Evaluation of the impact of cardiac implantable electronic devices on cine MRI for real-time adaptive cardiac radioablation on a 1.5 T MR-linac

被引:0
|
作者
Akdag, Osman [1 ]
Mandija, Stefano [1 ,2 ]
Borman, Pim T. S. [1 ]
Tzitzimpasis, Paris [1 ]
van Lier, Astrid L. H. M. W. [1 ]
Keesman, Rick [3 ]
Raaymakers, Bas W. [1 ]
Fast, Martin F. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Ctr Image Sci, Computat Imaging Grp MR Diag & Therapy, Utrecht, Netherlands
[3] Radboud Univ Nijmegen, Med Ctr, Dept Radiotherapy, Nijmegen, Netherlands
关键词
cardiac radiotherapy; MRI-guided radiotherapy; real-time; MAGNETIC-RESONANCE; TREATING PATIENTS; RADIATION; SAFETY;
D O I
10.1002/mp.17438
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Stereotactic arrhythmia radioablation (STAR) is a novel treatment approach for refractory ventricular tachycardia (VT). The risk of treatment-induced toxicity and geographic miss can be reduced with online MRI-guidance on an MR-linac. However, most VT patients carry cardiac implantable electronic devices (CIED), which compromise MR images. Purpose: Robust MR-linac imaging sequences are required for cardiac visualization and accurate motion monitoring in presence of a CIED during MRI-guided STAR. We optimized two clinically available cine sequences for cardiorespiratory motion estimation in presence of a CIED on a 1.5 T MR-linac. The image quality, motion estimation accuracy, and geometric fidelity using these cine sequences were evaluated. Methods: Clinically available 2D balanced steady-state free precession (bSSFP, voxel size = 3.0 3.0 10 mm(3), T-scan = 96 ms, bandwidth (BW) = 1884 Hz/px) and -spoiled gradient echo (-GRE, voxel size = 4.0 4.0 10 mm(3), T-scan = 97 ms, BW = 500 Hz/px) sequences were adjusted for real-time cardiac visualization and cardiorespiratory motion estimation on a 1.5 T Unity MR-linac (Elekta AB, Stockholm, Sweden), while complying with safety guidelines for MRI in presence of CIEDs (specific absorption rate < 2 W/kg and dB/dt< 80 mT/s). Cine acquisitions were performed in five healthy volunteers, with and without an implantable cardioverter- defibrillator (ICD) placed on the clavicle, and a VT patient. Generalized divergence-curl (GDC) deformable image registration (DIR) was used for automated landmark motion estimation in the left ventricle (LV). Gaussian processes (GP), a machine-learning technique, was trained using GDC landmarks and deployed for real-time cardiorespiratory motion prediction. B0-mapping was performed to assess geometric image fidelity in the presence of CIEDs. Results: CIEDs introduced banding artifacts partially obscuring cardiac structures in bSSFP acquisitions. In contrast, the T-1-GRE was more robust to CIED-induced artifacts at the expense of a lower signal-to-noise ratio. In presence of an ICD, image-based cardiorespiratory motion estimation was possible for 85% (100%) of the volunteers using the bSSFP (T-1-GRE) sequence. The in-plane 2D root-mean-squared deviation (RMSD) range between GDC-derived landmarks and manual annotations using the bSSFP (T-1-GRE) sequence was 3.1-3.3 (3.3-4.1) mm without ICD and 4.6-4.6 (3.2-3.3) mm with ICD. Without ICD, the RMSD between the GP-predictions and GDC-derived landmarks ranged between 0.9 and 2.2 mm (1.3-3.0 mm) for the bSSFP (T-1-GRE) sequence. With ICD, the RMSD between the GP-predictions and GDC-derived landmarks ranged between 1.3 and 2.2 mm (1.2-3.2 mm) using the bSSFP (T-1-GRE) sequence resulting in an RMSD-increase of 42%-143% (bSSFP) and -61%-142% (T-1-GRE). Lead-induced spatial distortions ranged between -0.2 and 0.2 mm (-0.7-1.2 mm) using the bSSFP (T1-GRE) sequence. The 98(th) percentile range of the spatial distortions in the gross target volume of the patient was between 0.0 and 0.4 mm (0.0-1.8 mm) when using bSSFP (T-1-GRE). Conclusions: Tailored bSSFP and T-1-GRE sequences can facilitate real-time cardiorespiratory estimation using GP trained with GDC-derived landmarks in the majority of landmark locations in the LV despite the presence of CIEDs. The need for high temporal resolution noticeably reduced achievable spatial resolution of the cine MRIs. However, the effect of the CIED-induced artifacts is device, patient and sequence dependent and requires specific assessment per case.
