Timing-synchronized passive ultrasound imaging of cavitation using eigenspace-based minimum variance beamforming and principal component analysis

被引:0
作者
Lu, Shukuan [1 ]
Su, Ruibo [1 ]
Ma, Yingping [1 ]
Wan, Mingxi [1 ]
机构
[1] Xi An Jiao Tong Univ, Dept Biomed Engn, Sch Life Sci & Technol, Key Lab Biomed Informat Engn,Minist Educ, Xian 710049, Peoples R China
基金
中国国家自然科学基金;
关键词
adaptive beamforming; cavitation; principal component analysis; timing-synchronized passive ultrasound imaging; SHORT PULSES; LOCALIZATION; HISTOTRIPSY; THERAPY;
D O I
10.1002/mp.17853
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundPassive ultrasound imaging (PUI) allows to spatially resolve cavitation triggered during ultrasound irradiation, its application in therapeutic ultrasound has been gaining attention in recent years. The diffraction mode of the imaging transducer greatly limits the PUI axial resolution, which can be improved by transmit-receive synchronization and employment of delay sum beamforming (DSB) when transmitting short pulses, however, DSB yields poor performance in resolution and anti-interference.PurposeInspired by adaptive beamforming and its low-complexity algorithm in active imaging field, this paper aims to develop an improved timing-synchronized PUI (TSPUI) algorithm for detection of short-pulse transmission-induced cavitation.MethodsThe passive array data collected by timing synchronization is processed by minimum variance beamforming (MVB), whose weights are optimized by projection on the eigendecomposed signal subspace, that is, eigenspace-based MVB (EMVB), with the sum of the flight times on the transmitting and receiving paths as the delay. Applying principal component analysis (PCA) on the pre-collected MVB weight samples, a conversion matrix is constructed to allow the matrix inversion and eigendecomposition involved in weight calculation to be performed in a low dimension. The algorithm performance is confirmed by experiments, where a high-intensity focused ultrasound transducer and a linear-array transducer configured in a common parallel or vertical manner are employed for cavitation induction and cavitation imaging, and evaluated with the established indicators.ResultsReducing the eigenvalue threshold coefficient allows more sidelobes to be removed, and choosing an appropriate principal component number can reduce the time cost while guaranteeing the reconstruction quality. EMVB-PCA provides high resolution and anti-interference performance relative to DSB, with a reduction of over 60% in the point spread area and over 14 dB in the sidelobe and noise level, meanwhile, its time cost is considerably lower than EMVB, with a reduction of over 80%. Additionally, constructing the conversion matrix by simulation is feasible and valid, providing convenience for real imaging.ConclusionsEMVB-PCA allows for high-quality TSPUI reconstruction of cavitation at a fast rate, providing an effective tool for detecting short-duration cavitation and further benefiting short-pulse therapeutic ultrasound applications.
引用
收藏
页数:14
相关论文
共 25 条
[1]   Post Hoc Analysis of Passive Cavitation Imaging for Classification of Histotripsy-Induced LIquefaction in Vitro [J].
Bader, Kenneth B. ;
Haworth, Kevin J. ;
Maxwell, Adam D. ;
Holland, Christy K. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (01) :106-115
[2]   Pulse-Inversion Subharmonic Ultrafast Active Cavitation Imaging in Tissue Using Fast Eigenspace-Based Adaptive Beamforming and Cavitation Deconvolution [J].
Bai, Chen ;
Xu, Shanshan ;
Duan, Junbo ;
Jing, Bowen ;
Yang, Miao ;
Wan, Mingxi .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (08) :1175-1193
[3]   Power cavitation-guided blood-brain barrier opening with focused ultrasound and microbubbles [J].
Burgess, M. T. ;
Apostolakis, I. ;
Konofagou, E. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (06)
[4]   Noninvasive and localized neuronal delivery using short ultrasonic pulses and microbubbles [J].
Choi, James J. ;
Selert, Kirsten ;
Vlachos, Fotios ;
Wong, Anna ;
Konofagou, Elisa E. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2011, 108 (40) :16539-16544
[5]   Passive acoustic mapping utilizing optimal beamforming in ultrasound therapy monitoring [J].
Coviello, Christian ;
Kozick, Richard ;
Choi, James ;
Gyoengy, Miklos ;
Jensen, Carl ;
Smith, Penny Probert ;
Coussios, Constantin-C. .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2015, 137 (05) :2573-2585
[6]   Imaging With Therapeutic Acoustic Wavelets-Short Pulses Enable Acoustic Localization When Time of Arrival is Combined With Delay and Sum [J].
Davies, Harry J. ;
Morse, Sophie, V ;
Copping, Matthew J. ;
Sujarittam, Krit ;
Bourgin, Victor D. ;
Tang, Meng-Xing ;
Choi, James J. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (01) :178-190
[7]   2D and 3D real-time passive cavitation imaging of pulsed cavitation ultrasound therapy in moving tissues [J].
Escudero, Daniel Suarez ;
Goudot, Guillaume ;
Vion, Michael ;
Tanter, Mickael ;
Pernot, Mathieu .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (23)
[8]   Combined Passive Detection and Ultrafast Active Imaging of Cavitation Events Induced by Short Pulses of High-Intensity Ultrasound [J].
Gateau, Jerome ;
Aubry, Jean-Francois ;
Pernot, Mathieu ;
Fink, Mathias ;
Tanter, Mickael .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (03) :517-532
[9]   Passive cavitation mapping for localization and tracking of bubble dynamics [J].
Gyoengy, Miklos ;
Coussios, Constantin-C. .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2010, 128 (04) :E175-E180
[10]   Passive Spatial Mapping of Inertial Cavitation During HIFU Exposure [J].
Gyongy, Miklos ;
Coussios, Constantin-C. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (01) :48-56