Long-Term Performance of Two Systems for Automated Insulin Delivery in Adults With Type 1 Diabetes: An Observational Study

被引:0
作者
Fisker, Sanne [1 ]
Christensen, Mia [2 ]
Bach, Ermina [1 ,3 ]
Bibby, Bo Martin [4 ]
Hansen, Klavs Wuergler [2 ,5 ]
机构
[1] Aarhus Univ Hosp, Steno Diabet Ctr Aarhus, Aarhus, Denmark
[2] Silkeborg Reg Hosp, Med Diagnost Ctr, Silkeborg, Denmark
[3] Viborg Reg Hosp, Med Diagnost Ctr, Viborg, Denmark
[4] Aarhus Univ, Dept Publ Hlth, Sect Biostat, Aarhus, Denmark
[5] Aarhus Univ, Dept Internal Med, Aarhus, Denmark
关键词
automated insulin delivery; continuous glucose monitoring; glucose rate of change; glucose time in range; type; 1; diabetes; REAL-WORLD USERS; INTERNATIONAL CONSENSUS; GLUCOSE; RANGE; TIME;
D O I
10.1002/edm2.70043
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims To compare glycaemic outcomes for two automated insulin delivery (AID) systems, the Tandem Control IQ (CIQ) and the MiniMed 780G (MM780G). Material and Methods In this observational study, we evaluated 60 days of glycaemic data from 139 persons with type 1 diabetes (CIQ: 79 persons, MM780G: 60 persons), who had an active glucose sensor time >= 85%. Results The time with AID was median 620 (IQR, 439-755) days for CIQ users and 509 (429-744) days for MM780G users (p = 0.26). The last HbA1c before initiation of AID was 59.7 mmol/mol in CIQ and 60.1 mmol/mol in MM780G (p = 0.88). The time with an active glucose sensor was higher for CIQ than MM780G (median 98.5 (97.4-98.0)% vs. 96.5 (94.9-97.0)%, p < 0.001). Time in range (TIR, glucose 3.9-10.0 mmol/L) was lower in CIQ than MM780G (mean 68.9% +/- 11.4% vs. 73.7% +/- 12.0%, p = 0.02) as was time in tight range (TITR) (glucose 3.9-7.8 mmol/L) (43.0% +/- 12.2% vs. 48.4% +/- 12.7%, p = 0.01). The difference in TIR (4.2 (95% CI 1.0-7.5)%, p = 0.01) and TITR (5.0 (1.4-8.6)%, p < 0.01) remained statistically significant in a multiple regression model controlling for various baseline variables. Time with an absolute rate of glucose change > 1.5 mmol/L/15 min was higher in CIQ than MM780G (9.4 (IQR, 7.2-13.3)% vs. 7.4 (5.2-10.4)%, p < 0.001). Conclusions The CIQ system had a higher active glucose sensor time but a lower TIR, TITR, and a higher time with a rapid glucose rate of change than the MM780G system.
引用
收藏
页数:10
相关论文
共 41 条
[1]   One-year follow-up comparison of two hybrid closed-loop systems in Italian children and adults with type 1 diabetes [J].
Bassi, Marta ;
Patti, Luca ;
Silvestrini, Irene ;
Strati, Marina Francesca ;
Ponzano, Marta ;
Minuto, Nicola ;
Maggi, Davide .
FRONTIERS IN ENDOCRINOLOGY, 2023, 14
[2]   A Comparison of Two Hybrid Closed-Loop Systems in Italian Children and Adults With Type 1 Diabetes [J].
Bassi, Marta ;
Teliti, Marsida ;
Lezzi, Marilea ;
Iosca, Arianna ;
Strati, Marina Francesca ;
Carmisciano, Luca ;
d'Annunzio, Giuseppe ;
Minuto, Nicola ;
Maggi, Davide .
FRONTIERS IN ENDOCRINOLOGY, 2022, 12
[3]   Continuous glucose monitoring and metrics for clinical trials: an international consensus statement [J].
Battelino, Tadej ;
Alexander, Charles M. ;
Amiel, Stephanie A. ;
Arreaza-Rubin, Guillermo ;
Beck, Roy W. ;
Bergenstal, Richard M. ;
Buckingham, Bruce A. ;
Carroll, James ;
Ceriello, Antonio ;
Chow, Elaine ;
Choudhary, Pratik ;
Close, Kelly ;
Danne, Thomas ;
Dutta, Sanjoy ;
Gabbay, Robert ;
Garg, Satish ;
Heverly, Julie ;
Hirsch, Irl B. ;
Kader, Tina ;
Kenney, Julia ;
Kovatchev, Boris ;
Laffel, Lori ;
Maahs, David ;
Mathieu, Chantal ;
Mauricio, Didac ;
Nimri, Revital ;
Nishimura, Rimei ;
Scharf, Mauro ;
Del Prato, Stefano ;
Renard, Eric ;
Rosenstock, Julio ;
Saboo, Banshi ;
Ueki, Kohjiro ;
Umpierrez, Guillermo E. ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
LANCET DIABETES & ENDOCRINOLOGY, 2023, 11 (01) :42-57
[4]   Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range [J].
