Failure to treat ventricular tachycardia or ventricular fibrillation by implantable cardioverter-defibrillators

被引:0
作者
Hauser, Robert G. [1 ]
Sharkey, Scott W. [1 ]
Bahbah, Ali [1 ]
Kapphahn-Bergs, Melanie [1 ]
Zishiri, Edwin T. [1 ]
Witt, Dawn [1 ]
Sengupta, Jay N. [1 ]
Swerdlow, Charles D. [2 ]
机构
[1] Minneapolis Heart Inst Fdn, Joseph F Novogratz Family Heart Rhythm Sci Ctr, Minneapolis, MN USA
[2] Cedars Sinai Med Ctr, Smidt Heart Inst, Los Angeles, CA USA
关键词
Implantable cardioverter-defibrillator; Cardiac resynchronization defibrillator; Ventricular tachycardia; Ventricular fibrillation; Malfunction; Injury; Death; Detection; Discriminator; Undersensing; UNDERDETECTION;
D O I
10.1016/j.hrthm.2024.08.055
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Normally functioning implantable cardioverter-defibrillators (ICDs) with intact lead systems occasionally fail to deliver therapy for ventricular tachycardia/fibrillation (VT/VF) or deliver it only after clinically significant delays (failure-to-treat). OBJECTIVE We aimed to investigate ICD failure-to-treat VT/VF in a large patient cohort. METHODS We searched the US Food and Drug Administration's online Manufacturer and User Facility Device Experience (MAUDE) database from 2019 to 2023 for manufacturer-verified reports in which normally functioning ICDs failed to treat VT/VF. RESULTS We identified 854 reports classified as deaths (n = 96 [11.2%]), injuries (n = 585 [68.5%]), or malfunctions (n = 173 [20.3%]) for normally functioning ICDs. The most common causes were misclassification as supraventricular tachycardia (SVT) or atrial fibrillation (AF; 54.8%), undersensing (21.1%), and failure to satisfy programmed rate/duration criteria (8.7%). Most events caused by misclassification as SVT/AF (89.5%) and failure to satisfy rate/duration criteria (70.3%) were VT; most caused by undersensing were either VF (54.4%) or not specified as VT or VF (19.4%). Undersensing caused 65.6% of deaths, although it represented only 21.1% of reports. In the United States, the number of reports increased faster than that of ICD patients. CONCLUSION In the largest reported series of failure-to-treat VT/VF by normally functioning ICDs, the most common cause was misclassification of VT as SVT/AF; the most common cause of death was undersensing of VF. Although relatively few patients with normally functioning ICDs experience failure-to-treat VT/VF, the absolute number of verified MAUDE reports suggests that more work is needed to quantify the magnitude of the problem, to identify root causes, and to develop solutions.
引用
收藏
页码:1249 / 1254
页数:6
相关论文
共 16 条
[1]   Fatal Undersensing of Ventricular Fibrillation due to Intermittent High-Amplitude R Waves [J].
Anguera, Ignasi ;
Sabate, Xavier ;
Sugranes, Gemma ;
Cequier, Angel .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (09) :e284-e286
[2]   Underdetection of ventricular fibrillation despite ICD testing and high sinus R waveUndersensing von Kammerflimmern trotz ICD-Testung und hoher Sinus-R-Wellen-Amplitude [J].
Dirk Bastian .
Herzschrittmachertherapie + Elektrophysiologie, 2013, 24 (2) :131-135
[3]  
Brignoli Marcello, 2023, HeartRhythm Case Rep, V9, P929, DOI 10.1016/j.hrcr.2023.10.002
[4]   Electrogram morphology discriminators in implantable cardioverter defibrillators: A comparative evaluation [J].
Frontera, Antonio ;
Strik, Marc ;
Eschalier, Romain ;
Biffi, Mauro ;
Pereira, Bruno ;
Welte, Nicolas ;
Chauvel, Remi ;
Mondoly, Pierre ;
Laborderie, Julien ;
Bernis, Jean-Paul ;
Clementy, Nicolas ;
Reuter, Sylvain ;
Garrigue, Stephane ;
Deplagne, Antoine ;
Vernooy, Kevin ;
Pillois, Xavier ;
Haissaguerre, Michel ;
Dubois, Remi ;
Ritter, Philippe ;
Bordachar, Pierre ;
Ploux, Sylvain .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (06) :1493-1506
[5]   Mechanisms of Undersensing by a Noise Detection Algorithm That Utilizes Far-Field Electrograms With Near-Field Bandpass Filtering [J].
Koneru, Jayanthi N. ;
Swerdlow, Charles D. ;
Ploux, Sylvain ;
Sharma, Parikshit S. ;
Kaszala, Karoly ;
Tan, Alex Y. ;
Huizar, Jose F. ;
Vijayaraman, Pugazhendi ;
Kenigsberg, David ;
Ellenbogen, Kenneth A. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (02) :224-232
[6]   Underdetection of ventricular tachycardia by an implantable cardioverter-defibrillator due to a locally prolonged arrhythmia cycle length [J].
Lee, Illkyu-Oliver ;
Ukena, Christian ;
Boehm, Michael ;
Buob, Axel ;
Neuberger, Hans-Ruprecht .
CLINICAL RESEARCH IN CARDIOLOGY, 2013, 102 (12) :923-925
[7]   Inappropriate ICD classification of ventricular fibrillation as sinus tachycardia. What is the etiology? [J].
Mahmood, Rafat ;
Gehi, Anil K. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2024, 35 (05) :1026-1032
[8]   ICD-unresponsive ventricular arrhythmias [J].
Michaud, Josee ;
Horduna, Irina ;
Dubuc, Marc ;
Khairy, Paul .
HEART RHYTHM, 2009, 6 (12) :1827-1829
[9]  
Stadler Robert W., 2012, Official Gazette of the United States Patent and Trademark Office Patents, Patent No. US 08306614
[10]   UNDERDETECTION OF VENTRICULAR-TACHYCARDIA BY ALGORITHMS TO ENHANCE SPECIFICITY IN A TIERED-THERAPY CARDIOVERTER-DEFIBRILLATOR [J].
SWERDLOW, CD ;
AHERN, T ;
CHEN, PS ;
HWANG, C ;
GANG, E ;
MANDEL, W ;
KASS, RM ;
PETER, CT .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (02) :416-424