Final report on the EURAMET key comparison EURAMET AUV.V-K5

被引:1
作者
Winther, J. H.
Bruns, Th
Schiavi, A.
Bilgic, E.
Prasil, Milan
Hynek, Stanislav
Siejda, Z.
Ruiz, S.
Canu, A.
Alsubaey, F. N.
Hof, C.
Andersson, H.
机构
[1] BKSV-DPLA, Skodborgvej 307, Nærum
[2] Ptb Ag, 1.71 z.Hd. Bundesallee 100, Braunschweig
[3] Mechanics - Stalt, Str. delle Cacce 91, Torino
[4] Tübi˙tak Ume, P.O. Box: 54, Gebze-KOCAELI
[5] V Botanice 4/1504, Praha
[6] ul. Elektoralna 2, Warszawa
[7] C/Alfar 2. Tres Cantos, Madrid
[8] 29 avenue Roger Hennequin, Trappes
[9] Riyadh - Al Muhammadiyah - in front of King Saud University
[10] Federal Institute of Metrology, Metas Laboratory Acoustics and Vibration, Lindenweg 50, Bern-Wabern
[11] Håkan Andersson Brinellgatan 4, Borås
关键词
D O I
10.1088/0026-1394/60/1A/09002
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Main text The results of the fifth EURAMET key comparison in the area of primary vibration calibration of the complex sensitivity, here concerning sinusoidal acceleration, are presented. This regional key comparison followed the previous CCAUV comparison over the extended frequency range from 10 Hz to 20 kHz. Measurement results reported by 11 participants for magnitude and phase shift of complex charge sensitivity of single ended and back-to-back accelerometers were compared. A link to the CCAUV.V-K5 with the respective Degrees of Equivalence is reported. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). © 2023 BIPM & IOP Publishing Ltd.
引用
收藏
页数:2
相关论文
共 7 条
[1]  
Methods for the calibration of vibration and shock transducers Part 1: Basic concepts, (1998)
[2]  
Methods for the calibration of vibration and shock transducers Part 11: Primary vibration calibration by laser interferometry, (1999)
[3]  
Taubner A., Schlaak H, Brucke M, Bruns Th, The influence of different vibration exciter systems on high frequency primary calibration of single-ended accelerometers, Metrologia, 47, 1
[4]  
Bruns Th, Link A, Taubner A, The influence of different vibration exciter systems on high frequency primary calibration of single-ended accelerometers: II, Metrologia, 49, 1
[5]  
General requirements for the competence of testing and calibration laboratories
[6]  
Uncertainty of measurement Part 3: Guide to the expression of uncertainty in measurement, (2008)
[7]  
Propagation of distributions using a Monte Carlo method, (2008)