Comparison of Complex Open Domain Electrical Impedance Tomography Methods

被引:0
作者
Doussan, Allaire F. [3 ]
Murphy, Ethan K. [1 ]
Lloyd, Sophie A. [1 ]
Halter, Ryan J. [1 ,2 ]
机构
[1] Dartmouth Coll USA, Thayer Sch Engn, Hanover, NH USA
[2] Dartmouth Coll USA, Geisel Sch Med, Hanover, NH USA
[3] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
Electrical impedance tomography; Electrodes; Impedance; Image reconstruction; Probes; Muscles; Cancer; Voltage measurement; Surgery; Accuracy; inverse problem; solid tumor cancer; surgical margin assessment; POSITIVE SURGICAL MARGINS; BREAST-CONSERVING SURGERY; RADICAL PROSTATECTOMY; CLINICAL-SIGNIFICANCE; REAL-TIME; IMPACT; CANCER; TISSUE; TUMOR; RECONSTRUCTIONS;
D O I
10.1109/TBME.2025.3547239
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: This study evaluates the potential of complex 3D Electrical Impedance Tomography (EIT) for intraoperative surgical margin assessment (SMA) using an ex vivo bovine model. Methods: A custom electrode array was used to collect impedance data across multiple frequencies (100 Hz-1 MHz) from 57 tissue samples. An optimal pressure range was identified to ensure proper electrode contact with minimal tissue deformation (2 kPa to 5 kPa) and a saline calibration was used to minimize model-data mismatch. A novel best fit factor, beta, was introduced to scale the reference data of difference EIT and initial guess for absolute EIT to eliminate inversions in permittivity. The accuracy of complex EIT reconstructions were investigated using pixel-based tissue classification. Results: Experimental difference EIT, particularly for conductivity (sigma) at 100 Hz, achieved high sensitivity (0.92) and specificity (0.90) with fast reconstruction speeds of 1.67 seconds. While absolute EIT performed slightly better in quantitative metrics for sigma imaging (AUC = 0.90, Accuracy = 0.86), difference EIT was found to be more suitable for real-time applications due to its faster processing time. Simulations suggested that permittivity (& varepsilon;) has sufficient contrast for classifying muscle and adipose tissue. However, experimental & varepsilon; reconstructions exhibited lower performance, suggesting the need for hardware improvements. Conclusion: Difference EIT is the most promising method for real-time applications, balancing high accuracy with reconstruction speeds. Absolute EIT, while slightly more accurate, is less feasible due to longer processing times (150.6 seconds). Significance: This study advances the potential for real-time intraoperative SMA using a novel complex EIT reconstruction method.
引用
收藏
页码:2555 / 2566
页数:12
相关论文
共 53 条
[1]   A prospective, randomized, controlled, multicenter study of a real-time, intraoperative probe for positive margin detection in breast-conserving surgery [J].
Allweis, Tanir M. ;
Kaufman, Zvi ;
Lelcuk, Shlomo ;
Pappo, Itzhak ;
Karni, Tami ;
Schneebaum, Shlomo ;
Spector, Rona ;
Schindel, Asher ;
Hershko, Dan ;
Zilberman, Moshe ;
Sayfan, Joel ;
Berlin, Yuri ;
Hadary, Amram ;
Olsha, Oded ;
Paran, Haim ;
Gutman, Mordechai ;
Carmon, Moshe .
AMERICAN JOURNAL OF SURGERY, 2008, 196 (04) :483-489
[2]   The clinical significance of the positive surgical margin in oral cancer [J].
Binahmed, Abdulaziz ;
Nason, Richard W. ;
Abdoh, Ahmed A. .
ORAL ONCOLOGY, 2007, 43 (08) :780-784
[3]   3D Electric Impedance Tomography Reconstruction on Multi-Core Computing Platforms [J].
Borsic, Andrea ;
Hartov, Alexander ;
Paulsen, Keith D. ;
Manwaring, Preston .
2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vols 1-8, 2008, :1175-1177
[4]   THE CLINICAL-SIGNIFICANCE OF PATHOLOGICAL FINDINGS IN SURGICALLY RESECTED MARGINS OF THE PRIMARY TUMOR IN HEAD AND NECK-CARCINOMA [J].
CHEN, TY ;
EMRICH, LJ ;
DRISCOLL, DL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (06) :833-837
[5]   Margin control in open radical prostatectomy: What are the real outcomes? [J].
Cookson, Michael S. ;
Chang, Sam S. .
UROLOGIC ONCOLOGY-SEMINARS AND ORIGINAL INVESTIGATIONS, 2010, 28 (02) :205-209
[6]   Classification of breast tissue by electrical impedance spectroscopy [J].
da Silva, JE ;
de Sá, JPM ;
Jossinet, J .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2000, 38 (01) :26-30
[7]   Intra-operative assessment of excised breast tumour margins using ClearEdge imaging device [J].
Dixon, J. M. ;
Renshaw, L. ;
Young, O. ;
Kulkarni, D. ;
Saleem, T. ;
Sarfaty, M. ;
Sreenivasan, R. ;
Kusnick, C. ;
Thomas, J. ;
Williams, L. J. .
EJSO, 2016, 42 (12) :1834-1840
[8]   Bioimpedance of soft tissue under compression [J].
Dodde, R. E. ;
Bull, J. L. ;
Shih, A. J. .
PHYSIOLOGICAL MEASUREMENT, 2012, 33 (06) :1095-1109
[9]   Towards intraoperative surgical margin assessment: Validation of an electrical impedance-based probe with ex vivo bovine tissue [J].
Doussan, Allaire F. ;
Lloyd, Sophie ;
Murphy, Ethan K. ;
Halter, Ryan J. .
2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
[10]   Analysis of resting noise characteristics of three EIT systems in order to compare suitability for time difference imaging with scalp electrodes during epileptic seizures [J].
Fabrizi, L. ;
McEwan, A. ;
Woo, E. ;
Holder, D. S. .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (07) :S217-S236