Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy

被引:0
|
作者
Rong-Cheng Han [1 ,2 ]
Yong-Jiang Li [3 ]
Yue-Hu Pu [3 ,4 ]
机构
[1] Shanghai Institute of Applied Physics, Chinese Academy of Sciences
[2] University of Chinese Academy of Sciences
关键词
Collection efficiency; Monitor ionization chamber; Pencil beam scanning; Proton therapy;
D O I
暂无
中图分类号
R730.5 [肿瘤治疗学];
学科分类号
100214 ;
摘要
The collection efficiency of monitor parallel plate ionization chambers is the main uncertainty in the beam control of pencil beam scanning systems. Existing calculation methods for collection efficiency in photon or passive scattering proton systems have not considered the characteristics of non-uniform charge density in pencil beam scanning systems. In this study, Boag’s theory was applied to a proton pencil beam scanning system. The transverse distribution of charge density in the ionization chamber was considered to be a Gaussian function and an analytical solution was derived to calculate collection efficiency in the beam spot area. This calculation method is called the integral method and it was used to investigate the effects of beam parameters on collection efficiency. It was determined that collection efficiency is positively correlated with applied voltage, beam size, and beam energy,but negatively correlated with beam current intensity.Additionally, it was confirmed that collection efficiency is improved when the air filling the monitor parallel plate ionization chamber is replaced with nitrogen.
引用
收藏
页码:3 / 12
页数:10
相关论文
共 50 条
  • [1] Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy
    Rong-Cheng Han
    Yong-Jiang Li
    Yue-Hu Pu
    Nuclear Science and Techniques, 2020, 31
  • [2] Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy
    Han, Rong-Cheng
    Li, Yong-Jiang
    Pu, Yue-Hu
    NUCLEAR SCIENCE AND TECHNIQUES, 2020, 31 (02)
  • [3] Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance
    Lin, Liyong
    Kang, Minglei
    Solberg, Timothy D.
    Mertens, Thierry
    Baumer, Christian
    Ainsley, Christopher G.
    McDonough, James E.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 270 - 276
  • [4] A 2D strip ionization chamber array with high spatiotemporal resolution for proton pencil beam scanning FLASH radiotherapy
    Yang, Yunjie
    Shi, Chengyu
    Chen, Chin-Cheng
    Tsai, Pingfang
    Kang, Minglei
    Huang, Sheng
    Lin, Chih-Hsun
    Chang, Fu-Xiong
    Chhabra, Arpit M.
    Choi, J. Isabelle
    Tome, Wolfgang A.
    Simone, Charles B.
    Lin, Haibo
    MEDICAL PHYSICS, 2022, 49 (08) : 5464 - 5475
  • [5] Contour scanning for penumbra improvement in pencil beam scanned proton therapy
    Meier, G.
    Leiser, D.
    Besson, R.
    Mayor, A.
    Safai, S.
    Weber, D. C.
    Lomax, A. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (06) : 2398 - 2416
  • [6] Investigating beam matching for multi-room pencil beam scanning proton therapy
    Suresh Rana
    Jaafar Bennouna
    Physical and Engineering Sciences in Medicine, 2020, 43 : 1241 - 1251
  • [7] Proton beam monitor chamber calibration
    Goma, C.
    Lorentini, S.
    Meer, D.
    Safai, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (17) : 4961 - 4971
  • [8] Investigating beam matching for multi-room pencil beam scanning proton therapy
    Rana, Suresh
    Bennouna, Jaafar
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (04) : 1241 - 1251
  • [9] Increased Efficiency of Range Verification in Routine QA for Pencil-Beam Scanning Proton Therapy
    Jeong, Seonghoon
    Yoon, Myonggeun
    Son, Jaeman
    Chung, Kwangzoo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 73 (07) : 983 - 989
  • [10] Increased Efficiency of Range Verification in Routine QA for Pencil-Beam Scanning Proton Therapy
    Seonghoon Jeong
    Myonggeun Yoon
    Jaeman Son
    Kwangzoo Chung
    Journal of the Korean Physical Society, 2018, 73 : 983 - 989