Battelino, Tadej ;
Danne, Thomas ;
Bergenstal, Richard M. ;
Amiel, Stephanie A. ;
Beck, Roy ;
Biester, Torben ;
Bosi, Emanuele ;
Buckingham, Bruce A. ;
Cefalu, William T. ;
Close, Kelly L. ;
Cobelli, Claudio ;
Dassau, Eyal ;
DeVries, J. Hans ;
Donaghue, Kim C. ;
Dovc, Klemen ;
Doyle, Francis J. ;
Garg, Satish ;
Grunberger, George ;
Heller, Simon ;
Heinemann, Lutz ;
Hirsch, Irl B. ;
Hovorka, Roman ;
Jia, Weiping ;
Kordonouri, Olga ;
Kovatchev, Boris ;
Kowalski, Aaron ;
Laffel, Lori ;
Levine, Brian ;
Mayorov, Alexander ;
Mathieu, Chantal ;
Murphy, Helen R. ;
Nimri, Revital ;
Norgaard, Kirsten ;
Parkin, Christopher G. ;
Renard, Eric ;
Rodbard, David ;
Saboo, Banshi ;
Schatz, Desmond ;
Stoner, Keaton ;
Urakami, Tatsuiko ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
DIABETES CARE, 2019, 42 (08) :1593-1603
[5]   A Multicenter Prospective Evaluation of the Benefits of Two Advanced Hybrid Closed-Loop Systems in Glucose Control and Patient-Reported Outcomes in a Real-world Setting [J].
Beato-Vibora, Pilar Isabel ;
Chico, Ana ;
Moreno-Fernandez, Jesus ;
Bellido-Castaneda, Virginia ;
Nattero-Chavez, Lia ;
Picon-Cesar, Maria Jose ;
Martinez-Brocca, Maria Asuncion ;
Gimenez-Alvarez, Marga ;
Aguilera-Hurtado, Eva ;
Climent-Biescas, Elisenda ;
Azriel-Mir, Sharona ;
Rebollo-Roman, Angel ;
Yoldi-Vergara, Carmen ;
Pazos-Couselo, Marcos ;
Alonso-Carril, Nuria ;
Quiros, Carmen .
DIABETES CARE, 2024, 47 (02) :216-224
[6]   A Comparison of Continuous Glucose Monitoring-Measured Time-in-Range 70-180 mg/dL Versus Time-in-Tight-Range 70-140 mg/dL [J].
Beck, Roy W. ;
Raghinaru, Dan ;
Calhoun, Peter ;
Bergenstal, Richard M. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2024, 26 (03) :151-155
[7]   Glucose Management Indicator (GMI): A New Term for Estimating A1C From Continuous Glucose Monitoring [J].
Bergenstal, Richard M. ;
Beck, Roy W. ;
Close, Kelly L. ;
Grunberger, George ;
Sacks, David B. ;
Kowalski, Aaron ;
Brown, Adam S. ;
Heinemann, Lutz ;
Aleppo, Grazia ;
Ryan, Donna B. ;
Riddlesworth, Tonya D. ;
Cefalu, William T. .
DIABETES CARE, 2018, 41 (11) :2275-2280
[8]   Sustained 3-Year Improvement of Glucose Control With Hybrid Closed Loop in Children With Type 1 Diabetes While Going Through Puberty [J].
Bismuth, Elise ;
Tubiana-Rufi, Nadia ;
Rynders, Corey A. ;
Dalla-Vale, Fabienne ;
Bonnemaison, Elisabeth ;
Coutant, Regis ;
Farret, Anne ;
Poidvin, Amelie ;
Bouhours-Nouet, Natacha ;
Storey, Caroline ;
Donzeau, Aurelie ;
Deboer, Mark D. ;
Breton, Marc D. ;
Villard, Orianne ;
Renard, Eric .
DIABETES CARE, 2024, 47 (09) :1696-1703
[9]   Time in Tight Glucose Range in Type 1 Diabetes: Predictive Factors and Achievable Targets in Real-World Users of the MiniMed 780G System [J].
Castaneda, Javier ;
Arrieta, Arcelia ;
van den Heuvel, Tim ;
Battelino, Tadej ;
Cohen, Ohad .
DIABETES CARE, 2024, 47 (05) :790-797
[10]   Predictors of time in target glucose range in real-world users of the MiniMed 780G system [J].
Castaneda, Javier ;
Mathieu, Chantal ;
Aanstoot, Henk-Jan ;
Arrieta, Arcelia ;
Da Silva, Julien ;
Shin, John ;
Cohen, Ohad .
DIABETES OBESITY & METABOLISM, 2022, 24 (11) :2212-2221