Fabrication of low-cost heterogeneous anthropomorphic thyroid neck phantom for CT

被引:0
作者
Akbar, Obaidullah Ehtesham [1 ]
Hameed, Kamran [1 ]
Jaffar, Hussain [1 ]
Yousef, Mauth [1 ]
Al Hashim, Maryam [2 ,3 ]
机构
[1] Imam Abdulrahman Bin Faisal Univ, Dept Biomed Engn, POB 1982, Dammam 31441, Saudi Arabia
[2] Imam Abdulrahman Bin Faisal Univ, Coll Med, Radiol Dept, POB 1982, Dammam 31441, Saudi Arabia
[3] King Fahad Specialist Hosp Dammam, Med Imaging Serv Ctr, Dammam 32253, Saudi Arabia
关键词
Bone - Computerized tomography - Dosimeters - Dosimetry - Electron density measurement - Muscle - Phantoms;
D O I
10.1016/j.radphyschem.2025.112752
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This article details the creation of a low-cost, heterogeneous anthropomorphic neck phantom with thyroid carcinoma for CT imaging. The phantom was designed to enable academic researchers and imaging system manufacturers to test and calibrate imaging machines and calculate dosimetry using thermoluminescent dosimeters in the neck and thyroid region. It was constructed using tissue-mimicking materials representing the thyroid, carcinoma, trachea, esophagus, spinal bones, muscle tissue, and adipose tissue. These materials were developed based on the elemental compositions outlined in the International Commission on Radiation Units and Measurements (ICRU) Report 44 and the International Commission on Radiological Protection Report 110 (ICRP). The phantom was modeled and segmented from a CT scan of a 52-year-old female thyroid cancer patient using segmentation software. Segmented parts were 3D printed and molded with silicone caulking, followed by the fabrication of tissue-mimicking materials cast into a cylinder filled with muscle tissue material. Validation showed that the tissue-mimicking materials generally achieved appropriate ionizing radiation parameters, such as tissue density, electron density, and effective atomic number. However, notable discrepancies were observed in the densities of the muscle and spinal bone materials compared to reference values. Mass attenuation coefficient (mu/rho) graphs generated via the XCOM database showed minimal deviations from reference data provided by the ICRU and the ICRP. Validation under a standard thyroid imaging protocol on a CT machine demonstrated reasonable anatomical accuracy and Hounsfield unit ratios between tissues. However, challenges in achieving consistent material properties for muscle and spinal bone materials were identified, suggesting areas for future improvement.
引用
收藏
页数:12
相关论文
共 25 条
  • [1] The Mass Attenuation Coefficients, Electronic, Atomic, and Molecular Cross Sections, Effective Atomic Numbers, and Electron Densities for Compounds of Some Biomedically Important Elements at 59.5 keV
    Akca, Burcu
    Erzeneoglu, Salih Z.
    [J]. SCIENCE AND TECHNOLOGY OF NUCLEAR INSTALLATIONS, 2014, 2014
  • [2] Evaluation of 3D printing materials for fabrication of a novel multi-functional 3D thyroid phantom for medical dosimetry and image quality
    Alssabbagh, Moayyad
    Tajuddin, Abd Aziz
    Abdulmanap, Mahayuddin
    Zainon, Rafidah
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2017, 135 : 106 - 112
  • [3] Thyroid cancer
    Cabanillas, Maria E.
    McFadden, David G.
    Durante, Cosimo
    [J]. LANCET, 2016, 388 (10061) : 2783 - 2795
  • [4] HG M., 2009, Ann ICRP, DOI DOI 10.1016/J.ICRP.2009.09.001
  • [5] X-RAY ATTENUATION COEFFICIENTS OF ELEMENTS AND MIXTURES
    JACKSON, DF
    HAWKES, DJ
    [J]. PHYSICS REPORTS-REVIEW SECTION OF PHYSICS LETTERS, 1981, 70 (03): : 169 - 233
  • [6] Karimi M, 2018, J Biomed Phys Eng, V8, P217
  • [7] Kirk S., 2016, The cancer genome atlas thyroid cancer collection (TCGA-THCA)
  • [8] Development and validation of the low-cost pregnant female physical phantom for fetal dosimetry in MV photon radiotherapy
    Kopacin, Vjekoslav
    Brkic, Hrvoje
    Ivkovic, Ana
    Kasabasic, Mladen
    Knezevic, Zeljka
    Majer, Marija
    Nodilo, Marijana
    Turk, Tajana
    Faj, Dario
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (02):
  • [9] Reproduction of a conventional anthropomorphic female chest phantom by 3D-printing: Comparison of image contrasts and absorbed doses in CT
    Kunert, Patrizia
    Schlattl, Helmut
    Trinkl, Sebastian
    Giussani, Augusto
    Klein, Lea
    Janich, Martin
    Reichert, Detlef
    Brix, Gunnar
    [J]. MEDICAL PHYSICS, 2023, 50 (08) : 4734 - 4743
  • [10] Tissue equivalence of 3D printing materials with respect to attenuation and absorption of X-rays used for diagnostic and interventional imaging
    Kunert, Patrizia
    Trinkl, Sebastian
    Giussani, Augusto
    Reichert, Detlef
    Brix, Gunnar
    [J]. MEDICAL PHYSICS, 2022, 49 (12) : 7766 - 7778