Numerical modeling and parametric study of a hollow fiber dialyzer using double porous media approach

被引:4
作者
Goodarzi, Mohammad Reza [1 ]
Mohebbi-Kalhori, Davod [1 ]
机构
[1] Univ Sistan & Baluchestan, Fac Engn, Dept Chem Engn, POB 98164161, Zahedan, Iran
关键词
Hollow fiber dialyzer; Membrane; Modeling of hemodialyzer; Double porous media; Clearance; SOLUTE TRANSPORT; MASS-TRANSFER; PERFORMANCE; FLOW; CLEARANCE; EQUATIONS; VOLUME;
D O I
10.1016/j.sajce.2022.07.001
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Mathematical modeling is one of the efficient methods to improve the design of hemodialyzer and study affecting parameters on dialyzer performance. In this paper, mathematical modeling of hollow fiber dialyzer is done to investigate the effect of various parameters on the clearance of urea and creatinine. The computational domain is considered two-dimensional axisymmetry. The hollow fibers bundle is defined as a double porous media. The Darcy-BrinkmanForchheimer equation is employed to determine flows in porous media. Modified Kedem-Katchalsky equations are used to calculate the transmembrane flow. Continuity, Navier-Stokes and concentration equations are solved by finite element method. The numerical results showed a change in the thickness of hollow fibers from 100 to 25 mu m and a decrease in their inner diameter from 240 to 150 mu m, could improve clearance of urea by 66.7 and 20.7 percent respectively. The increase in blood flow (40 to 120 mL.min-1), dialysate flow (80 to 200 mL.min-1), dialyzer diameter (28 to 38 mm) and length of hollow fibers (15 to 30 cm) increase the clearance of urea by 88.2, 20.4, 35 and 26.2 percent, respectively. The model has been validated and shows good agreement with experimental results. Such studies advance our understanding of the parameters affect the performance of the dialyzer.
引用
收藏
页码:1 / 11
页数:11
相关论文
共 30 条
[1]   SOLUTE TRANSPORT CHARACTERISTICS IN HEMODIAFILTRATION [J].
CHANG, YL ;
LEE, CJ .
JOURNAL OF MEMBRANE SCIENCE, 1988, 39 (02) :99-111
[2]   Enhanced clearance of highly protein-bound drugs by albumin-supplemented dialysate during modeled continuous hemodialysis [J].
Churchwell, Mariann D. ;
Pasko, Deborah A. ;
Smoyer, William E. ;
Mueller, Bruce A. .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2009, 24 (01) :231-238
[3]   Double porous media model for mass transfer of hemodialyzers [J].
Ding, WP ;
He, LQ ;
Gang, Z ;
Zhang, HF ;
Shu, ZQ ;
Gao, DY .
INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2004, 47 (22) :4849-4855
[4]   Computational flow modeling in hollow-fiber dialyzers [J].
Eloot, S ;
De Wachter, D ;
Van Tricht, I ;
Verdonck, P .
ARTIFICIAL ORGANS, 2002, 26 (07) :590-599
[5]  
Eloott S., 2006, Computer Methods in Biomechanics and Biomedical Engineering, V9, P363, DOI 10.1080/10255840601002728
[6]   Visualization of concentration fields in hemodialyzers by computed tomography [J].
Frank, A ;
Lipscomb, GG ;
Dennis, M .
JOURNAL OF MEMBRANE SCIENCE, 2000, 175 (02) :239-251
[7]   Numerical investigation of UF membrane to reduce energy consumption using double porosity approach [J].
Ghotbi, S. ;
Pirzadeh, B. ;
Mohebbi-Kalhori, Davod ;
Abdollahi, A. .
WATER SCIENCE AND TECHNOLOGY, 2018, 77 (12) :2907-2916
[8]   Non-Newtonian blood flow in human right coronary arteries: steady state simulations [J].
Johnston, BM ;
Johnston, PR ;
Corney, S ;
Kilpatrick, D .
JOURNAL OF BIOMECHANICS, 2004, 37 (05) :709-720
[9]   Dialyzer fiber bundle volume during Hemodialysis using large-surface dialyzers [J].
Leblanc, Martine ;
Lafrance, Jean-Philippe ;
Renald, Alexandra .
ASAIO JOURNAL, 2006, 52 (06) :646-651
[10]   A theoretical model to predict the in vitro performance of hemodiafilters [J].
Legallais, C ;
Catapano, G ;
von Harten, B ;
Baurmeister, U .
JOURNAL OF MEMBRANE SCIENCE, 2000, 168 (1-2) :3-15