Comparison of accuracy, efficiency and safety between robotic-assisted and non-robotic-assisted deep brain stimulation: systematic review and/or meta-analysis

被引:0
作者
Wan, Junge [1 ]
Jin, Zhizhong [2 ]
Wang, Yongfeng [3 ]
Han, Shuai [1 ]
机构
[1] China Med Univ, Dept Neurosurg, Shengjing Hosp, Sanhao St 36, Shenyang 110055, Peoples R China
[2] China Med Univ, Hosp 1, Dept Neurosurg, Shenyang, Peoples R China
[3] China Med Univ, Hosp 1, Dept Radiol, 155 North Nanjing St, Shenyang 110001, Peoples R China
关键词
accuracy; deep brain stimulation; efficiency; meta-analysis; robotic-assisted; safety; SUBTHALAMIC NUCLEUS; SURGERY; LOCALIZATION; IMPLANTATION; INFECTIONS; ELECTRODES; CHECKLIST; ERROR; RISK;
D O I
10.1097/JS9.0000000000002436
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective:This meta-analysis aims to compare robotic-assisted deep brain stimulation (RA-DBS) and non-robotic-assisted deep brain stimulation (nRA-DBS) regarding accuracy, efficiency and safety.Methods:We searched six databases to retrieve relevant studies. Two independent reviewers selected the studies and assessed the risk of bias using the Cochrane risk-of-bias tool for randomized trials version 2 and the Methodological index for nonrandomized studies score. Statistical analysis was completed by Revman 5.4.Results:A total of seven trials with 341 participants entered our analysis. Our meta-analysis showed that RA-DBS demonstrated a statistically significant reduction in target point error (MD: -0.30, 95%CI: [-0.58, -0.02], I2 = 0, P = 0.03) and deviation outliers compared to nRA-DBS. (OR: 0.15, 95%CI: [0.04, 0.51], I2 = 0, P = 0.002). RA-DBS and nRA-DBS demonstrated comparable efficiency metrics in terms of operation room time (MD: 19.37, 95%CI: [-62.85,102.59], I2 = 99%, P = 0.65), operating time (MD: -17.04, 95%CI: [-84.95, 50.87], I2 = 98%, P = 0.62) and total anesthesia time (MD: 14.24, 95%CI: [-96.26, -124.73], I2 = 97%, P = 0.80). Two groups were comparable in terms of complication rates (OR: 1.79, 95%CI: [0.79, 4.05], I2 = 5%, P = 0.17) and intracranial hemorrhage rates (OR: 0.80, 95%CI: [0.23, 2.74], I2 = 0, P = 0.72).Conclusions:RA-DBS exhibits efficiency and safety comparable to nRA-DBS, serving as a viable alternative to nRA-DBS. Although RA-DBS shows promise in accuracy, further high-quality studies are needed to establish its clinical superiority.
引用
收藏
页码:4704 / 4714
页数:11
相关论文
共 54 条
[1]   Effective accuracy of stereoelectroencephalography: robotic 3D versus Talairach orthogonal approaches [J].
Bourdillon, Pierre ;
Chatillon, Claude-Edouard ;
Moles, Alexis ;
Rheims, Sylvain ;
Catenoix, Helene ;
Montavont, Alexandra ;
Ostrowsky-Coste, Karine ;
Boulogne, Sebastien ;
Isnard, Jean ;
Guenot, Marc .
JOURNAL OF NEUROSURGERY, 2019, 131 (06) :1938-1946
[2]   Prolonged Operative Duration Increases Risk of Surgical Site Infections: A Systematic Review [J].
Cheng, Hang ;
Chen, Brian Po-Han ;
Soleas, Ireena M. ;
Ferko, Nicole C. ;
Cameron, Chris G. ;
Hinoul, Piet .
SURGICAL INFECTIONS, 2017, 18 (06) :722-735
[3]   Deep brain stimulation for movement disorders [J].
Collins, Kelly L. ;
Lehmann, Emily M. ;
Patil, Parag G. .
NEUROBIOLOGY OF DISEASE, 2010, 38 (03) :338-345
[4]   Feasibility of an Operational Standardized Checklist for Movement Disorder Surgery A Pilot Study [J].
Connolly, Patrick J. ;
Kilpatrick, Michaux ;
Jaggi, Jurg L. ;
Church, Ephraim ;
Baltuch, Gordon H. .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2009, 87 (02) :94-100
[5]   REOPERATION FOR SUBOPTIMAL OUTCOMES AFTER DEEP BRAIN STIMULATION SURGERY [J].
Ellis, Tina-Marie ;
Foote, Kelly D. ;
Fernandez, Hubert H. ;
Sudhyadhom, Atchar ;
Rodriguez, Ramon L. ;
Zeilman, Pamela ;
Jacobson, Charles E., IV ;
Okun, Michael S. .
NEUROSURGERY, 2008, 63 (04) :754-760
[6]   Cranial neurosurgical robotics [J].
Elsabeh, Rami ;
Singh, Sukhbir ;
Shasho, Jeff ;
Saltzman, Yoni ;
Abrahams, John M. .
BRITISH JOURNAL OF NEUROSURGERY, 2021, 35 (05) :532-540
[7]   Frameless Robot-Assisted vs Frame-Based Awake Deep Brain Stimulation Surgery: An Evaluation of Technique and New Challenges [J].
Fenoy, Albert J. ;
Conner, Christopher R. .
OPERATIVE NEUROSURGERY, 2022, 22 (03) :171-178
[8]   Robotic Stereotaxy in Cranial Neurosurgery: A Qualitative Systematic Review [J].
Fomenko, Anton ;
Serletis, Demitre .
NEUROSURGERY, 2018, 83 (04) :642-650
[9]   Accuracy of the robot-assisted procedure in deep brain stimulation [J].
Goia, Alice ;
Gilard, Vianney ;
Lefaucheur, Romain ;
Welter, Marie-Laure ;
Maltete, David ;
Derrey, Stephane .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2019, 15 (06)
[10]   Robot-assisted vs. manually guided stereoelectroencephalography for refractory epilepsy: a systematic review and meta-analysis [J].
Gomes, Fernando Cotrim ;
Larcipretti, Anna Laura Lima ;
Nager, Gabriela ;
Dagostin, Caroline Serafim ;
Udoma-Udofa, Ofonime Chantal ;
Pontes, Julia Pereira Muniz ;
de Oliveira, Jessica Sales ;
de Souza, Justine Hellen Cavalcanti ;
Bannach, Matheus de Andrade .
NEUROSURGICAL REVIEW, 2023, 46 (01)