Development, validity and reliability of a new pressure air biofeedback device (PAB) for measuring isometric extension strength of the lumbar spine

被引:5
作者
Pienaar A.W. [1 ]
Barnard J.G. [1 ]
机构
[1] Biokinetics Laboratory, Department of Sport Science, Stellenbosch University, Stellenbosch
关键词
air pressure; isometric strength; low back pain (LBP); PAB[!sup]®[!/sup; reliability;
D O I
10.1080/03091902.2016.1253794
中图分类号
学科分类号
摘要
This study describes the development of a new portable muscle testing device, using air pressure as a biofeedback and strength testing tool. For this purpose, a pressure air biofeedback device (PAB®) was developed to measure and record the isometric extension strength of the lumbar multifidus muscle in asymptomatic and low back pain (LBP) persons. A total of 42 subjects (age 47.58 years, ±18.58) participated in this study. The validity of PAB® was assessed by comparing a selected measure, air pressure force in millibar (mb), to a standard criterion; calibrated weights in kilograms (kg) during day-to-day tests. Furthermore, clinical trial-to-trial and day-to-day tests of maximum voluntary isometric contraction (MVIC) of L5 lumbar multifidus were done to compare air pressure force (mb) to electromyography (EMG) in microvolt (μV) and to measure the reliability of PAB®. A highly significant relationship were found between air pressure output (mb) and calibrated weights (kg). In addition, Pearson correlation calculations showed a significant relationship between PAB® force (mb) and EMG activity (μV) for all subjects (n = 42) examined, as well as for the asymptomatic group (n = 24). No relationship was detected for the LBP group (n = 18). In terms of lumbar extension strength, we found that asymptomatic subjects were significantly stronger than LBP subjects. The results of the PAB® test differentiated between LBP and asymptomatic subject’s lumbar isometric extension strength without any risk to the subjects and also indicate that the lumbar isometric extension test with the new PAB® device is reliable and valid. © 2016 Informa UK Limited, trading as Taylor & Francis Group.
引用
收藏
页码:216 / 222
页数:6
相关论文
共 30 条
[1]  
Henn L., Schier K., Brian T., Et al., Back pain in Poland and Germany: a survey of prevalence and association with demographic characters, Biomed Research Int, 2014, pp. 1-5, (2014)
[2]  
Manchikanti L., Singh V., Falco F.J.E., Et al., Epidemiology of low back pain in adults, Neuromodulation, 17, pp. 3-10, (2014)
[3]  
McGill S.M., Building better rehabilitation programs for low back injuries, Low back disorders: evidence-based prevention and rehabilitation, pp. 166-188, (2007)
[4]  
Richardson C., Hides J., Open chain segmental control and progression into function, Therapeutic exercise for lumbopelvic stabilization. A motor control approach for the treatment and prevention of low back pain, pp. 233-242, (2005)
[5]  
Nobori H., Maruyama H., Invention and making of “Power Measure”, a simple instrument to measure muscular strength–validation of the measurement values, J Physical Therapy Sci, 19, pp. 9-13, (2007)
[6]  
Axen K., Haas F., Schicchi J., Et al., Progressive resistance neck exercises using a compressible ball coupled with an air pressure gauge, J Orthopaedic Sports Physical Therapy, 16, pp. 1-9, (1992)
[7]  
Helewa A., Goldsmith C.H., Smythe H.A., The modified sphygmomanometer–an instrument to measure muscle strength: a validation study, J Chronic Dis, 34, pp. 353-361, (1981)
[8]  
Storheim K., Bo K., Pederstad O., Et al., Intra-tester reproducibility of pressure biofeedback in measurement of transverses abdominis function, Physiotherapy Res Int, 7, pp. 239-249, (2002)
[9]  
Li R.C., Jasiewicz J.M., Middleton J., Et al., The development, validity, and reliability of a manual muscle testing device with integrated limb position sensors, Arch Physical Med Rehabil, 87, pp. 411-417, (2006)
[10]  
Verkhoshansky Y., Siff M.C., The methods of special strength training, Supertraining, Sixth Edition–Expanded Version, pp. 255-312, (2009)