Dual Energy CT Imaging in Cone-Beam Micro-CT for Improved Attenuation Coefficient Measurement

被引:3
作者
唐智伟 [1 ]
胡广书 [1 ]
机构
[1] Department of Biomedical Engineering, Tsinghua University
关键词
micro-CT; cone-beam; linear attenuation coefficient; beam hardening; dual energy;
D O I
暂无
中图分类号
TH774 [医用放射线设备];
学科分类号
1004 ;
摘要
In order to improve micro-CT’s capability of accurate quantification of linear attenuation coefficient μ, a dual energy method was developed to correct beam hardening artifacts caused by the polychromatic spectra of X-ray tubes. In this method, two sets of scans, taken at different energy levels, were combined to create a synthetic monochromatic image. A physical polychromatic model of μ in dual energy imaging was developed with an iterative method to solve the model for a few selected pixels. To find a high-speed and effective computing approach, the physics model was approximated by a polynomial function of the measured intensities. The method was tested on a PMMA-aluminum phantom and CaCl2 admixtures. The results show that streak and cupping artifacts are completely eliminated and that the measurement of the reconstructed attenuation coefficient μ is observed to be over 95% accurate.
引用
收藏
页码:352 / 357
页数:6
相关论文
共 3 条
  • [1] A microtomographic system for the nondestructive evaluation of bone architecture
    Ruegsegger, P
    Koller, B
    Muller, R
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1996, 58 (01) : 24 - 29
  • [2] A frameworkfor spectral artifact corrections in X-ray CT .2 J.Peter Stonestrom,R.Alvarez,A.Macovski. IEEETransactions on Biomedical Engineering . 1981
  • [3] Short-Scan Cone Beam Algorithm for Circular and Non-Circular Detector Orbits .2 Zeng,G. L,and G. T. Gullberg. SPIE Medical Imaging IV: Image Processing . 1990