Consequences of Pausing Heated Humidification During Invasive Ventilation

被引:0
作者
Lellouche, Francois [1 ]
Bouchard, Pierre-Alexandre [1 ]
机构
[1] Univ Laval, Ctr Rech Inst Univ Cardiol & Pneumol Quebec, 2725 Chemin St Foy, Quebec City, PQ G1V 4G5, Canada
关键词
nebulization; heated humidifier; absolute humidity; psychrometry; endotracheal tube occlusions; METERED-DOSE INHALER; MECHANICAL VENTILATION; MOISTURE EXCHANGERS; AEROSOL DELIVERY; IN-VITRO; CARE; PERFORMANCE; HUMIDIFIERS; DEVICES;
D O I
10.4187/respcare.12084
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: During invasive mechanical ventilation, where medical gases are very dry and the upper airway is bypassed, appropriate gas conditioning and humidification are mandatory at all times. Results of in vitro studies suggest that dry gases may improve lung deposition during nebulization, but this has not been confirmed through in vivo studies. The objective of this study was to measure gas humidity under multiple conditions to better describe gas hygrometry when heated humidifiers are turned off. METHODS: We measured, on a bench, the hygrometry of different gases at steady state: medical gases, at the Y-piece without humidifier, with the humidifier switched off, and with humidifier switched on. We measured gas humidity every 10-60 s during dynamic conditions after switching off the heated humidifier and after switching on the heated humidifier. Hygrometry was measured by using the psychrometric method with at least 3 measurements for each tested condition. RESULTS: We performed 287 psychrometric measurements in different situations. The mean 6 SD gas absolute humidity at steady state during different conditions were the following: 1.6 6 0.2 mg H2O/L for the medical gases, 4.5 6 0.9 mg H2O/L at the Y-piece without humidifier, 9.1 6 0.3 mg H2O/L at the Y-piece with heated humidifier turned off, and 34.2 6 2.2 mg H2O/L at the Y-piece with the heated humidifier turned on. During the dynamic evaluation, after turning off the humidifier, humidity was < 30 mg H2O/L after a few minutes, attained 15 mg H2O/L after 15 min, and was below 10 mg H2O/L after 1 h but never reached the level of dry medical gases. After turning on the heated humidifier, the gas hygrometry reached 30 mg H2O/L after 5 min. CONCLUSIONS: When heated humidifiers are turned off, gas humidity levels are very low but not as low as medical gases. The clinical impact of repeated shutdowns is unknown. As recommended, heated humidifiers should never be turned off during nebulization.
引用
收藏
页码:1239 / 1244
页数:6
相关论文
共 33 条
  • [1] Lellouche F., Humidification during invasive mechanical ventilation, Humidification in the intensive care-the essentials Switzerland, pp. 97-120, (2023)
  • [2] Ari A, Areabi H, Fink JB., Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation, Respir Care, 55, 7, pp. 837-844, (2010)
  • [3] Diot P, Morra L, Smaldone GC., Albuterol delivery in a model of mechanical ventilation. Comparison of metered-dose inhaler and nebulizer efficiency, Am J Respir Crit Care Med, 152, 4, pp. 1391-1394, (1995)
  • [4] Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ., Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors, Am J Respir Crit Care Med, 159, 1, pp. 63-68, (1999)
  • [5] Li J, Liu K, Lyu S, Jing G, Dai B, Dhand R, Et al., Aerosol therapy in adult critically ill patients: a consensus statement regarding aerosol administration strategies during various modes of respiratory support, Ann Intensive Care, 13, 1, (2023)
  • [6] Dugernier J, Le Pennec D, Maerckx G, Allimonnier L, Hesse M, Castanares-Zapatero D, Et al., Inhaled drug delivery: a randomized study in intubated patients with healthy lungs, Ann Intensive Care, 13, 1, (2023)
  • [7] Moustafa IOF, Ali MRA-A, Al Hallag M, Rabea H, Fink JB, Dailey P, Abdelrahim MEA., Lung deposition and systemic bioavailability of different aerosol devices with and without humidification in mechanically ventilated patients, Heart Lung, 46, 6, pp. 464-467, (2017)
  • [8] Fink JB, Dhand R, Duarte AG, Jenne JW, Tobin MJ., Aerosol delivery from a metered-dose inhaler during mechanical ventilation. An in vitro model, Am J Respir Crit Care Med, 154, 2, pp. 382-387, (1996)
  • [9] Lellouche F, Taille S, Lefrancois F, Deye N, Maggiore SM, Jouvet P, Et al., Humidification performance of 48 passive airway humidifiers: comparison with manufacturer data, Chest, 135, 2, pp. 276-286, (2009)
  • [10] Lellouche F, Taille S, Maggiore SM, Qader S, L'her E, Deye N, Brochard L., Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers, Am J Respir Crit Care Med, 170, 10, pp. 1073-1079, (2004)