Monte Carlo calculated beam quality correction factors for high energy photon and electron fields

被引:0
|
作者
Czarnecki, Damian [1 ]
Zink, Klemens [1 ,2 ]
机构
[1] Univ Appl Sci THM, Inst Med Phys & Radiat Protect, Wiesenstr 14, D-35390 Giessen, Germany
[2] Marburg Univ Hosp, Dept Radiotherapy & Radiooncol, D-35043 Marburg, Germany
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2025年 / 131卷
关键词
Dosimetry; Beam quality correction factor; Ionization chamber; Monte carlo simulation; REFERENCE DOSIMETRY;
D O I
10.1016/j.ejmp.2025.104939
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Although several studies provide data on reference dosimetry, ionization chambers are in clinical use worldwide, for which no or only a few correction factors for beam quality are available. The aim of this study was to calculate beam quality correction factors in electron and photon fields for four cylindrical ionization chamber, i.e. CC08, CC04, Razor T Chamber and Razor T Nano Chamber (all IBA, Dosimetry, Schwarzenbruck, Germany). Method: Monte Carlo simulations were performed with the EGSnrc code system to calculate the beam quality correction factor k(Q). Therefore, a detailed Monte Carlo based model of the four ionization chambers were created to calculate the dose in the sensitive volume. Electron and photon spectra of clinical linear accelerators within the entire energy range of therapeutic linear accelerators were used as radiation source. Results: k(Q) values as a function of the beam quality specifiers TPR20,10 and R-50 were calculated for high energy photon and electron fields, respectively. The data was fitted against recommended equations for photon and electron dosimetry with a chi(2) between 6.4e-7 and 4.2e-6. Conclusion: Electron and photon beam quality correction factors k(Q) were calculated by Monte Carlo simulations for the cylindrical ionization chambers CC08, CC04, Razor (TM) Chamber and Razor (TM) Nano Chamber to provide k(Q) values for absolute dosimetry according to the TRS 398 dosimetry protocol.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Monte Carlo calculated and experimentally determined output correction factors for small field detectors in Leksell Gamma Knife Perfexion beams
    Benmakhlouf, H.
    Johansson, J.
    Paddick, I.
    Andreo, P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (10) : 3959 - 3973
  • [32] A study of the beam quality correction factor in clinical proton beam conditions using Monte Carlo simulations
    Yong-Cheol Kwon
    Sang-Il Pak
    Hyon-Suk Jo
    Seonghoon Jeong
    Se Byeong Lee
    In Jung Kim
    Chul-Young Yi
    Wook-Geun Shin
    Journal of the Korean Physical Society, 2022, 81 : 1128 - 1134
  • [33] Beam hardening correction based on Monte Carlo simulation
    Zeng, G
    Yu, ZQ
    Yan, YL
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2006, 30 (02): : 178 - 182
  • [34] Monte Carlo source model for photon beam radiotherapy: photon source characteristics
    Fix, MK
    Keall, PJ
    Dawson, K
    Siebers, JV
    MEDICAL PHYSICS, 2004, 31 (11) : 3106 - 3121
  • [35] Electron beam quality kQ,Q0 factors for various ionization chambers: a Monte Carlo investigation with PENELOPE
    Erazo, F.
    Brualla, L.
    Lallena, A. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (21) : 6673 - 6691
  • [36] Estimation of the cable effect in megavoltage photon beam by measurement and Monte Carlo simulation
    Yamauchi, Ryohei
    Igari, Mitsunobu
    Kasai, Yuya
    Hariu, Masatsugu
    Suda, Yuhi
    Kawachi, Toru
    Katayose, Tetsurou
    Mizuno, Norifumi
    Miyasaka, Ryohei
    Saitoh, Hidetoshi
    MEDICAL PHYSICS, 2020, 47 (10) : 5324 - 5332
  • [37] Monte Carlo calculation of the photon beam quality correction factor kQ, Q0 for ionization chambers of very small volume: Use of variance reduction techniques driven with an ant colony algorithm
    Garvi, A. M.
    Zamora, L. I.
    Anguiano, M.
    Lallena, A. M.
    Garcia-Pareja, S.
    RADIATION PHYSICS AND CHEMISTRY, 2024, 225
  • [38] Optimization of low-energy electron beam proximity lithography stencil mask structure factors by Monte Carlo simulation
    Kim, C
    Kang, IY
    Chung, YC
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2004, 43 (03): : 1196 - 1198
  • [39] The replacement correction factors for cylindrical chambers in high-energy photon beams
    Wang, L. L. W.
    Rogers, D. W. O.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (06) : 1609 - 1620
  • [40] Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy
    Fix, MK
    Keller, H
    Rüegsegger, P
    Born, EJ
    MEDICAL PHYSICS, 2000, 27 (12) : 2739 - 2747