Evaluating smartwatch-based detection of supraventricular tachycardia and atrial fibrillation in the emergency department

被引:0
作者
Alpar, Suleyman [1 ]
Tatliparmak, Ali Cankut [2 ]
机构
[1] Istanbul Beykent Univ, Istanbul Beykent Univ Hosp, Dept Emergency Med, Istanbul, Turkiye
[2] Uskudar Univ, Dept Emergency Med, Fac Med, Istanbul, Turkiye
关键词
Emergency department; Supraventricular tachycardia; Atrial fibrillation; Photoplethysmography; Electrocardiogram;
D O I
10.1016/j.ajem.2025.05.037
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: In this study, we categorized smartwatches into two groups: those with electrocardiogram (ECG) capability and those utilizing photoplethysmography (PPG) sensors. The aim was to evaluate the diagnostic performance of these devices in differentiating atrial fibrillation (AF) from supraventricular tachycardia (SVT) among patients presenting to the emergency department (ED) with an irregular rhythm alert. Methods: This retrospective, single-center observational study was conducted in the ED of a tertiary hospital over a two-year period. The study included adult patients (>= 22 years) who presented to the ED following an irregular rhythm alert generated by their smartwatch. Smartwatches were classified as either ECG-capable devices or PPG-based devices. Final rhythm diagnoses were confirmed by two independent emergency physicians based on 12-lead ECG findings obtained during the ED evaluation. Results: A total of 424 patients were included. Of these, 63.7 % (n = 270) used ECG-capable smartwatches, and 66.3 % (n = 281) received an AF alert. Among all alerts, 60.8 % (n = 258) were ECG-based and 39.2 % (n = 166) were PPG-based. Common presenting symptoms included palpitations (56.1 %), dizziness (21.7 %), and dyspnea (23.6 %), while 8.5 % of patients were asymptomatic. In total, 44.6 % (n = 189) required pharmacologic or electrical interventions. Smartwatches demonstrated moderate accuracy for AF and SVT detection. PPG-based detection showed higher sensitivity than ECG-based detection (AF: 54.3 % vs. 44.9 %; SVT: 83.6 % vs. 79.2 %). Agreement with final ECG diagnoses was poor (Cohen's Kappa = 0), with significant misclassification for both arrhythmias (McNemar's p < 0.001), except for PPG-based SVT (p = 0.072). No significant predictors of false-positive AF alerts were found (area under the curve [AUC] = 0.538). False-positive SVT alerts were more common in younger patients (odds ratio [OR] = 0.92, p = 0.039), females (OR = 0.55, p = 0.048), and less frequent in diabetic patients (OR = 0.24, p = 0.001; AUC = 0.685). Conclusion: Among patients presenting to the ED with smartwatch-reported irregular rhythm alerts, PPG-based detection demonstrated higher sensitivity for SVT compared to ECG-based methods; however, the overall diagnostic agreement for both AF and SVT was limited. (c) 2025 Elsevier Inc. All rights are reserved, including those for text and data mining, AI training, and similar tech-nologies.
引用
收藏
页码:101 / 106
页数:6
相关论文
共 17 条
[1]   Smartwatch Electrocardiograms for Automated and Manual Diagnosis of Atrial Fibrillation: A Comparative Analysis of Three Models [J].
Abu-Alrub, Saer ;
Strik, Marc ;
Ramirez, F. Daniel ;
Moussaoui, Nadir ;
Racine, Hugo Pierre ;
Marchand, Hugo ;
Buliard, Samuel ;
Haissaguerre, Michel ;
Ploux, Sylvain ;
Bordachar, Pierre .
FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
[2]  
afibinstitute, 2022, Atrial Fibrillation: A Guide to Wearable ECG Smart Watches
[3]  
[Anonymous], 2021, Eur Heart J, V42, P507, DOI 10.1093/eurheartj/ehaa798
[4]  
Aydin OF, 2025, Gumushane Saglik Bilimleri Dergisi, V14, P41, DOI [10.37989/gumussagbil.1616151, DOI 10.37989/GUMUSSAGBIL.1616151]
[5]   Sensitivity and Specificity of Wearables for Atrial Fibrillation in Elderly Populations: A Systematic Review [J].
Babar, Faiza ;
Cheema, Abdul Manan ;
Ahmad, Zubair ;
Sarfraz, Azza ;
Sarfraz, Zouina ;
Ashraff, Huma ;
Tayyab, Hamnah ;
Bajwa, Ali Mustafa ;
Ali, Haider ;
Sarfraz, Muzna ;
Shahzad, Anum ;
Sabzwari, Syed Rafay Ali .
CURRENT CARDIOLOGY REPORTS, 2023, 25 (07) :761-779
[6]  
Castaneda Denisse, 2018, Int J Biosens Bioelectron, V4, P195, DOI 10.15406/ijbsbe.2018.04.00125
[7]   The WATCH AF Trial: SmartWATCHes for Detection of Atrial Fibrillation [J].
Doerr, Marcus ;
Nohturfft, Vivien ;
Brasier, Noe ;
Bosshard, Emil ;
Djurdjevic, Aleksandar ;
Gross, Stefan ;
Raichle, Christina J. ;
Rhinisperger, Mattias ;
Stoeckli, Raphael ;
Eckstein, Jens .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (02) :199-208
[8]   Mobile Photoplethysmographic Technology to Detect Atrial Fibrillation [J].
Guo, Yutao ;
Wang, Hao ;
Zhang, Hui ;
Liu, Tong ;
Liang, Zhaoguang ;
Xia, Yunlong ;
Yan, Li ;
Xing, Yunli ;
Shi, Haili ;
Li, Shuyan ;
Liu, Yanxia ;
Liu, Fan ;
Feng, Mei ;
Chen, Yundai ;
Lip, Gregory Y. H. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 74 (19) :2365-2375
[9]  
Margulescu AD, 2017, EXPERT REV CARDIOVAS, V15, P601, DOI 10.1080/14779072.2017.1355237
[10]   Contribution of atrial fibrillation to incidence and outcome of ischemic stroke - Results from a population-based study [J].
Marini, C ;
De Santis, F ;
Sacco, S ;
Russo, T ;
Olivieri, L ;
Totaro, R ;
Carolei, A .
STROKE, 2005, 36 (06) :1115-1119