Intra and postoperative electrocochleography in cochlear implant recipients: retrospective chart review

被引:0
作者
Ordonez Ordonez, Leonardo Elias [1 ,2 ,3 ]
Angulo Martinez, Esther Sofia [3 ]
Vanegas, Silvia Carolina [1 ,3 ]
Hernandez, Estefany Catherine [1 ]
Peroza Fuentes, Eduardo [4 ]
Dwyer, Robert T. [5 ]
Martinho-Carvalho, Ana Claudia [5 ]
机构
[1] Clin Univ Colombia, Dept Otolaryngol, Clin Colsanitas SA, Calle 23 66-46, Keralty, Bogota, Colombia
[2] Fdn Univ Sanitas Unisanitas, Fac Med, Dept Otolaryngol, Keralty, Bogota, Colombia
[3] Anaboleas Res Team, Clin Univ Colombia & Fdn Univ Sanitas Unisanitas, Colciencias 2021 Minist Sci Technol & Innovat Colo, Bogota, Colombia
[4] San Roque IPS, Dept Audiol, Bogota, Colombia
[5] Adv Bion LLC, Valencia, CA USA
关键词
Cochlear implants; intracochlear electrocochleography; intraoperative electrocochleography; hearing preservation; sensorineural hearing loss; SOUND SOURCE LOCALIZATION; HEARING-PRESERVATION; INTRACOCHLEAR ELECTROCOCHLEOGRAPHY; STIMULATION; THRESHOLDS; ELECTRODE; TRIAL;
D O I
10.1080/14670100.2025.2515703
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
ObjectivesThe purpose of the study was (1) to retrospectively analyze ECochG data and compare it with the current literature. (2) to explore the relationship between ECochG and post-op audiological assessments.MethodsThis retrospective study included all patients with intraoperative and postoperative ECochG measurements and postoperative audiologic testing. ECochG were performed during CI insertion regardless of the preoperative hearing status. Preoperative and postoperative assessments were extracted from the medical records.ResultsA total of 35 patients (42 ears) met the inclusion criteria. Intra-op ECochG response was measurable in 17 of 42 ears (40.5%). The mean peak amplitude found was 103.3 mu volts. The type of electrode array was not associated with the ECochG amplitude (p = .630). Post-op ECochG thresholds were significantly correlated with behavioral thresholds after a month of CI implantation (p < 0.05).ConclusionsECochG responses were measurable in individuals with mean preop pure tone thresholds of 66 dB HL at 250 Hz, 77 dB HL at 500 Hz, 84 dB HL at 1 kHz and 92 dB HL at 2kHz. The type of electrode was not associated with the ECochG peak nor with greater preservation of residual hearing. Post-op ECochG thresholds were significantly correlated with post-op thresholds.
引用
收藏
页数:11
相关论文
共 34 条
[1]   Acoustic component programming in children with cochlear implants using electrocochleography [J].
Agrawal, Smita ;
Coulthurst, Sarah ;
Nachman, Alison ;
Koka, Kanthaiah ;
Murray, Michael .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2022, 61 (09) :736-743
[2]   Postoperative Intracochlear Electrocochleography in Pediatric Cochlear Implant Recipients: Association to Audiometric Thresholds and Auditory Performance [J].
Attias, Joseph ;
Ulanovski, David ;
Hilly, Ohad ;
Greenstein, Tally ;
Sokolov, Merav ;
HabibAllah, Suhail ;
Mormer, Hen ;
Raveh, Eyal .
EAR AND HEARING, 2020, 41 (05) :1135-1143
[3]  
Brockmeier S J, 2011, Cochlear Implants Int, V12, P10, DOI 10.1179/146701010X12677899497236
[4]   Unilateral Cochlear Implants for Severe, Profound, or Moderate Sloping to Profound Bilateral Sensorineural Hearing Loss A Systematic Review and Consensus Statements [J].
Buchman, Craig A. ;
Gifford, Rene H. ;
Haynes, David S. ;
Lenarz, Thomas ;
O'Donoghue, Gerard ;
Adunka, Oliver ;
Biever, Allison ;
Briggs, Robert J. ;
Carlson, Matthew L. ;
Dai, Pu ;
Driscoll, Colin L. ;
Francis, Howard W. ;
Gantz, Bruce J. ;
Gurgel, Richard K. ;
Hansen, Marlan R. ;
Holcomb, Meredith ;
Karltorp, Eva ;
Kirtane, Milind ;
Larky, Jannine ;
Mylanus, Emmanuel A. M. ;
Roland, J. Thomas, Jr. ;
Saeed, Shakeel R. ;
Skarzynski, Henryk ;
Skarzynski, Piotr H. ;
Syms, Mark ;
Teagle, Holly ;
Van de Heyning, Paul H. ;
Vincent, Christophe ;
Wu, Hao ;
Yamasoba, Tatsuya ;
Zwolan, Terry .
JAMA OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2020, 146 (10) :942-953
[5]   Clinical experiences with intraoperative electrocochleography in cochlear implant recipients and its potential to reduce insertion trauma and improve postoperative hearing preservation [J].
Buechner, Andreas ;
Bardt, Michael ;
Haumann, Sabine ;
Geissler, Gunnar ;
Salcher, Rolf ;
Lenarz, Thomas .
PLOS ONE, 2022, 17 (04)
[6]   Comparison of Pure-Tone Thresholds and Cochlear Microphonics Thresholds in Pediatric Cochlear Implant Patients [J].
Coulthurst, Sarah ;
Nachman, Alison J. ;
Murray, Mike T. ;
Koka, Kanthaiah ;
Saoji, Aniket A. .
EAR AND HEARING, 2020, 41 (05) :1320-1326
[7]   Combining acoustic and electric stimulation in the service of speech recognition [J].
Dorman, Michael F. ;
Gifford, Rene H. .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2010, 49 (12) :912-919
[8]   A Multicenter Comparison of 1-yr Functional Outcomes and Programming Differences Between the Advanced Bionics Mid-Scala and SlimJ Electrode Arrays [J].
Eitutis, Susan T. ;
Vickers, Deborah A. ;
Tebbutt, Karen ;
Thomas, Tisa ;
Jiang, Dan ;
de Klerk, Anel ;
Clemesha, Jennifer ;
Chung, Mark ;
Bance, Manohar L. .
OTOLOGY & NEUROTOLOGY, 2023, 44 (10) :E730-E738
[9]  
Friedland David R, 2009, Trends Amplif, V13, P124, DOI 10.1177/1084713809336422
[10]  
Gfeller Kate E, 2006, Audiol Neurootol, V11 Suppl 1, P12, DOI 10.1159/000095608