Biomechanical comparison of the non-locking bone plate, locking bone plate, and double-rod clamp internal fixation in a canine femoral model

被引:0
作者
Maneewan, Rutjathorn [1 ]
Chantarapanich, Nattapon [2 ]
Morimoto, Takuma [3 ]
Thitiyanaporn, Chaiyakorn [4 ]
机构
[1] Kasetsart Univ, Vet Teaching Hosp, Fac Vet Med, Bangkok 10900, Thailand
[2] Kasetsart Univ, Fac Engn Sriracha, Dept Mech Engn, Sriracha Campus, Laem Chabang 20230, Thailand
[3] Kochi Univ Technol, Dept Aerosp Engn, Kami 7828502, Japan
[4] Kasetsart Univ, Fac Vet Med, Dept Compan Anim Clin Sci, Bangkok 10900, Thailand
关键词
bone plate; canine; clamp; femoral fracture; finite element analysis; INTRAMEDULLARY ROD; OSTEOSYNTHESIS; COMPLICATIONS; FRACTURES; FEMUR;
D O I
10.14202/vetworld.2025.773-781
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
Background and Aim: Canine femoral fractures are prevalent in veterinary medicine, necessitating effective fixation methods to ensure stability and promote healing. Conventional bone plate fixation methods, including non-locking and locking plates, have inherent limitations, such as periosteal damage and mechanical failure. This study aims to evaluate the biomechanical performance of three fixation methods-non-locking bone plates, locking bone plates, and a novel double-rod clamp internal fixation system-using finite element analysis (FEA). Materials and Methods: A computed tomography-based canine femur model was created to simulate a midshaft comminuted fracture with a 20 mm gap. Three fixation configurations were modeled: A non-locking bone plate, a locking bone plate, and a double-rod clamp system. FEA was performed to assess implant stress and proximal fragment displacement under physiological axial loading. Mesh refinement and multiple loading conditions were incorporated to enhance computational accuracy. Results: The non-locking bone plate exhibited the highest implant stress (1160.22 MPa), surpassing the material yield strength and indicating a risk of mechanical failure. The double-rod clamp system demonstrated lower stress (628.34 MPa), whereas the locking bone plate had the lowest stress (446.63 MPa). Proximal fragment displacement was highest in the non-locking bone plate (2.37 mm), followed by the double-rod clamp system (0.99 mm), with the locking bone plate exhibiting the least displacement (0.34 mm), suggesting superior stability. Conclusion: The double-rod clamp system emerged as a promising alternative, offering a balance between stability and stress distribution while minimizing periosteal damage. While the locking bone plate provided the greatest stability, the double-rod clamp fixation demonstrated favorable mechanical properties and could serve as a cost-effective and minimally invasive alternative in veterinary orthopedics.
引用
收藏
页码:773 / 781
页数:9
相关论文
共 34 条
[1]   Clinical Outcomes and Complications after Open Reduction and Internal Fixation Utilizing Conventional Plates in 65 Distal Radial and Ulnar Fractures of Miniature- and Toy-Breed Dogs [J].
Aikawa, Takeshi ;
Miyazaki, Yuta ;
Shimatsu, Taichi ;
Iizuka, Kyoko ;
Nishimura, Masaaki .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2018, 31 (03) :214-217
[2]  
Aumarm W., 2012, J. Thai Vet. Pract., V24, P15
[3]   Orthopedic clinical techniques femur fracture repair [J].
Beale, B .
CLINICAL TECHNIQUES IN SMALL ANIMAL PRACTICE, 2004, 19 (03) :134-150
[4]   A comparative finite element simulation of locking compression plate materials for tibial fracture treatment [J].
Beirami, Sami ;
Nikkhoo, Mohammad ;
Hassani, Kamran ;
Karimi, Alireza .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (10) :1064-1072
[5]   Pitfalls and limits of locking plates [J].
Bel, Jean-Christophe .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2019, 105 (01) :S103-S109
[6]   An in vitro biomechanical study of bone plate and interlocking nail in a canine diaphyseal femoral fracture model [J].
Bernarde, A ;
Diop, A ;
Maurel, N ;
Viguier, E .
VETERINARY SURGERY, 2001, 30 (05) :397-408
[7]  
Bidari K., 2023, Pharma Innov. J., V12, P531
[8]  
Brunnberg L, 1998, KLEINTIERPRAXIS, V43, P579
[9]   Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate [J].
Chao, Peini ;
Conrad, Bryan P. ;
Lewis, Daniel D. ;
Horodyski, MaryBeth ;
Pozzi, Antonio .
BMC VETERINARY RESEARCH, 2013, 9
[10]  
Demner D, 2014, VET COMP ORTHOPAED, V27, P422, DOI 10.3415/VCOT-14-03-0040