzung Lin
    Tsung-Hui Hu
    BMC Gastroenterology, 12
  • [3] Surveillance cultures of samples obtained from biopsy channels and automated endoscope reprocessors after high-level disinfection of gastrointestinal endoscopes
    Chiu, King-Wah
    Tsai, Ming-Chao
    Wu, Keng-Liang
    Chiu, Yi-Chun
    Lin, Ming-Tzung
    Hu, Tsung-Hui
    BMC GASTROENTEROLOGY, 2012, 12
  • [4] Surveillance Culture of Endoscope to Monitor the Quality of High-Level Disinfection of Gastrointestinal Reprocessing
    Chiu, King-Wah
    Fong, Tze-Vun
    Wu, Keng-Liang
    Chiu, Yi-Chun
    Chou, Yeh-Pin
    Kuo, Chung-Mou
    Chuah, Seng-Kee
    Kuo, Chung-Huang
    Chiou, Shue-Shien
    ChangChien, Chi-Sin
    HEPATO-GASTROENTEROLOGY, 2010, 57 (99-100) : 531 - 534
  • [5] Automated High-Level Disinfection of Nonchanneled Flexible Endoscopes: Duty Cycles and Endoscope Repair
    Statham, Melissa McCarty
    Willging, J. Paul
    LARYNGOSCOPE, 2010, 120 (10) : 1946 - 1949
  • [6] High-level endoscope disinfection processes in emerging economies: financial impact of manual process versus automated endoscope reprocessing
    Funk, S. E.
    Reaven, N. L.
    JOURNAL OF HOSPITAL INFECTION, 2014, 86 (04) : 250 - 254
  • [7] Recent Update on Microbiological Monitoring of Gastrointestinal Endoscopes after High-Level Disinfection
    Shin, Suk Pyo
    Kim, Won Hee
    CLINICAL ENDOSCOPY, 2015, 48 (05) : 369 - 373
  • [8] Surveillance culture monitoring of double-balloon enteroscopy reprocessing with high-level disinfection
    Chiu, King-Wah
    Lu, Long-Sheng
    Wu, Keng-Liang
    Lin, Ming-Tzung
    Hu, Ming-Luen
    Tai, Wei-Chen
    Chiu, Yi-Chun
    Chuah, Seng-Kee
    Hu, Tsung-Hui
    EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 2012, 42 (04) : 427 - 431
  • [9] Evaluation of an automated high-level disinfection technology for ultrasound transducers
    Vickery, Karen
    Gorgis, Vivian Zaiya
    Burdach, Jon
    Patel, Dipika
    JOURNAL OF INFECTION AND PUBLIC HEALTH, 2014, 7 (02) : 153 - 160
  • [10] Improper positioning of the elevator lever of duodenoscopes may lead to sequestered bacteria that survive disinfection by automated endoscope reprocessors
    Alfa, Michelle J.
    Singh, Harminder
    Duerksen, Donald R.
    Schultz, Gale
    Reidy, Carol
    DeGagne, Pat
    Olson, Nancy
    AMERICAN JOURNAL OF INFECTION CONTROL, 2018, 46 (01) : 73 - 75