Comparison of the Point-Of-Care Coagulation Device Quantra With the TEG-5000 for Congenital Cardiac Surgery-A Pilot Study

被引:0
|
作者
Ochocinski, Dominik [1 ,2 ]
Navaratnam, Manchula [1 ]
Babb, Amy [1 ]
De Souza, Elizabeth [1 ]
Kamra, Komal [1 ]
Sleasman, Justin R. [3 ]
Margetson, Tristan Day [3 ]
Bhamidipati, Jyoti K. [1 ]
Ramamoorthy, Chandra [1 ]
Schmidt, Alexander R. [1 ,4 ,5 ]
机构
[1] Stanford Univ, Sch Med, Dept Anesthesiol Perioperat & Pain Med, Stanford, CA 94305 USA
[2] Univ Calif San Francisco, Dept Anesthesia & Perioperat Care, San Francisco, CA USA
[3] Stanford Med Childrens Hlth, Perfus Perioperat Serv, Palo Alto, CA USA
[4] Univ Childrens Hosp, Dept Anesthesia, Zurich, Switzerland
[5] Univ Childrens Hosp, Childrens Res Ctr, Zurich, Switzerland
关键词
coagulation; pediatric cardiac anesthesia; point-of-care; validation; SEER SONORHEOMETRY; THROMBOELASTOGRAPHY; TRANSFUSION; COMPONENT; FIBRINOGEN; CHILDREN; SYSTEM; IMPACT;
D O I
10.1111/pan.15097
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BackgroundCongenital cardiac surgery with cardiopulmonary bypass (CPB) alters patients' hemostasis. Viscoelastic testing is a modern technology identifying coagulation abnormalities. A new device, the Quantra-QPlus System (HemoSonics LLC, Charlottesville, VA) has not yet been investigated during congenital cardiac surgery.AimThis prospective observational pilot study investigated the correlation of Quantra and TEG-5000 in children undergoing cardiac surgery at Stanford Medicine Children's Health (SMCH).MethodsPatients (0-8 years) undergoing cardiac surgery with CPB were included after parental consent. Per standard of care at SMCH, a TEG-5000 and Clauss Fibrinogen were measured on CPB during rewarming (T1) as well as after administration of blood components and coagulation factors (T2). For the study purpose, Quantra measurements were performed simultaneously at T1 and T2. Quantra results were correlated with TEG-5000 results and Clauss Fibrinogen. In addition, the agreement for normal and abnormal results was calculated, and a post hoc simulated transfusion algorithm using TEG-5000 and Quantra thresholds was compared to patients' clinical management.ResultsFrom October 2022 to May 2023, 289 congenital cardiac surgeries were performed, 97 met inclusion criteria, 63 patients were consented, and 40 patients (12 females [30%] 28 males [70%]) included in the analysis. Median age and weight were 0.5 years and 6.33 kg, respectively. Correlation of Quantra/Clauss Fibrinogen was "moderate" and correlation of Quantra/TEG-5000 parameters ranged from "weak" to "very strong" Levels of agreement ranged from 15% to 97%. The post hoc simulated transfusion algorithm for TEG-5000 showed an agreement of 56% for FFP, 56% for Fibrinogen, and 95% for platelets and for Quantra, it was 40% for FFP, 87.5% for Fibrinogen, and 98% for platelets.ConclusionsDespite weaknesses in correlation and agreement both VET devices suggested appropriate coagulation management based on the simulated transfusion algorithms. Quantra and TEG-5000 are not interchangeable and none of them can be considered as "the gold standard".Trial RegistrationNTC 05295693
引用
收藏
页数:12
相关论文
共 16 条
  • [1] Point-of-Care Viscoelastic Hemostatic Assays in Cardiac Surgery Patients: Comparison of Thromboelastography 6S, Thromboelastometry Sigma, and Quantra
    Demailly, Zoe
    Wurtz, Veronique
    Barbay, Virginie
    Surlemont, Elisabeth
    Scherrer, Vincent
    Billoir, Paul
    Clavier, Thomas
    Besnier, Emmanuel
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2023, 37 (06) : 948 - 955
  • [2] Clinical utility of the Quantra® point-of-care haemostasis analyser during urgent cardiac surgery
    Zghaibe, W.
    Scheuermann, S.
    Munting, K.
    Blaudszun, G.
    Besser, M.
    Ortmann, E.
    Klein, A. A.
    ANAESTHESIA, 2020, 75 (03) : 366 - 373
  • [3] Point-of-care haemostasis and coagulation monitoring in cardiac surgery at IRCCS Policlinico San Donato
    Baryshnikova, Ekaterina
    Ranucci, Marco
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2016, 18 (0E) : E42 - E48
  • [4] RETRACTED ARTICLE: Point-of-care (POC) measurement of coagulation after cardiac surgery
    J. Boldt
    G. Walz
    J. Triem
    S. Suttner
    B. Kumle
    Intensive Care Medicine, 1998, 24 : 1187 - 1193
  • [5] Comparison of three point-of-care testing devices to detect hemostatic changes in adult elective cardiac surgery: a prospective observational study
    Espinosa, Aurora
    Stenseth, Roar
    Videm, Vibeke
    Pleym, Hilde
    BMC ANESTHESIOLOGY, 2014, 14
  • [6] RETRACTED: Point-of-care (POC) measurement of coagulation after cardiac surgery (Retracted Article)
    Boldt, J
    Walz, G
    Triem, J
    Suttner, S
    Kumle, B
    INTENSIVE CARE MEDICINE, 1998, 24 (11) : 1187 - 1193
  • [7] Evaluation of the GEM®PCL Plus point-of-care device for neonatal coagulation assessment: An observational study on cord blood
    De Luca, Roberta
    Fontana, Pierre
    Poncet, Antoine
    De Moerloose, Philippe
    Pfister, Riccardo E.
    THROMBOSIS RESEARCH, 2014, 134 (02) : 474 - 478
  • [8] Comparison of three point-of-care testing devices to detect hemostatic changes in adult elective cardiac surgery: a prospective observational study
    Aurora Espinosa
    Roar Stenseth
    Vibeke Videm
    Hilde Pleym
    BMC Anesthesiology, 14
  • [9] Comparison of international normalized ratio determined by point-of-care to standard laboratory testing before and after reversal of heparin in cardiac surgery
    Fabbro, Michael, II
    Abalo, Miguel
    Fermin, Lilibeth
    Andrews, David M.
    Patel, Prakash A.
    BLOOD COAGULATION & FIBRINOLYSIS, 2020, 31 (02) : 140 - 144
  • [10] Point-of-care thromboelastography/thromboelastometry-based coagulation management in cardiac surgery: a meta-analysis of 8332 patients
    Deppe, Antje-Christin
    Weber, Carolyn
    Zimmermann, Julia
    Kuhn, Elmar W.
    Slottosch, Ingo
    Liakopoulos, Oliver J.
    Choi, Yeong-Hoon
    Wahlers, Thorsten
    JOURNAL OF SURGICAL RESEARCH, 2016, 203 (02) : 424 - 433