Comparison of Enzyme-Linked Immunosorbent Assay and Lateral Flow Assay to Measure Thyroid-Stimulating Hormone and Free T4 in Human Serum
被引:0
作者:
Advincula-Espino, Rolando
[1
]
Li, Jair
[2
]
Rosales-Rimache, Jaime
[3
]
机构:
[1] Univ Nacl Mayor San Marcos, Fac Farm & Bioquim, Lima, Peru
[2] Univ Nacl Federico Villarreal, Fac Tecnol Med, Lima, Peru
[3] Univ Continental, Carrera Med, Huancayo, Peru
关键词:
enzyme-linked immunosorbent assay;
free tetraiodothyronine;
lateral flow assay;
thyroid-stimulating hormone;
IFCC WORKING GROUP;
FUNCTION TESTS;
STANDARDIZATION;
AUTOANTIBODIES;
IMMUNOASSAY;
THYROXINE;
D O I:
10.1002/jcla.70032
中图分类号:
R446 [实验室诊断];
R-33 [实验医学、医学实验];
学科分类号:
1001 ;
摘要:
Background: Thyroid disorders affect millions of people around the world, mainly women. Measurement of hormones is critical in their identification, treatment, and monitoring; Therefore, the methods must be as reliable and valid as possible. Methods: We designed a cross-sectional study to compare thyroid-stimulating hormone (TSH) and free tetraiodothyronine (fT4) concentrations using ELISA (Enzyme-Linked Immunosorbent Assay) and LFA (Lateral Flow Assay) in human serum samples. We obtained 96 serum samples that were evaluated with a commercial kit for ELISA (Accubind Monobite, Lake Forest, California, US) and LFA (Micropofit Fluorecare, Guangdong, Shenzhen, China). We used the Wilcoxon nonparametric test, Bland-Altman, and Passing-Bablok regression to compare TSH and fT4 concentrations obtained by ELISA and LFA. Results: The median TSH concentrations obtained by ELISA and LFA were 1.92 and 2.11 mu IU/mL, and for fT4 were 1.14 and 1.10 ng/dL. On the other hand, the Spearman's rho between ELISA and LFA for TSH was 0.845. The TSH concentrations between ELISA and LFA had significant differences (p < 0.05). About fT4, the Spearman's rho between methods was 0.348. The bias for TSH was -0.315, and for fT4 was -0.013. The Deming regression for TSH (p = 0.309) and fT4 (=0.938) shows that the levels obtained by both methods do not present significant differences, and the Passing-Bablok regression identifies significant bias between both methods, especially for fT4, in the range of concentrations studied. Conclusion: TSH measurement by LFA may be a viable alternative to evaluate thyroid diseases, but fT4 by LFA is not precise and presents a high bias.
引用
收藏
页数:8
相关论文