引用
收藏
页码:99 / 112
页数:14
相关论文
共 16 条
  • [1] Experimental demonstration of real-time cardiac physiology-based radiotherapy gating for improved cardiac radioablation on an MR-linac
    Akdag, Osman
    Borman, Pim T. S.
    Mandija, Stefano
    Woodhead, Peter L.
    Uijtewaal, Prescilla
    Raaymakers, Bas W.
    Fast, Martin F.
    MEDICAL PHYSICS, 2024, 51 (04) : 2354 - 2366
  • [2] Real-time 4D MRI using MR signature matching (MRSIGMA) on a 1.5T MR-Linac system
    Wu, Can
    Murray, Victor
    Siddiq, Syed S.
    Tyagi, Neelam
    Reyngold, Marsha
    Crane, Christopher
    Otazo, Ricardo
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (18)
  • [3] Evaluation of highly accelerated real-time cardiac cine MRI in tachycardia
    Bassett, Elwin C.
    Kholmovski, Eugene G.
    Wilson, Brent D.
    DiBella, Edward V. R.
    Dosdall, Derek J.
    Ranjan, Ravi
    McGann, Christopher J.
    Kim, Daniel
    NMR IN BIOMEDICINE, 2014, 27 (02) : 175 - 182
  • [4] First experimental exploration of real-time cardiorespiratory motion management for future stereotactic arrhythmia radioablation treatments on the MR-linac
    Akdag, O.
    Borman, P. T. S.
    Woodhead, P.
    Uijtewaal, P.
    Mandija, S.
    Van Asselen, B.
    Verhoeff, J. J. C.
    Raaymakers, B. W.
    Fast, M. F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (06)
  • [5] Highly Accelerated Real-Time Free-Breathing Cine CMR for Patients With a Cardiac Implantable Electronic Device
    Hong, Sungtak
    Hong, KyungPyo
    Culver, Austin E.
    Pathrose, Ashitha
    Allen, Bradley D.
    Wilcox, Jane E.
    Lee, Daniel C.
    Kim, Daniel
    ACADEMIC RADIOLOGY, 2021, 28 (12) : 1779 - 1786
  • [6] Free breathing cardiac real-time cine MR without ECG triggering
    Beer, Meinrad
    Stamm, Heimo
    Machann, Wolfram
    Weng, Andreas
    Goltz, Jan P.
    Breunig, Frank
    Weidemann, Frank
    Hahn, Dietbert
    Koestler, Herbert
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 145 (02) : 380 - 382
  • [7] Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE
    Feng, Li
    Srichai, Monvadi B.
    Lim, Ruth P.
    Harrison, Alexis
    King, Wilson
    Adluru, Ganesh
    Dibella, Edward V. R.
    Sodickson, Daniel K.
    Otazo, Ricardo
    Kim, Daniel
    MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (01) : 64 - 74
  • [8] Safety and Quality of 1.5-T MRI in Patients With Conventional and MRI-Conditional Cardiac Implantable Electronic Devices After Implementation of a Standardized Protocol
    Camacho, Juan C.
    Moreno, Courtney Coursey
    Shah, Anand D.
    Mittal, Pardeep K.
    Mengistu, Andenet
    Lloyd, Michael S.
    El-Chami, Mikhael F.
    Lerakis, Stamatios
    Saindane, Amit M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (03) : 599 - 604
  • [9] Real-time SPARSE-SENSE cardiac cine MR imaging: optimization of image reconstruction and sequence validation
    Goebel, Juliane
    Nensa, Felix
    Bomas, Bettina
    Schemuth, Haemi P.
    Maderwald, Stefan
    Gratz, Marcel
    Quick, Harald H.
    Schlosser, Thomas
    Nassenstein, Kai
    EUROPEAN RADIOLOGY, 2016, 26 (12) : 4482 - 4489
  • [10] Comparison of cardiac volumetry using real-time MRI during free-breathing with standard cine MRI during breath-hold in children
    Roewer, Lena Maria
    Radke, Karl Ludger
    Hussmann, Janina
    Malik, Halima
    Uelwer, Tobias
    Voit, Dirk
    Frahm, Jens
    Wittsack, Hans-Joerg
    Harmeling, Stefan
    Pillekamp, Frank
    Klee, Dirk
    PEDIATRIC RADIOLOGY, 2022, 52 (08) : 1462 - 1475