Beam data commissioning, validation and clinical implementation of SRS gyroscopic system

被引:0
作者
Souto, Catarina [1 ,2 ,4 ]
Aguilar, Borja [1 ,2 ]
Dias, Vania [1 ,2 ]
Pascual, Enrique [1 ]
Farinha, Gabriel [2 ]
Marcelino, Carlos [2 ]
Rodrigues, Carlos [1 ,2 ]
Teles, Pedro [4 ,5 ]
Perez-Rozos, Alberto [1 ,2 ,3 ]
Sallabanda, Morena [1 ]
Sallabanda, Kita [1 ]
机构
[1] Hosp Viamed Santa Elena, IRCA, Madrid, Spain
[2] Mercurius Hlth, Porto Salvo, Portugal
[3] Serv Andaluz Salud, Seville, Spain
[4] Univ Porto, DFA, Fac Ciencias, Porto, Portugal
[5] CI IPO Porto, Porto, Portugal
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2025年 / 133卷
关键词
ZAP-X (R); SRS; Small field dosimetry; Commissioning; STEREOTACTIC RADIOSURGERY; QUALITY-ASSURANCE; DOSIMETRY; SIMULATION;
D O I
10.1016/j.ejmp.2025.104952
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The very unconventional design and features of the new gyroscopic radiosurgery ZAP-X (R) system (ZAP Surgical Systems) generates a challenge for beam data acquisition, characterization and validation. Treatment Planning System commissioning was performed with different detectors, filling the absence of methodologies and correction factors published in international guidelines. The congruence of detectors response was compared with corresponding SciMoCaTM Monte Carlo (MC) synthetic data (Scientific RT GmbH). A machine and patient- specific quality assurance (QA) protocol was also implemented. Materials and Methods: All data were acquired using a microsilicon and a microdiamond detectors, three ionization chambers (PTW Freiburg) and a plastic scintillation detector (PSD) (Blue Physics LLC). For patient-specific QA, an anthropomorphic phantom (RT Safe Inc), a 2D array (Sun Nuclear Corporation) and SciMoCaTM were used for 1D dose verification, 2D and 3D evaluation gamma assessment, respectively. Results: Agreement between detectors, for all PDDs, was better than 1% beyond the depth of dmax. For profiles, the microsilicon detector showed steeper penumbras due to the smaller detector width. A good congruence with MC data was found for the solid-state detectors. Output factors measurements demonstrated strong agreement between SciMoCaTM data and PSD data, with a maximum difference of 2.1%. The patient QA results validated the quality of the commissioning data. Conclusions: Our results showed strong agreement between different measurement methods. MC simulations proved to be an important tool for commissioning validation. The methods and materials discussed in this article may be of use to other facilities where this SRS gyroscope system is installed.
引用
收藏
页数:12
相关论文
共 53 条
  • [1] Adler JR, 2017, CUREUS J MED SCIENCE, V9, DOI 10.7759/cureus.1663
  • [2] AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    Almond, PR
    Biggs, PJ
    Coursey, BM
    Hanson, WF
    Huq, MS
    Nath, R
    Rogers, DWO
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1847 - 1870
  • [3] Nanomaterials and nanotechnology for the delivery of agrochemicals: strategies towards sustainable agriculture
    An, Changcheng
    Sun, Changjiao
    Li, Ningjun
    Huang, Bingna
    Jiang, Jiajun
    Shen, Yue
    Wang, Chong
    Zhao, Xiang
    Cui, Bo
    Wang, Chunxin
    Li, Xingye
    Zhan, Shenshan
    Gao, Fei
    Zeng, Zhanghua
    Cui, Haixin
    Wang, Yan
    [J]. JOURNAL OF NANOBIOTECHNOLOGY, 2022, 20 (01)
  • [4] [Anonymous], 2023, Netherlands Commission on Radiation Dosimetry NCR Report n 36, DOI [10.25030/ncs-036, DOI 10.25030/NCS-036]
  • [5] [Anonymous], 2015, Netherlands Commission on Radiation Dosimetry NCR Report n 25 Process Management and Quality Assurance for Intracranial Stereotactic Treatment, DOI [10.25030/ncs-025, DOI 10.25030/NCS-025]
  • [6] [Anonymous], 2017, Technical Report Series, V483
  • [7] [Anonymous], ZAP-X Radiosurgery System Beam Data Manual
  • [8] Aspradakis M., 2010, IPEM REPORT 103 SMAL
  • [9] COMP report: CPQR technical quality control guidelines for Gamma Knife radiosurgery
    Berndt, Anita
    van Prooijen, Monique
    Guillot, Mathieu
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 365 - 367
  • [10] BluePhysics, 2023, Model 10