Artifacts in magnetic resonance imaging of the head and neck: Unwanted effects caused by implant-supported restorations fabricated with different alloys

被引:1
作者
Bohner, Lauren [1 ]
Dirksen, Dieter [2 ]
Hanisch, Marcel [1 ]
Sesma, Newton [3 ]
Kleinheinz, Johannes [1 ]
Meier, Norbert [4 ]
机构
[1] Univ Hosp Muenster, Dept Craniomaxillofacial Surg, Waldeyerstr 12, D-48149 Munster, Germany
[2] Univ Hosp Muenster, Dept Prosthodont & Biomat, Munster, Germany
[3] Univ Sao Paulo, Sch Dent, Dept Prosthodont, Sao Paulo, Brazil
[4] Univ Hosp Muenster, Dept Radiol, Munster, Germany
关键词
SEMAC-VAT; REDUCTION; MRI;
D O I
10.1016/j.prosdent.2023.08.018
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. Implant-supported restorations may cause artifacts in magnetic resonance imaging (MRI) of the head and neck area. However, the effect of different alloys remains unclear. Purpose. The purpose of this in vitro study was to assess artifacts in head and neck MRI caused by implant-supported restorations with different alloys. Material and methods. Three dry mandibles were prepared to receive bilateral dental implants at the second premolar and second molar sites. Different alloy combinations were evaluated: titanium implants+cobalt chromium restorations; titanium-zirconium implants+cobalt chromium restorations; and zirconia implants+ceramic restorations. Specimens were imaged by using a 3-Tesla magnetic resonance scanner system (Achieva 3.0T TX; Philips GmbH) with a turbo-spin-echo sequence. Scan protocols were adjusted to optimize metal artifact reduction and shorten scan time. Artifact volumes were assessed and statistically analyzed by using the Kruskal-Wallis and Spearman tests (alpha=.05). Results. A statistically significant difference was found among artifact volume caused by different materials (P=.002). The presence of titanium alloy was correlated with the artifact volume (r=-.87). Artifacts were greater for titanium and fewer for titanium-zirconium alloys, whereas zirconia implants found only minimal artifacts. Conclusions. The dimension of artifacts produced by implant-supported restorations varied according to the material. (J Prosthet Dent 2025;133:1574-1579)
引用
收藏
页码:1574 / 1579
页数:6
相关论文
共 23 条
[1]   SEMAC-VAT and MSVAT-SPACE Sequence Strategies for Metal Artifact Reduction in 1.5T Magnetic Resonance Imaging [J].
Ai, Tao ;
Padua, Abraham ;
Goerner, Frank ;
Nittka, Mathias ;
Gugala, Zbigniew ;
Jadhav, Siddharth ;
Trelles, Miguel ;
Johnson, Raleigh F. ;
Lindsey, Ronald W. ;
Li, Xiaoming ;
Runge, Val M. .
INVESTIGATIVE RADIOLOGY, 2012, 47 (05) :267-276
[2]   Hybrid PET/MR Imaging: An Algorithm to Reduce Metal Artifacts from Dental Implants in Dixon-Based Attenuation Map Generation Using a Multiacquisition Variable-Resonance Image Combination Sequence [J].
Burger, Irene A. ;
Wurnig, Moritz C. ;
Becker, Anton S. ;
Kenkel, David ;
Delso, Gaspar ;
Veit-Haibach, Patrick ;
Boss, Andreas .
JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (01) :93-97
[3]   Reducing MRI susceptibility artefacts in implants using additively manufactured porous Ti-6Al-4V structures [J].
Carter, Luke N. ;
Addison, Owen ;
Naji, Nashwan ;
Seres, Peter ;
Wilman, Alan H. ;
Shepherd, Duncan E. T. ;
Grover, Liam ;
Cox, Sophie .
ACTA BIOMATERIALIA, 2020, 107 :338-348
[4]  
Chockattu Sherin Jose, 2018, Restor Dent Endod, V43, pe39, DOI [10.5395/rde.2018.43.e39, 10.5395/rde.2018.43.e39]
[5]   Evaluation of artifacts generated by titanium, zirconium, CT, and CBCT images:A phantom study [J].
Kocasarac, Husniye Demirturk ;
Ustaoglu, Gulbahar ;
Bayrak, Seval ;
Katkar, Rujuta ;
Geha, Hassem ;
Deahl, S. Thomas, II ;
Mealey, Brian L. ;
Danaci, Murat ;
Noujeim, Marcel .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2019, 127 (06) :535-544
[6]   Magnetic resonance imaging in zirconia-based dental implantology [J].
Duttenhoefer, Fabian ;
Mertens, Marianne E. ;
Vizkelety, Josef ;
Gremse, Felix ;
Stadelmann, Vincent A. ;
Sauerbier, Sebastian .
CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (10) :1195-1202
[7]   Material-Dependent Implant Artifact Reduction Using SEMAC-VAT and MAVRIC: A Prospective MRI Phantom Study [J].
Filli, Lukas ;
Jud, Lukas ;
Luechinger, Roger ;
Nanz, Daniel ;
Andreisek, Gustav ;
Runge, Val M. ;
Kozerke, Sebastian ;
Farshad-Amacker, Nadja A. .
INVESTIGATIVE RADIOLOGY, 2017, 52 (06) :381-387
[8]   Compressed Sensing SEMAC: 8-fold Accelerated High Resolution Metal Artifact Reduction MRI of Cobalt-Chromium Knee Arthroplasty Implants [J].
Fritz, Jan ;
Ahlawat, Shivani ;
Demehri, Shadpour ;
Thawait, Gaurav K. ;
Raithel, Esther ;
Gilson, Wesley D. ;
Nittka, Mathias .
INVESTIGATIVE RADIOLOGY, 2016, 51 (10) :666-676
[9]   Body MRI artifacts in clinical practice: A physicist's and radiologist's perspective [J].
Graves, Martin J. ;
Mitchell, Donald G. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (02) :269-287
[10]   Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI [J].
Gunzinger J.M. ;
Delso G. ;
Boss A. ;
Porto M. ;
Davison H. ;
von Schulthess G.K. ;
Huellner M. ;
Stolzmann P. ;
Veit-Haibach P. ;
Burger I.A. .
EJNMMI Physics, 1 (1) :1-14