Improvement of Anatomic Alignment and Image Quality Using a Respiratory Motion Reduction Block in Oncologic PET/CT

被引:0
作者
Ichikawa, Hajime [1 ]
Shibutani, Takayuki [2 ]
Kato, Toyohiro [1 ]
Banno, Tomoya [1 ]
Terabe, Mitsuaki [1 ]
Shimada, Hideki [1 ]
机构
[1] Toyohashi Municipal Hosp, Dept Radiol, Toyohashi, Japan
[2] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Dept Quantum Med Technol, Kanazawa, Japan
关键词
PET; respiratory gating; respiratory motion reduction; anatomic alignment; BREATH-HOLD PET/CT; CT; PROTOCOL;
D O I
10.2967/jnmt.124.269096
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
18F-FDG PET/CT is crucial for cancer diagnosis; however, respiratory motion often causes misregistration between PET and CT images. This study aimed to evaluate the impact of a custommade respiratory motion reduction block (RRB) in reducing misregistration and improving image quality in 18F-FDG PET/CT. Methods: The RRB was developed to minimize the effects of respiratory motion. It is a pentagonal block made of Styrofoam designed to compress the upper abdomen. This study included 170 patients who underwent whole-body 18F-FDG PET/CT. Patients were categorized into 4 groups based on the CT and PET scanning technique used: a control free-breathing (FB) group (breath-holding [BH] CT without RRB and respiration-gated [RG] PET), a control RG group (BH CT without RRB with RG PET), a BH RRB group (BH CT with RRB), and a FB RRB group (FB CT with RRB). Various parameters, including the incidence rate and distance of anatomic misregistration, and signal-to-noise ratio, were measured and compared. Results: For the control FB, control RG, BH RRB, and FB RRB groups, the incidence rates of anatomic misregistration between PET and CT images were 47.1%, 27.5%, 34.0%, and 16.7%, respectively. The mean misregistration distances in these groups were 5.1 +/- 6.0, 3.1 +/- 5.5, 3.8 +/- 6.2, and 1.3 +/- 3.2 mm, respectively. Furthermore, the RRB improved the signal-to-noise ratio of the liver. Conclusion: The RRB effectively improved registration between PET and CT images. Our approach is cost-effective and provides respiratory motion suppression not only during PET but also during CT scans. The use of RRB with FB PET scanning was found to be more beneficial than respiratory gating.
引用
收藏
页码:146 / 151
页数:6
相关论文
共 28 条
[1]   Improvement in PET/CT Image Quality with a Combination of Point-Spread Function and Time-of-Flight in Relation to Reconstruction Parameters [J].
Akamatsu, Go ;
Ishikawa, Kaori ;
Mitsumoto, Katsuhiko ;
Taniguchi, Takafumi ;
Ohya, Nobuyoshi ;
Baba, Shingo ;
Abe, Koichiro ;
Sasaki, Masayuki .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (11) :1716-1722
[2]   FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0 [J].
Boellaard, Ronald ;
Delgado-Bolton, Roberto ;
Oyen, Wim J. G. ;
Giammarile, Francesco ;
Tatsch, Klaus ;
Eschner, Wolfgang ;
Verzijlbergen, Fred J. ;
Barrington, Sally F. ;
Pike, Lucy C. ;
Weber, Wolfgang A. ;
Stroobants, Sigrid ;
Delbeke, Dominique ;
Donohoe, Kevin J. ;
Holbrook, Scott ;
Graham, Michael M. ;
Testanera, Giorgio ;
Hoekstra, Otto S. ;
Zijlstra, Josee ;
Visser, Eric ;
Hoekstra, Corneline J. ;
Pruim, Jan ;
Willemsen, Antoon ;
Arends, Bertjan ;
Kotzerke, Joerg ;
Bockisch, Andreas ;
Beyer, Thomas ;
Chiti, Arturo ;
Krause, Bernd J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (02) :328-354
[3]   Imaging and dosimetric errors in 4D PET/CT-guided radiotherapy from patient-specific respiratory patterns: a dynamic motion phantom end-to-end study [J].
Bowen, S. R. ;
Nyflot, M. J. ;
Herrmann, C. ;
Groh, C. M. ;
Meyer, J. ;
Wollenweber, S. D. ;
Stearns, C. W. ;
Kinahan, P. E. ;
Sandison, G. A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (09) :3731-3746
[4]   Implementation of an Automated Respiratory Amplitude Gating Technique for PET/CT: Clinical Evaluation [J].
Chang, Guoping ;
Chang, Tingting ;
Pan, Tinsu ;
Clark, John W., Jr. ;
Mawlawi, Osama R. .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (01) :16-24
[5]   Using Cine-Averaged CT With the Shallow Breathing Pattern to Reduce Respiration-Induced Artifacts for Thoracic Cavity PET/CT Scans [J].
Changlai, Sheng-Pin ;
Huang, Cheng-Kai ;
Luzhbin, Dmytro ;
Lin, Fang-Yi ;
Wu, Jay .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2019, 213 (01) :140-146
[6]   Multi-bed-position acquisition technique for deep inspiration breath-hold PET/CT: a preliminary result for pulmonary lesions [J].
Daisaki, Hiromitsu ;
Shinohara, Hiroyuki ;
Terauchi, Takashi ;
Murano, Takeshi ;
Shimada, Naoki ;
Moriyama, Noriyuki ;
Tateishi, Ukihide .
ANNALS OF NUCLEAR MEDICINE, 2010, 24 (03) :179-188
[7]   Clinical evaluation of a breathing protocol for PET/CT [J].
de Juan, R ;
Seifert, B ;
Berthold, T ;
von Schulthess, GK ;
Goerres, GW .
EUROPEAN RADIOLOGY, 2004, 14 (06) :1118-1123
[8]   Clinical feasibility and impact of data-driven respiratory motion compensation studied in 200 whole-body 18F-FDG PET/CT scans [J].
Dias, Andre H. ;
Schleyer, Paul ;
Vendelbo, Mikkel H. ;
Hjorthaug, Karin ;
Gormsen, Lars C. ;
Munk, Ole L. .
EJNMMI RESEARCH, 2022, 12 (01)
[9]   Initial clinical results for breath-hold CT-based processing of respiratory-gated PET acquisitions [J].
Fin, Loic ;
Daouk, Joeel ;
Morvan, Julie ;
Bailly, Pascal ;
El Esper, Isabelle ;
Saidi, Lazhar ;
Meyer, Marc-Etienne .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (11) :1971-1980
[10]   Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 2.0 [J].
Fukukita, Hiroyoshi ;
Suzuki, Kazufumi ;
Matsumoto, Keiichi ;
Terauchi, Takashi ;
Daisaki, Hiromitsu ;
Ikari, Yasuhiko ;
Shimada, Naoki ;
Senda, Michio .
ANNALS OF NUCLEAR MEDICINE, 2014, 28 (07) :693-705