Does Leak Matter? A Novel Dynamic Leak Model to Simulate Leak for Performance Testing of Manual Neonatal Resuscitation Devices. A Bench Study

被引:0
作者
Morakeas, Stephanie [1 ,2 ]
Tracy, Mark Brian [1 ,3 ]
Hinder, Murray [1 ,3 ]
Gruber, Viktoria [4 ]
Mcewan, Alistair [2 ]
Drevhammar, Thomas [5 ]
机构
[1] Westmead Hosp, Neonatal Intens Care Unit, Sydney, NSW, Australia
[2] Univ Sydney, BMET Inst, Fac Engn & Informat Technol, Sydney, NSW, Australia
[3] Univ Sydney, Dept Paediat & Child Hlth, Sydney, NSW, Australia
[4] Med Univ Graz, Dept Paediat & Adolescent Med, Div Neonatol, Graz, Austria
[5] Karolinska Inst, Dept Womens & Childrens Hlth, Stockholm, Sweden
关键词
leak; mask leak; neonatal resuscitation; positive pressure ventilation; MASK VENTILATION; VOLUME; COMPENSATION; MECHANICS;
D O I
10.1002/ppul.71043
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
IntroductionNewborn resuscitation is commonly performed in the presence of face mask leak. Leak is highly variable, pressure-dependent and often unrecognized. The effectiveness of resuscitation devices to deliver adequate inflations in the presence of leak is unknown. Bench models simulating continuous leak have the disadvantage of not accurately reflecting leak occurring during clinical resuscitation. A dynamic leak model based on pressure-release valves was thus developed.MethodsThis study investigates self-inflating bag (SIB) and T-piece resuscitator (TPR) ventilation performance in the presence of dynamic (DLM) compared to continuous (CLM) leak models in a bench study. Five predefined leak levels were tested for each leak model (0%-87%). Resuscitation devices were connected to a test lung (compliance 0.6 mL/cmH2O) and respiratory parameters were measured using respiratory function monitors before (patient interface) and after (actual) an induced leak at 40, 60, 80 inflations/min.ResultsThree thousand six hundred inflations were analyzed. DLM showed a decrease in actual tidal volumes from 0%-87% leak with tidal volume differences (SIB 4.8 mL, TPR 2.9 mL), contrasting to minimal change for CLM (SIB -0.6 mL, TPR 0.3 mL). CLM demonstrated larger differences between patient interface and actual leak. The absolute difference at 60 inflations/min at 87% leak were SIB 37.5%, TPR 18.2% for CLM compared to SIB 4.6%, TPR 1.4% for DLM.ConclusionsCLM may underestimate the impact of resuscitation device performance with poor correlation between patient interface and actual delivered volume. DLM demonstrates several advantages with a more accurate representation of face mask leak and will prove useful in modeling all systems delivering PPV.
引用
收藏
页数:8
相关论文
共 27 条
[1]  
( ANZCOR) AaNZCoR, 2021, ANZCOR.org: ANZCOR
[2]   Unraveling the Links Between the Initiation of Ventilation and Brain Injury in Preterm Infants [J].
Barton, Samantha K. ;
Tolcos, Mary ;
Miller, Suzie L. ;
Roehr, Charles C. ;
Schmolzer, Georg M. ;
Davis, Peter G. ;
Moss, Timothy J. M. ;
LaRosa, Domenic A. ;
Hooper, Stuart B. ;
Polglase, Graeme R. .
FRONTIERS IN PEDIATRICS, 2015, 3
[3]   2023 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations: Summary From the Basic Life Support; Advanced Life Support; Pediatric Life Support; Neonatal Life Support; Education, Implementation, and Teams; and First Aid Task Forces [J].
