Simultaneous ultra-high resolution multi-energy cardiac imaging in a dual-source photon counting detector CT system

被引:0
作者
Shanblatt, Elisabeth [1 ]
Rajendran, Kishore [2 ]
Swicklik, Joseph R. [2 ]
Allmendinger, Thomas [3 ]
Schmidt, Bernhard
Flohr, Thomas [4 ]
Mccollough, Cynthia H. [2 ]
Leng, Shuai [2 ]
机构
[1] Siemens Med Solut USA Inc, Malvern, PA 19355 USA
[2] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[3] Siemens Healthineers, D-91301 Forchheim, Germany
[4] Friedrich Alexander Univ, Inst Radiol, D-91054 Erlangen, Germany
基金
美国国家卫生研究院;
关键词
photon counting; cardiac CT; multi-energy imaging; ultra-high-spatial resolution; RECONSTRUCTION; ALGORITHM;
D O I
10.1093/bjr/tqaf069
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives To assess quantitative and qualitative image characteristics in phantoms and patients using a simultaneous ultra-high resolution (UHR) and multi-energy (ME) dual-source cardiac scan mode in a photon counting detector (PCD) CT scanner.Methods UHR-ME cardiac scans were performed to determine the spatial resolution and multi-energy performance. UHR reconstructions were completed to measure transverse and axial spatial resolution in terms of modulation transfer function (MTF) and slice sensitivity profile (SSP). Virtual monoenergetic images (VMIs), virtual non-contrast (VNC) images, and iodine concentration images were created using moderately sharp reconstruction kernels to assess ME performance. With IRB approval, coronary CT angiograms were acquired in 19 patients using the UHR-ME mode, from which iodine contrast-to-noise ratio (CNR) was measured. Various ME post-processing tasks were performed to show the utility of the spectral capabilities.Results VMIs showed a mean CT number error of 6.0%, while iodine accuracy was 0.71 mg/cc (root-mean-square-error). Spatial resolution matched well with a standard UHR mode. Patient images showed a trend of higher CNR at lower keV, while higher keV images showed decreased calcium blooming. ME techniques to separate iodine and calcium were used successfully.Conclusions The combination of UHR-ME cardiac mode retains UHR image quality while providing flexibility for ME post-processing.Advances in knowledge Cardiac UHR-ME PCD-CT scanning offers simultaneous UHR and ME imaging without loss of spatial or temporal resolution or quantitative accuracy.
引用
收藏
页数:7
相关论文
共 26 条
  • [1] Photon-Counting Detector CT-Based Vascular Calcium Removal Algorithm Assessment Using a Cardiac Motion Phantom
    Allmendinger, Thomas
    Nowak, Tristan
    Flohr, Thomas
    Klotz, Ernst
    Hagenauer, Junia
    Alkadhi, Hatem
    Schmidt, Bernhard
    [J]. INVESTIGATIVE RADIOLOGY, 2022, 57 (06) : 399 - 405
  • [2] Radiology: Cardiothoracic Imaging Highlights 2023
    Aquino, Gilberto J.
    Mastrodicasa, Domenico
    Alabed, Samer
    Abohashem, Shady
    Wen, Lingyi
    Gill, Ritu R.
    Bardo, Dianna M. E.
    Abbara, Suhny
    Hanneman, Kate
    [J]. RADIOLOGY-CARDIOTHORACIC IMAGING, 2024, 6 (02):
  • [3] Berger M.J., 2010, XCOM: Photon Cross Sections Database
  • [4] Diagnostic Accuracy of Computed Tomography-Derived Fractional Flow Reserve A Systematic Review
    Cook, Christopher M.
    Petraco, Ricardo
    Shun-Shin, Matthew J.
    Ahmad, Yousif
    Nijjer, Sukhjinder
    Al-Lamee, Rasha
    Kikuta, Yuetsu
    Shiono, Yasutsugu
    Mayet, Jamil
    Francis, Darrel P.
    Sen, Sayan
    Davies, Justin E.
    [J]. JAMA CARDIOLOGY, 2017, 2 (07) : 803 - 810
  • [5] Myocardial extracellular volume quantification with computed tomography-current status and future outlook
    Cundari, Giulia
    Galea, Nicola
    Mergen, Victor
    Alkadhi, Hatem
    Eberhard, Matthias
    [J]. INSIGHTS INTO IMAGING, 2023, 14 (01)
  • [6] Coronary Calcium Scoring with First Generation Dual-Source Photon-Counting CT-First Evidence from Phantom and In-Vivo Scans
    Eberhard, Matthias
    Mergen, Victor
    Higashigaito, Kai
    Allmendinger, Thomas
    Manka, Robert
    Flohr, Thomas
    Schmidt, Bernhard
    Euler, Andre
    Alkadhi, Hatem
    [J]. DIAGNOSTICS, 2021, 11 (09)
  • [7] Coronary Computed Tomography Angiography-Based Calcium Scoring In Vitro and In Vivo Validation of a Novel Virtual Noniodine Reconstruction Algorithm on a Clinical, First-Generation Dual-Source Photon Counting-Detector System
    Emrich, Tilman
    Aquino, Gilberto
    Schoepf, U. Joseph
    Braun, Franziska M.
    Risch, Franka
    Bette, Stefanie J.
    Woznicki, Piotr
    Decker, Josua A.
    O'Doherty, Jim
    Brandt, Verena
    Allmendinger, Thomas
    Nowak, Tristan
    Schmidt, Bernhard
    Flohr, Thomas
    Kroencke, Thomas J.
    Scheurig-Muenkler, Christian
    Varga-Szemes, Akos
    Schwarz, Florian
    [J]. INVESTIGATIVE RADIOLOGY, 2022, 57 (08) : 536 - 543
  • [8] Improved Detection of Small and Low-Density Plaques in Virtual Noncontrast Imaging-based Calcium Scoring at Photon-Counting Detector CT
    Fink, Nicola
    Emrich, Tilman
    Schoepf, U. Joseph
    Zsarnoczay, Emese
    O'Doherty, Jim
    Halfmann, Moritz C.
    Griffith, Joseph P., III
    Pinos, Daniel
    Suranyi, Pal
    Baruah, Dhiraj
    Kabakus, Ismail M.
    Ricke, Jens
    Varga-Szemes, Akos
    [J]. RADIOLOGY-CARDIOTHORACIC IMAGING, 2024, 6 (04):
  • [9] Image reconstruction and image quality evaluation for a dual source CT scanner
    Flohr, T. G.
    Bruder, H.
    Stierstorfer, K.
    Petersilka, M.
    Schmidt, B.
    McCollough, C. H.
    [J]. MEDICAL PHYSICS, 2008, 35 (12) : 5882 - 5897
  • [10] First performance evaluation of a dual-source CT (DSCT) system
    Flohr, TG
    McCollough, CH
    Bruder, H
    Petersilka, M
    Gruber, K
    Süss, C
    Grasruck, M
    Stierstorfer, K
    Krauss, B
    Raupach, R
    Primak, AN
    Küttner, A
    Achenbach, S
    Becker, C
    Kopp, A
    Ohnesorge, BM
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (02) : 256 - 268