Blood Pressure Difference Between Cuff Inflation and Deflation by Auscultatory Method: Impact of Hypertension Grade

被引:0
|
作者
Coccina, Francesca [1 ]
Pizzicannella, Jacopo [2 ]
Trubiani, Oriana [1 ]
Pierdomenico, Sante D. [1 ]
机构
[1] Univ Gabriele dAnnunzio, Dept Innovat Technol Med & Dent, I-66100 Chieti, Italy
[2] Univ Gabriele dAnnunzio, Dept Engn & Geol, I-66100 Chieti, Italy
关键词
blood pressure; measurement; auscultatory method; hypertension; PAIN PERCEPTION; MANAGEMENT;
D O I
10.3390/diagnostics15060687
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The aim of this study was to evaluate blood pressure (BP) difference between cuff inflation and deflation and to investigate whether hypertension grade and patients' characteristics could be influential. Methods: We selected 328 asymptomatic patients, 219 with mild systolic/diastolic hypertension, 75 with moderate systolic/diastolic hypertension and 34 with severe systolic/diastolic hypertension. Results: Hypertension grade was a determinant of the difference between systolic/diastolic BP during cuff inflation and deflation; the difference was progressively and significantly higher from mild to moderate to severe hypertension (1.80 +/- 1.03/1.21 +/- 0.56 vs. 5.32 +/- 1.09/3.04 +/- 0.81 vs. 9.74 +/- 1.46/4.88 +/- 0.73 mmHg, respectively (all differences were significant). Age, gender, body mass index, smoking habits and laboratory parameters were not associated with BP differences. The observed differences led to a reclassification of 24% of patients with moderate and 32% of patients with severe hypertension to a lower grade, but all were classified as hypertensive patients during cuff inflation. Conclusions: Hypertension grade influences the difference in systolic/diastolic BP during cuff inflation and deflation. This difference leads to a reclassification of hypertension grade during cuff inflation within the hypertensive range but does not influence the definition of hypertensive status. Future studies are needed to confirm whether the differences in systolic/diastolic BP between cuff inflation and deflation are due to BP increases induced by sympathetic activation, as well as a potential different behavior of the brachial artery during closing or opening, or both.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Blood Pressure Difference between the Measurements taken during Cuff Inflation and Deflation
    Zheng, Dingchang
    Di Marco, Luigi Y.
    Murray, Alan
    2012 COMPUTING IN CARDIOLOGY (CINC), VOL 39, 2012, 39 : 729 - 732
  • [2] Oscillometric Waveform Difference between Cuff Inflation and Deflation during Blood Pressure Measurement
    Liu, Chengyu
    Zheng, Dingchang
    Griffiths, Clive
    Murray, Alan
    2014 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), VOL 41, 2014, 41 : 849 - 852
  • [3] THE EFFECT OF CUFF PRESSURE DEFLATION RATE ON ACCURACY IN INDIRECT MEASUREMENT OF BLOOD-PRESSURE WITH THE AUSCULTATORY METHOD
    YONG, PG
    GEDDES, LA
    JOURNAL OF CLINICAL MONITORING, 1987, 3 (03): : 155 - 159
  • [5] Evaluation of the oscillometric method for noninvasive blood pressure measurement during cuff deflation and cuff inflation with reference to intra-arterial blood pressure
    Celler, Branko G.
    Yong, Andy
    Rubenis, Imants
    Butlin, Mark
    Argha, Ahmadreza
    Rehan, Rajan
    Avolio, Alberto
    JOURNAL OF HYPERTENSION, 2024, 42 (07) : 1235 - 1247
  • [6] Effect of mechanical behavior of the brachial artery on blood pressure measurement during cuff inflation and cuff deflation
    Zheng, Dingchang
    Pan, Fan
    Murray, Alan
    BLOOD PRESSURE MONITORING, 2014, 19 (02) : 120 - 121
  • [7] Comparison of cuff inflation and cuff deflation brachial sphygmomanometry with intra-arterial blood pressure as reference
    Celler, Branko G.
    Yong, Andy
    Rubenis, Imants
    Butlin, Mark
    Argha, Ahmadreza
    Rehan, Rajan
    Avolio, Alberto
    JOURNAL OF HYPERTENSION, 2024, 42 (06) : 968 - 976
  • [8] Effect of mechanical behaviour of the brachial artery on blood pressure measurement during both cuff inflation and cuff deflation
    Zheng, Dingchang
    Pan, Fan
    Murray, Alan
    BLOOD PRESSURE MONITORING, 2013, 18 (05) : 265 - 271
  • [9] A Blood Pressure Monitor with Robust Noise Reduction System under Linear Cuff Inflation and Deflation
    Usuda, Takashi
    Kobayashi, Naoki
    Takeda, Sunao
    Kotake, Yoshifumi
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 1226 - 1229
  • [10] Measurement of non-invasive oscillometric blood pressure during cuff inflation, rather than deflation
    Morioka, N
    Ozaki, M
    Ozaki, K
    Sessler, DI
    Matsukawa, T
    Suzuki, H
    ANESTHESIOLOGY, 1997, 87 (03) : A431 - A431