Performance of a Dual-Sensor Cardiopulmonary Resuscitation Feedback System in Estimating Compression Depth and Rate During Simulated Chest Compressions

被引:0
作者
Gould, Jeffrey R. [1 ]
Mostofi, Naghmeh [2 ]
Badin, Jill K. [1 ]
Giacometti, Paolo [2 ]
Lampe, Joshua W. [1 ]
机构
[1] ZOLL Med Corp, Sci Affairs, Chelmsford, MA 01824 USA
[2] ZOLL Med Corp, Adv Dev, Chelmsford, MA 01824 USA
来源
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME | 2025年 / 19卷 / 02期
关键词
cardiopulmonary resuscitation; CPR feedback; chest compressions; compression depth; accelerometry; accuracy; HEART-ASSOCIATION GUIDELINES; ADVANCED LIFE-SUPPORT; QUALITY; SURVIVAL;
D O I
10.1115/1.4068089
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study evaluated the performance of a dual-sensor cardiopulmonary resuscitation (CPR) feedback system in estimating chest compression depth and rate over a range of depth and rate combinations during rigid and compliant surface conditions using a computer-controlled motion system to simulate chest compressions. Ten dual-sensor CPR pads were tested using a computer-controlled motion system which simulated chest compressions at target depths of 1.9, 3.8, 4.8, 6.4, and 8.9 cm and target rates of 60, 80, 100, 120, and 140 compressions per minute (cpm). A rigid surface was simulated by applying motion only to the anterior sensor, and a compliant surface was simulated by applying motion to both the anterior and posterior sensor, challenging the algorithm to calculate a net compression depth by subtracting motion of the posterior sensor. For all simulated compressions, including every rate and depth combination and for both rigid and compliant surface simulations, the mean (+/- sd) depth error was 0.05 (+/- 0.08) cm and the rate error was -0.55 (+/- 1.44) cpm. A dual-sensor CPR system accurately estimates compression depth within +/- 0.25 cm and compression rate within +/- 3 cpm over a wide range and combination of clinically relevant chest compression depths and rates during both rigid and compliant surface simulations. Use of a computer-controlled motion system provides a more direct assessment of accuracy than manual compressions performed on instrumented manikins.
引用
收藏
页数:6
相关论文
共 24 条
[1]   Compression depth estimation for CPR quality assessment using DSP on accelerometer signals [J].
Aase, SO ;
Myklebust, H .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (03) :263-268
[2]   Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest [J].
Abella, BS ;
Alvarado, JP ;
Myklebust, H ;
Edelson, DP ;
Barry, A ;
O'Hearn, N ;
Vanden Hoek, TL ;
Becker, LB .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 293 (03) :305-310
[3]   Part 11: Pediatric Basic Life Support and Cardiopulmonary Resuscitation Quality 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care [J].
Atkins, Dianne L. ;
Berger, Stuart ;
Duff, Jonathan P. ;
Gonzales, John C. ;
Hunt, Elizabeth A. ;
Joyner, Benny L. ;
Meaney, Peter A. ;
Niles, Dana E. ;
Samson, Ricardo A. ;
Schexnayder, Stephen M. .
CIRCULATION, 2015, 132 (18) :S519-S525
[4]   Accurate feedback of chest compression depth on a manikin on a soft surface with correction for total body displacement [J].
Beesems, Stefanie G. ;
Koster, Rudolph W. .
RESUSCITATION, 2014, 85 (11) :1439-1443
[5]   The Influence of Scenario-Based Training and Real-Time Audiovisual Feedback on Out-of-Hospital Cardiopulmonary Resuscitation Quality and Survival From Out-of-Hospital Cardiac Arrest [J].
Bobrow, Bentley J. ;
Vadeboncoeur, Tyler F. ;
Stolz, Uwe ;
Silver, Annemarie E. ;
Tobin, John M. ;
Crawford, Scott A. ;
Mason, Terence K. ;
Schirmer, Jerome ;
Smith, Gary A. ;
Spaite, Daniel W. .
ANNALS OF EMERGENCY MEDICINE, 2013, 62 (01) :47-56
[6]   Measuring and improving cardiopulmonary resuscitation quality inside the emergency department [J].
Crowe, Christopher ;
Bobrow, Bentley J. ;
Vadeboncoeur, Tyler F. ;
Dameff, Christian ;
Stolz, Uwe ;
Silver, Annemarie ;
Roosa, Jason ;
Page, Rianne ;
LoVecchio, Frank ;
Spaite, Daniel W. .
RESUSCITATION, 2015, 93 :8-13
[7]   Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest [J].
Edelson, Dana P. ;
Abella, Benjamin S. ;
Kramer-Johansen, Jo ;
Wik, Lars ;
Myklebust, Helge ;
Barry, Anne M. ;
Merchant, Raina M. ;
Vanden Hoek, Terry L. ;
Steen, Petter A. ;
Becker, Lance B. .
RESUSCITATION, 2006, 71 (02) :137-145
[8]   Effect of real-time feedback during cardiopulmonary resuscitation outside hospital: prospective, cluster-randomised trial [J].
Hostler, David ;
Everson-Stewart, Siobhan ;
Rea, Thomas D. ;
Stiell, Ian G. ;
Callaway, Clifton W. ;
Kudenchuk, Peter J. ;
Sears, Gena K. ;
Emerson, Scott S. ;
Nichol, Graham .
BRITISH MEDICAL JOURNAL, 2011, 342 :371
[9]   Quality of out-of-hospital cardiopulmonary resuscitation with real time automated feedback: A prospective interventional study [J].
Kramer-Johansen, Jo ;
Myklebust, Helge ;
Wik, Lars ;
Fellows, Bob ;
Svensson, Leif ;
Sorebo, Hallstein ;
Steen, Petter Andreas .
RESUSCITATION, 2006, 71 (03) :283-292
[10]   Mechanical chest compressions with trapezoidal waveform improve haemodynamics during cardiac arrest [J].
Kramer-Johansen, Jo ;
Pytte, Morten ;
Tomlinson, Ann-Elin ;
Sunde, Kjetil ;
Dorph, Elizabeth ;
Svendsen, Jan Vegard H. ;
Eriksen, Morten ;
Stromme, Taevje A. ;
Wik, Lars .
RESUSCITATION, 2011, 82 (02) :213-218