Geometric calibration and correction for a lens-coupled detector in x-ray phase-contrast imaging

被引:1
作者
George A. [1 ]
Chen P.Y. [1 ]
Morales-Martinez A. [1 ]
Panna A. [1 ]
Gomella A.A. [1 ]
Bennett E.E. [1 ]
Wen H. [1 ]
机构
[1] National Institutes of Health, National Heart, Lung and Blood Institute, Biochemistry and Biophysics Center, Imaging Physics Laboratory, Bethesda, MD
关键词
geometric correction; image distortion; phase contrast; tilted phosphor; x-ray;
D O I
10.1117/1.JMI.4.1.013507
中图分类号
学科分类号
摘要
A lens-coupled x-ray camera with a tilted phosphor collects light emission from the x-ray illuminated (front) side of phosphor. Experimentally, it has been shown to double x-ray photon capture efficiency and triple the spatial resolution along the phosphor tilt direction relative to the same detector at normal phosphor incidence. These characteristics benefit grating-based phase-contrast methods, where linear interference fringes need to be clearly resolved. However, both the shallow incident angle on the phosphor and lens aberrations of the camera cause geometric distortions. When tiling multiple images of limited vertical view into a full-field image, geometric distortion causes blurring due to image misregistration. Here, we report a procedure of geometric correction based on global polynomial transformation of image coordinates. The corrected image is equivalent to one obtained with a single full-field flat panel detector placed at the sample plane. In a separate evaluation scan, the position deviations in the horizontal and vertical directions were reduced from 0.76 and 0.028 mm, respectively, to 0.006 and 0.009 mm, respectively, by the correction procedure, which were below the 0.028-mm pixel size of the imaging system. In a demonstration of a phase-contrast imaging experiment, the correction reduced blurring of small structures. © 2017 Society of Photo-Optical Instrumentation Engineers (SPIE).
引用
收藏
相关论文
共 24 条
[1]  
Boone J.M., Et al., Analysis and correction of imperfections in the image intensifier-Tv-digitizer imaging chain, Med. Phys., 18, 2, pp. 236-242, (1991)
[2]  
Fahrig R., Moreau M., Holdsworth D.W., Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Correction of image intensifier distortion, Med. Phys., 24, 7, pp. 1097-1106, (1997)
[3]  
Gronenschild E., The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain, Med. Phys., 24, 12, pp. 1875-1888, (1997)
[4]  
Gronenschild E., Correction for geometric image distortion in the x-ray imaging chain: Local technique versus global technique, Med. Phys., 26, 12, pp. 2602-2616, (1999)
[5]  
Liu R.R., Rudin S., Bednarek D.R., Super-global distortion correction for a rotational C-arm x-ray image intensifier, Med. Phys., 26, 9, pp. 1802-1810, (1999)
[6]  
Liu H., Et al., Lens distortion in optically coupled digital x-ray imaging, Med. Phys., 27, 5, pp. 906-912, (2000)
[7]  
Cerveri P., Et al., Distortion correction for x-ray image intensifiers: Local unwarping polynomials and RBF neural networks, Med. Phys., 29, 8, pp. 1759-1771, (2002)
[8]  
Cerveri P., Et al., Hierarchical radial basis function networks and local polynomial un-warping for x-ray image intensifier distortion correction: A comparison with global techniques, Med. Biol. Eng. Comput., 41, 2, pp. 151-163, (2003)
[9]  
Chung H.T., Kim D.G., Distortion correction for digital subtraction angiography imaging: PC based system for radiosurgery planning, Comput. Meth. Progr. Biomed., 71, 2, pp. 165-173, (2003)
[10]  
Fantozzi S., Cappello A., Leardini A., A global method based on thin-plate splines for correction of geometric distortion: An application to fluoroscopic images, Med. Phys., 30, 2, pp. 124-131, (2003)