Finite Element Analysis of the Stress and Deformation in Bone Caused by CFR-PEEK, Titanium, and Zirconia Ceramic Implants

被引:1
作者
Jameel, Ahmad Saib [1 ]
Sahoo, Amiya Ranjan [2 ]
Kharkar, Viraj [3 ]
Kalsi, Harjit Singh [3 ]
Singh, Akanksha [4 ]
Chougule, Vaibhav T. [5 ]
Behera, Sharmila [6 ]
机构
[1] Sanjeevni Hosp, Consultant Prosthodont & Maxillofacial Prosthet, New Delhi, India
[2] Awadh Dent Coll, Dept Oral & Maxillofacial Surg, Jemshedpur, India
[3] Bharati Vidyapeeth Deemed Univ, Dent Coll & Hosp, Dept Oral & Maxillofacial Surg, Navi Mumbai, Maharashtra, India
[4] Inst Dent Sci, Dept Periodont, Associate Prof, PhD Scholar, Bareilly, Uttar Pradesh, India
[5] Bharati Vidyapeeth Deemed Univ Dent Coll & Hos, Dept Paediat & Prevent Dent, Sangli, Maharastra, India
[6] KIIT Deemed Univ, Kalinga Inst Dent Sci, Dept Oral Med & Radiol, Bhubaneswar, Odisha, India
关键词
CFR-PEEK; dental implant; finite analysis; stress; titanium; zirconia;
D O I
10.4103/jpbs.jpbs_45_24
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Introduction: Glass or carbon fiber-reinforced polyetheretherketone (CFR-PEEK) composites and zirconia ceramic have been offered as more modern implant biomaterials. Objectives: This study examined the effects of zirconia ceramic implants, titanium, and CFR-PEEK on bone stress and deformation. Materials and Methods: A geometric model of a mandibular molar replaced with an implant-supported crown was created for this in vitro finite element analysis investigation. The dimensions of the implant employed in the study are 11.0 mm in length and 4.5 mm in diameter. Using finite element analysis, three implant assemblies composed of zirconium, titanium, and CFR-PEEK were produced. 150 N loads were applied both vertically and obliquely on the long axis of the implant. The von-Mises stresses and deformation were compared using ANSYS Workbench 18.0 and finite element software, with a significance threshold of P < 0.05. Results: All three implant assemblies-zirconia, titanium, and CFR-PEEKshowed comparable stresses and deformation in bone with no appreciable variations. Conclusion: Zirconia and PEEK reinforced with carbon fibers (CFR-PEEK) have been found to be viable replacements for titanium in implant biomaterials.
引用
收藏
页码:S2078 / S2080
页数:3
相关论文
共 13 条
[1]  
Bathala Lakshmana, 2019, J Med Life, V12, P5, DOI [10.25122/jml-2019-0003, 10.25122/jml-2019-0003]
[2]   Polyetheretherketone (PEEK) in Dentistry [J].
Benakatti, Veena B. ;
Sajjanar, Jayashree A. ;
Acharya, Aditya .
JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2019, 13 (08) :ZE10-ZE12
[3]   Osseointegration of zirconia implants compared with titanium: An in vivo study [J].
Depprich R. ;
Zipprich H. ;
Ommerborn M. ;
Naujoks C. ;
Wiesmann H.-P. ;
Kiattavorncharoen S. ;
Lauer H.-C. ;
Meyer U. ;
Kübler N.R. ;
Handschel J. .
Head & Face Medicine, 4 (1)
[4]   SUSPECTED ASSOCIATION OF AN ALLERGIC REACTION WITH TITANIUM DENTAL IMPLANTS: A CLINICAL REPORT [J].
Egusa, Hiroshi ;
Ko, Nagakazu ;
Shimazu, Tsunetoshi ;
Yatani, Hirofumi .
JOURNAL OF PROSTHETIC DENTISTRY, 2008, 100 (05) :344-347
[5]   Application of finite element analysis in implant dentistry: A review of the literature [J].
Geng, JP ;
Tan, KBC ;
Liu, GR .
JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (06) :585-598
[6]   Finite element analysis to assess stress and deformation in bone with glass fiber-reinforced-poly-ether-ether-ketone, zirconia, and titanium implants [J].
Haseeb, Syeda Amtul ;
Kumar, Anju S. ;
Chaitra, M. P. ;
Vinaya, K. C. ;
Gudal, Soumya Shivananda ;
Rahmam, Fathima Parveen ;
Babaji, Prashant .
TZU CHI MEDICAL JOURNAL, 2023, 35 (03) :231-236
[7]   Osseointegration of one-piece zirconia implants compared with a titanium implant of identical design: a histomorphometric study in the dog [J].
Koch, F. P. ;
Weng, D. ;
Kraemer, S. ;
Biesterfeld, S. ;
Jahn-Eimermacher, A. ;
Wagner, W. .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (03) :350-356
[8]  
Mahalakshmi G, 2023, Dent Res J (Isfahan), V20, P57
[9]  
Moeen F, 2014, Pakistan Oral and Dental Journal, V34, P164
[10]  
Rahmitasari Fitria, 2017, Dent J (Basel), V5, DOI [10.3390/dj5040035, 10.3390/dj5040035]