Berg, Katherine M. ;
Bray, Janet E. ;
Ng, Kee-Chong ;
Liley, Helen G. ;
Greif, Robert ;
Carlson, Jestin N. ;
Morley, Peter T. ;
Drennan, Ian R. ;
Smyth, Michael ;
Scholefield, Barnaby R. ;
Weiner, Gary M. ;
Cheng, Adam ;
Djaerv, Therese ;
Abelairas-Gomez, Cristian ;
Acworth, Jason ;
Andersen, Lars W. ;
Atkins, Dianne L. ;
Berry, David C. ;
Bhanji, Farhan ;
Bierens, Joost ;
Bittencourt Couto, Thomaz ;
Borra, Vere ;
Boettiger, Bernd W. ;
Bradley, Richard N. ;
Breckwoldt, Jan ;
Cassan, Pascal ;
Chang, Wei-Tien ;
Charlton, Nathan P. ;
Chung, Sung Phil ;
Considine, Julie ;
Costa-Nobre, Daniela T. ;
Couper, Keith ;
Dainty, Katie N. ;
Dassanayake, Vihara ;
Davis, Peter G. ;
Dawson, Jennifer A. ;
de Almeida, Maria Fernanda ;
De Caen, Allan R. ;
Deakin, Charles D. ;
Dicker, Bridget ;
Douma, Matthew J. ;
Eastwood, Kathryn ;
El-Naggar, Walid ;
Fabres, Jorge G. ;
Fawke, Joe ;
Fijacko, Nino ;
Finn, Judith C. ;
Flores, Gustavo E. ;
Foglia, Elizabeth E. ;
Folke, Fredrik .
CIRCULATION, 2023, 148 (24) :E187-E280
[4]   Early initiation of basic resuscitation interventions including face mask ventilation may reduce birth asphyxia related mortality in low-income countries A prospective descriptive observational study [J].
Ersdal, Hege Langli ;
Mduma, Estomih ;
Svensen, Erling ;
Perlman, Jeffrey M. .
RESUSCITATION, 2012, 83 (07) :869-873
[5]   Assessment of volume and leak measurements during CPAP using a neonatal lung model [J].
Fischer, H. S. ;
Roehr, C. C. ;
Proquitte, H. ;
Wauer, R. R. ;
Schmalisch, G. .
PHYSIOLOGICAL MEASUREMENT, 2008, 29 (01) :95-107
[6]   Respiratory function monitoring during neonatal resuscitation: A systematic review [J].
Fuerch, Janene H. ;
Thio, Marta ;
Halamek, Louis P. ;
Liley, Helen G. ;
Wyckoff, Myra H. ;
Rabi, Yacov .
RESUSCITATION PLUS, 2022, 12
[7]   Delivery of Positive End-Expiratory Pressure Using Self-Inflating Bags during Newborn Resuscitation Is Possible Despite Mask Leak [J].
Gomo, Oystein Herwig ;
Eilevstjonn, Joar ;
Holte, Kari ;
Yeconia, Anita ;
Kidanto, Hussein ;
Ersdal, Hege Langli .
NEONATOLOGY, 2020, 117 (03) :341-348
[8]   Leak during Manual Neonatal Ventilation and Its Effect on the Delivered Pressures and Volumes: An in vitro Study [J].
Hartung, J. C. ;
te Pas, A. B. ;
Fischer, H. ;
Schmalisch, G. ;
Roehr, C. C. .
NEONATOLOGY, 2012, 102 (03) :190-195
[9]   Newborn resuscitation devices: The known unknowns and the unknown unknowns [J].
Hinder, Murray ;
Tracy, Mark .
SEMINARS IN FETAL & NEONATAL MEDICINE, 2021, 26 (02)
[10]   Performance of Leak Compensation in All-Age ICU Ventilators During Volume-Targeted Neonatal Ventilation: A Lung Model Study [J].
Itagaki, Taiga ;
Bennett, Desmond J. ;
Chenelle, Christopher T. ;
Fisher, Daniel F. ;
Kacmarek, Robert M. .
RESPIRATORY CARE, 2017, 62 (01) :10-21